Equinoxe 314-02-33Recalls & reported problems
Exactech, Inc. · Class 2 · Implant
All models of Equinoxe (305)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10885862175106), received between Dec 14, 2023 and Jul 6, 2026 — classified as 5 injuries.
The most frequently reported specific device problem is “Naturally Worn” (4 reports).
2 other recalls cover devices cleared under the same FDA submission (K103419); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Naturally Worn4
- Loosening of Implant Not Related to Bone-Ingrowth3
- Adverse Event Without Identified Device or Use Problem1
- Migration1
Patient problems
- Insufficient Information2
- Erosion1
- Failure of Implant1
- Joint Impingement1
- Muscle/Tendon Damage1
- Pain1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJul 6, 2026Naturally Worn
IT WAS REPORTED THAT A PATIENT, WHO HAD RIGHT TSA UNDERWENT A 1ST STAGE REVISION PROCEDURE. BASED ON THE CT SCAN, IT APPEARS THAT THE HUMERAL HEAD WAS NO LONGER FULLY ARTICULATING ON THE POLY GLENOID AND HAS ERODED INTO THE POSTERIOR ASPECT OF THE GLENOID. THE 2ND STAGE WILL BE A CUSTOM-MADE GLENOID COMPONENT OWING TO THE SEVERITY OF BONE LOSS, RETAINING THE HUMERAL STEM. THERE WERE NO SURGICAL DELAYS OR DEVICE BREAKAGE DURING THE PROCEDURE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO X-RAYS WERE ABLE TO BE OBTAINED. THE EXPLANTED DEVICES ARE AVAILABLE FOR RETURN. NO DEVICE IMAGES WERE PROVIDED. NO FURTHER INFORMATION. THIS IS 1 OF 2 REPORTS FOR THIS EVENT.
Patient problems: Erosion
MAUDE report 25725611 ↗InjuryMar 17, 2025Naturally Worn; Adverse Event Without Identified Device or Use Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS PATIENT'S RIGHT SHOULDER WAS REVISED APPROXIMATELY 8 YEARS 8 MONTHS POST INITIAL OPERATION. THE IMPLANTS WERE REMOVED DUE TO A ROTATOR CUFF TEAR. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.
Patient problems: Muscle/Tendon Damage
MAUDE report 21627492 ↗InjuryJan 13, 2025Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Migration; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Migration
IT WAS REPORTED THAT THIS PATIENT'S RIGHT SHOULDER WAS REVISED. THE PATIENT HAD A HUMERAL RESURFACING AND OVER TIME THE GLENOID COMPONENT FAILED. THE SURGEON CARRIED OUT A 1ST STAGE REVISION WITH THE INTENT TO DO A 2ND STAGE INSERTING AN EXACTECH REVERSE SHOULDER. THE PATIENT HAS A CEMENT SPACER INSITU AND IS AWAITING A 2ND STAGE REVISION.
Patient problems: Insufficient Information; Pain
MAUDE report 21138025 ↗InjuryNov 5, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
STUDY: (B)(6) SUBJECT: (B)(4). IMPLANT DATE: (B)(6) 2013; EX-PLANT DATE: (B)(6) 2023; DATE OF EVENT ONSET: 01-30-2023. APPROXIMATELY 9 YEAR(S), 10 MONTH(S) AND 17 DAY(S) POST-OPERATIVE OF A RIGHT TSA, IT WAS REPORTED VIA CLINICAL STUDY THAT THE PATIENT EXPERIENCED ASEPTIC GLENOID LOOSENING WITH MECHANICAL CATCHING. THE PATIENT UNDERWENT A STANDARD TOTAL REVISION WITH THE REPORTED REMOVAL OF THE REPLICATOR PLATE. THE CLINICAL REPORT INDICATES THAT THIS EVENT IS POSSIBLY RELATED TO THE DEVICE(S) AND/OR TO THE PROCEDURE. THIS EVENT WAS REPORTED THROUGH CLINICAL TRIAL STUDY DATA COLLECTION ACTIVITIES AND NO OTHER INFORMATION IS AVAILABLE AT THIS TIME. CONCOMITANT DEVICE(S): 300-01-11 - EQUINOXE, HUMERAL STEM PRIMARY, PRESS FIT 11MM 2585154. 310-01-50 - EQUINOXE, HUMERAL HEAD SHORT, 50MM (BETA) 1631444. 314-02-33 - UHMWPE POST AUG GLENOID MEDIUM, RIGHT 2149790.
Patient problems: Joint Impingement
MAUDE report 20612692 ↗InjuryDec 14, 2023Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED, APPROXIMATELY ONE YEAR POST INITIAL RIGHT TSA, THE 59 Y/O MALE PATIENT WAS REVISED AS THE INITIAL IMPLANT WAS POORLY CEMENTED IN. THERE WAS NO CEMENT IN THE PEG HOLES TO HOLD THE IMPLANT IN PLACE. ALL COMPONENTS, EXCEPT FOR THE HUMERAL STEM, WERE REVISED. THE IMPLANTS WERE CHANGED FROM A SHOULDER ANATOMIC TO A SHOULDER REVERSE. A 36MM LINER WAS IMPLANTED. THERE WAS NO BREAKAGE OF DEVICE OR SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO OTHER PATIENT INFORMATION/MEDICAL HISTORY REPORTED. IMAGES RECEIVED. SALES REP WAS UNABLE TO OBTAIN X-RAYS. THE DEVICES ARE NOT AVAILABLE FOR EVALUATION DUE TO THE DEVICES WERE DISPOSED AT HOSPITAL.
Patient problems: Failure of Implant; Insufficient Information
MAUDE report 18331029 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K103419. One submission often covers many models, so check the product and lot details in each record.
- Z-1406-2024 · Mar 6, 2024 · Open, ClassifiedExactech Equinoxe GLENOID, POSTERIOR AUGMENT,PEGGED, CEMENTED, 8o, LEFT, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-02-22, b) 41, 44, 47, 50 head, Medium, Item
- Z-1407-2024 · Mar 6, 2024 · Open, ClassifiedExactech Equinoxe GLENOID, POSTERIOR AUGMENT,PEGGED, CEMENTED, 8o, RIGHT, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-02-32, b) 41, 44, 47, 50 head, Medium, Item
Other Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page