Skip to content

Equinoxe 314-02-23Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Equinoxe (305)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
2

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885862175069), received between Sep 1, 2023 and Feb 20, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 6, 2024Open, Classified

    The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

    Z-1407-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  2. Mar 6, 2024Open, Classified

    The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

    Z-1406-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1
  • Insufficient Device Problem Information1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Unstable1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Aseptic loosening1
  • Insufficient Information1
  • Joint Dislocation1
  • Pain1
  • Subluxation1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 20, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    APPROXIMATELY 12 YEAR(S) AND 7 DAY(S) POST-OPERATIVE OF A LEFT RTSA, IT WAS REPORTED VIA CLINICAL STUDY THAT THE PATIENT EXPERIENCED ASEPTIC GLENOID LOOSENING WITH GRADUAL INCREASED PAIN, DECREASED FUNCTION, AND LOOSENING ON X-RAY. THE PATIENT UNDERWENT A STANDARD TOTAL REVISION WITH THE REPORTED REMOVAL OF THE REPLICATOR PLATE, TORQUE SCREW, HUMERAL HEAD, AND GLENOID COMPONENTS. ADDITIONALLY, THE PATIENT WAS REVISED TO A HEMIARTHROPLASTY AND THE OUTCOME OF THIS EVENT IS CONSIDERED RESOLVED. THE CASE REPORT FORM INDICATES THAT THIS EVENT IS UNLIKELY RELATED TO THE DEVICE(S) AND/OR TO THE PROCEDURE. THIS WAS REPORTED THROUGH CLINICAL TRIAL STUDY DATA COLLECTION ACTIVITIES, NO DEVICE RETURN IS ANTICIPATED, AND NO OTHER INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Aseptic loosening; Pain

    MAUDE report 21428194 ↗
  • InjuryDec 5, 2023Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    WHILE INVESTIGATING A 2023 REVISION FOR THIS PATIENT, IT WAS FOUND THAT A 63 YO MALE PATIENT HAD A PRIOR REVISION ON (B)(6) 2018. NO FURTHER INFORMATION KNOWN.

    Patient problems: Insufficient Information

    MAUDE report 18268505 ↗
  • InjurySep 1, 2023Unstable; Device Dislodged or Dislocated; Unstable; Device Dislodged or Dislocated

    IT WAS REPORTED BY THE EQUINOXE SHOULDER STUDY THAT THIS 56 Y/O MALE PATIENT HAD A SHOULDER DISLOCATION. POST OP X-RAY REVEALED SHOULDER WAS DISLOCATED. ACTION TAKEN: REDUCED IN THEATRE NEXT DAY. PATIENT WAS DISCHARGED IN A SLING ON (B)(6) 2019. THE CASE REPORT FORM INDICATES THIS EVENT IS POSSIBLY RELATED TO DEVICES AND DEFINITELY RELATED TO PROCEDURE. THIS EVENT REPORT WAS RECEIVED THROUGH CLINICAL DATA COLLECTION ACTIVITIES.

    Patient problems: Joint Dislocation; Subluxation

    MAUDE report 17669421 ↗

Other Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices

All Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page