Skip to content

Equinoxe 314-02-13 (UDI 10885862084514)Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Equinoxe (305)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10885862084514), received between Aug 3, 2021 and Aug 1, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

9 other recalls cover devices cleared under the same FDA submission (K042021); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 6, 2024Open, Classified

    The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

    Z-1405-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth3
  • Fracture1
  • Naturally Worn1

Patient problems

  • Aseptic loosening2
  • Arthropathy1
  • Failure of Implant1
  • Inadequate Osseointegration1
  • Joint Laxity1
  • Muscle/Tendon Damage1
  • Osteolysis1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 1, 2025Fracture; Loosening of Implant Not Related to Bone-Ingrowth; Fracture; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED, APPROXIMATELY 11 YEARS AND 11 MONTHS POST THE INITIAL LEFT TOTAL SHOULDER ARTHROPLASTY, THE PATIENT COMPLAINED ABOUT LOOSENING IN THE SHOULDER JOINT. THE ROTATOR CUFF FAILED, LEADING TO GLENOID LOOSENING. EVERYTHING WAS REMOVED. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: Arthropathy; Aseptic loosening; Joint Laxity; Muscle/Tendon Damage

    MAUDE report 22690345 ↗
  • InjuryMar 18, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL TOTAL SHOULDER REPLACEMENT ON THE RIGHT SIDE. SUBSEQUENTLY, THE PATIENT EXPERIENCED ASEPTIC GLENOID LOOSENING, FAILED ATSA. AS A RESULT, APPROXIMATELY 13 YEARS AFTER INITIAL REPLACEMENT, THE PATIENT UNDERWENT A RIGHT SHOULDER REVISION. NO FURTHER IMPACT TO THE PATIENT WAS REPORTED. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: Aseptic loosening; Inadequate Osseointegration

    MAUDE report 21636681 ↗
  • InjuryNov 4, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THIS PATIENT'S RIGHT SHOULDER WAS REVISED APPROXIMATELY 8 YEARS POST OP. PATIENT HAD FAILED ANATOMIC SHOULDER REPLACEMENT. SEVERE GLENOID LOOSENING DUE TO OSTEOLYSIS. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. PRODUCTS NOT RETURNING: DISPOSED.

    Patient problems: Osteolysis

    MAUDE report 20599991 ↗
  • InjuryAug 3, 2021Naturally Worn; Naturally Worn

    AS REPORTED. APPROXIMATELY 12 YRS POSTOP THE INITIAL THIS PATIENT WAS CONVERTED TO A REVERSE SHOULDER YESTERDAY. CEMENTED GLENOID WAS WORN EXTENSIVELY IN THE SUPERIOR HALF WITH OSTEOLYSIS NOTED ON HUMERUS AND GLENOID. ALL COMPONENTS WERE REMOVED. REVERSE SHOULDER WAS THEN PERFORMED. THE SURGEON WAS PLEASED WITH THE EXTENDED BASEPLATE FIXATION. A SHORT STEM WAS CEMENTED IN THE HUMERUS. PATIENT WAS LEFT IN STABLE CONDITION. NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 12266275 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K042021. One submission often covers many models, so check the product and lot details in each record.

  • Z-0535-2026 · Sep 5, 2025 · Open, Classified
    Brand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 42mm Model/Catalog Number: 321-59-42 Software Version: N/A Product Description: rTSA Reamer, 42mm Component: N
  • Z-0536-2026 · Sep 5, 2025 · Open, Classified
    Brand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 46mm Model/Catalog Number: 321-59-46 Software Version: N/A Product Description: rTSA Reamer, 46mm Component: N
  • Z-0537-2026 · Sep 5, 2025 · Open, Classified
    Brand Name: Equinoxe Ergo Product Name: Ergo Augments Pilot Tip Converter Model/Catalog Number: 301-11-11 Software Version: N/A Product Description: Pilot Tip Converter Piece Compo
  • Z-1404-2024 · Mar 6, 2024 · Open, Classified
    Exactech Equinoxe GLENOID,KEELED, CEMENTED, mates with 47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-02-02, b) Medium, Item Number 314-02-03, c) Large, Item Nu
  • Z-1402-2024 · Mar 6, 2024 · Open, Classified
    Exactech Equinoxe GLENOID,KEELED, CEMENTED, Mates with 38, 41, 44 head, alpha curvature, Sizes: a), Small, Item Number 314-01-02; b) Medium, Item number 314-01-03; c) Large, Item
  • Z-1403-2024 · Mar 6, 2024 · Open, Classified
    Exactech Equinoxe GLENOID,KEELED, CEMENTED, mates with 47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-01-12, b) Medium, Item Number 314-01-13, c) Large, Item Nu
  • Z-0884-2021 · Nov 9, 2020 · Open, Classified
    Equinoxe Humeral Stem, Primary, Press Fit, 11mm
  • Z-1459-2015 · Feb 13, 2015 · Terminated
    Equinoxe 44mm Short Humeral Head Catalog Number 310-01-44. Orthopedic shoulder implant component.
  • Z-2115-2012 · Jun 15, 2012 · Terminated
    Equinoxe Anatomic Replicator Plate Intended for Primary and Reverse Total Shoulder Arthroplasty

Other Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices

All Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page