Equinoxe 315-07-06Recalls & reported problems
Exactech, Inc. · IMPACTOR TIP, GLENOID · Class 2
All models of Equinoxe (305)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10885862251824), received between May 8, 2020 and May 13, 2024 — classified as 5 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
10 other recalls cover devices cleared under the same FDA submission (K042021, K093430); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break7
- Fracture1
Patient problems
- No Consequences Or Impact To Patient5
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryMay 13, 2024Break; Fracture; Break; Fracture
IT WAS REPORTED THAT AFTER CEMENTING IN A CAGED GLENOID THE SURGEON SWITCHED FROM THE SPIDER IMPACTOR TO THE NORMAL IMPACTOR TO PRESSURISE THE IMPLANT. AFTER APPROX 5-10 MINUTES HE IMPACTED THE HANDLE ONCE MORE WITH A MALLET AND THE TIP BROKE. HE COLLATED ALL THE BROKEN PARTS MADE SURE HE GOT EVERYTHING OUT, WASHED OUT THE WOUND AND PROCEEDED WITH THE REST OF THE OPERATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19296351 ↗MalfunctionDec 21, 2021Break; Break
AS REPORTED, DURING THE SURGEON¿S ROUTINE IMPACTION OF AN ANATOMIC GLENOID IMPLANT, THE IMPACTOR TIP BROKE INTO TWO PIECES WHICH WERE BOTH EASILY RETRIEVED. NOTHING FELL INTO THE WOUND SITE. THE (B)(6) MALE WAS NOT AFFECTED BY THE BREAK. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. DEVICE IS TO BE RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13052787 ↗InjuryFeb 18, 2021Break; Break
DURING A SHOULDER PROCEDURE ON A MALE PATIENT, THE TRI DRIVE COLLAR BROKE DURING GLENOID REAMING. CAUSING THE NEED TO CHANGE REAMER HANDLE AND REMOVE BALL BEARING AND RING FROM OPEN WOUND. DURING THE SAME PROCEDURE, WHEN IMPACTING THE GLENOID, THE GLENOID IMPACTOR TIP BROKE AND ALSO HAD PARTS REMOVED FROM WOUND. NO HEALTH COMPLICATIONS FROM EXPERIENCE, JUST SURGICAL DELAY. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. DEVICES WILL RETURN. TRI DRIVE (315-25-00) STATED ON EXPERIENCE REPORT IS BEING CAPTURED UNDER (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11342061 ↗InjuryFeb 18, 2021Break; Break
DURING A SHOULDER PROCEDURE ON A MALE PATIENT, THE TRI DRIVE COLLAR BROKE DURING GLENOID REAMING. CAUSING THE NEED TO CHANGE REAMER HANDLE AND REMOVE BALL BEARING AND RING FROM OPEN WOUND. DURING THE SAME PROCEDURE, WHEN IMPACTING THE GLENOID, THE GLENOID IMPACTOR TIP BROKE AND ALSO HAD PARTS REMOVED FROM WOUND. NO HEALTH COMPLICATIONS FROM EXPERIENCE, JUST SURGICAL DELAY. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. DEVICES WILL RETURN. TRI DRIVE (315-25-00) STATED ON EXPERIENCE REPORT IS BEING CAPTURED UNDER CASE (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11342057 ↗MalfunctionJun 25, 2020Break; Break
THE GLENOID IMPACTOR TIP IS BROKEN. THE PATIENT IS FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10194606 ↗InjuryMay 8, 2020Break; Break
AS REPORTED, DURING A PROCEDURE ON A (B)(6) Y/O MALE PATIENT, THE SURGEON IMPACTED THE GLENOID COMPONENT WITH THE PLASTIC BLACK TIP, IT FRACTURED IN HALF. THE BROKEN IMPACTOR TIP PIECES WERE REMOVED FROM THE WOUND. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE DEVICE IS AVAILABLE FOR RETURN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10038630 ↗InjuryMay 8, 2020Break; Break
AS REPORTED, DURING A SHOULDER PROCEDURE ON THIS (B)(6) Y/O MALE, THE SECONDARY GLENOID IMPACTOR BROKE WHILE IMPACTING THE GLENOID. ALL BROKEN PIECES WERE REMOVED FROM THE WOUND. THE BROKEN PIECE WAS THROWN AWAY AFTER THE CASE. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10038532 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K042021, K093430. One submission often covers many models, so check the product and lot details in each record.
- Z-0536-2026 · Sep 5, 2025 · Open, ClassifiedBrand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 46mm Model/Catalog Number: 321-59-46 Software Version: N/A Product Description: rTSA Reamer, 46mm Compone
- Z-0535-2026 · Sep 5, 2025 · Open, ClassifiedBrand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 42mm Model/Catalog Number: 321-59-42 Software Version: N/A Product Description: rTSA Reamer, 42mm Compone
- Z-0537-2026 · Sep 5, 2025 · Open, ClassifiedBrand Name: Equinoxe Ergo Product Name: Ergo Augments Pilot Tip Converter Model/Catalog Number: 301-11-11 Software Version: N/A Product Description: Pilot Tip Converter Piece
- Z-1404-2024 · Mar 6, 2024 · Open, ClassifiedExactech Equinoxe GLENOID,KEELED, CEMENTED, mates with 47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-02-02, b) Medium, Item Number 314-02-03, c) Large, Item Nu
- Z-1402-2024 · Mar 6, 2024 · Open, ClassifiedExactech Equinoxe GLENOID,KEELED, CEMENTED, Mates with 38, 41, 44 head, alpha curvature, Sizes: a), Small, Item Number 314-01-02; b) Medium, Item number 314-01-03; c) Large, Item
- Z-1403-2024 · Mar 6, 2024 · Open, ClassifiedExactech Equinoxe GLENOID,KEELED, CEMENTED, mates with 47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-01-12, b) Medium, Item Number 314-01-13, c) Large, Item Nu
- Z-1405-2024 · Mar 6, 2024 · Open, ClassifiedExactech Equinoxe GLENOID,KEELED, CEMENTED,47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-02-12, b) Medium, Item Number 314-02-13, c) Large, Item Number 314-02
- Z-0884-2021 · Nov 9, 2020 · Open, ClassifiedEquinoxe Humeral Stem, Primary, Press Fit, 11mm
- Z-1459-2015 · Feb 13, 2015 · TerminatedEquinoxe 44mm Short Humeral Head Catalog Number 310-01-44. Orthopedic shoulder implant component.
- Z-2115-2012 · Jun 15, 2012 · TerminatedEquinoxe Anatomic Replicator Plate Intended for Primary and Reverse Total Shoulder Arthroplasty
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page