Skip to content

Equinoxe 315-07-06Recalls & reported problems

Exactech, Inc. · IMPACTOR TIP, GLENOID · Class 2

All models of Equinoxe (305)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10885862251824), received between May 8, 2020 and May 13, 2024 — classified as 5 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

10 other recalls cover devices cleared under the same FDA submission (K042021, K093430); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Fracture1

Patient problems

  • No Consequences Or Impact To Patient5
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 13, 2024Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT AFTER CEMENTING IN A CAGED GLENOID THE SURGEON SWITCHED FROM THE SPIDER IMPACTOR TO THE NORMAL IMPACTOR TO PRESSURISE THE IMPLANT. AFTER APPROX 5-10 MINUTES HE IMPACTED THE HANDLE ONCE MORE WITH A MALLET AND THE TIP BROKE. HE COLLATED ALL THE BROKEN PARTS MADE SURE HE GOT EVERYTHING OUT, WASHED OUT THE WOUND AND PROCEEDED WITH THE REST OF THE OPERATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19296351 ↗
  • MalfunctionDec 21, 2021Break; Break

    AS REPORTED, DURING THE SURGEON¿S ROUTINE IMPACTION OF AN ANATOMIC GLENOID IMPLANT, THE IMPACTOR TIP BROKE INTO TWO PIECES WHICH WERE BOTH EASILY RETRIEVED. NOTHING FELL INTO THE WOUND SITE. THE (B)(6) MALE WAS NOT AFFECTED BY THE BREAK. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. DEVICE IS TO BE RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13052787 ↗
  • InjuryFeb 18, 2021Break; Break

    DURING A SHOULDER PROCEDURE ON A MALE PATIENT, THE TRI DRIVE COLLAR BROKE DURING GLENOID REAMING. CAUSING THE NEED TO CHANGE REAMER HANDLE AND REMOVE BALL BEARING AND RING FROM OPEN WOUND. DURING THE SAME PROCEDURE, WHEN IMPACTING THE GLENOID, THE GLENOID IMPACTOR TIP BROKE AND ALSO HAD PARTS REMOVED FROM WOUND. NO HEALTH COMPLICATIONS FROM EXPERIENCE, JUST SURGICAL DELAY. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. DEVICES WILL RETURN. TRI DRIVE (315-25-00) STATED ON EXPERIENCE REPORT IS BEING CAPTURED UNDER (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11342061 ↗
  • InjuryFeb 18, 2021Break; Break

    DURING A SHOULDER PROCEDURE ON A MALE PATIENT, THE TRI DRIVE COLLAR BROKE DURING GLENOID REAMING. CAUSING THE NEED TO CHANGE REAMER HANDLE AND REMOVE BALL BEARING AND RING FROM OPEN WOUND. DURING THE SAME PROCEDURE, WHEN IMPACTING THE GLENOID, THE GLENOID IMPACTOR TIP BROKE AND ALSO HAD PARTS REMOVED FROM WOUND. NO HEALTH COMPLICATIONS FROM EXPERIENCE, JUST SURGICAL DELAY. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. DEVICES WILL RETURN. TRI DRIVE (315-25-00) STATED ON EXPERIENCE REPORT IS BEING CAPTURED UNDER CASE (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11342057 ↗
  • MalfunctionJun 25, 2020Break; Break

    THE GLENOID IMPACTOR TIP IS BROKEN. THE PATIENT IS FINE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10194606 ↗
  • InjuryMay 8, 2020Break; Break

    AS REPORTED, DURING A PROCEDURE ON A (B)(6) Y/O MALE PATIENT, THE SURGEON IMPACTED THE GLENOID COMPONENT WITH THE PLASTIC BLACK TIP, IT FRACTURED IN HALF. THE BROKEN IMPACTOR TIP PIECES WERE REMOVED FROM THE WOUND. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE DEVICE IS AVAILABLE FOR RETURN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10038630 ↗
  • InjuryMay 8, 2020Break; Break

    AS REPORTED, DURING A SHOULDER PROCEDURE ON THIS (B)(6) Y/O MALE, THE SECONDARY GLENOID IMPACTOR BROKE WHILE IMPACTING THE GLENOID. ALL BROKEN PIECES WERE REMOVED FROM THE WOUND. THE BROKEN PIECE WAS THROWN AWAY AFTER THE CASE. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10038532 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K042021, K093430. One submission often covers many models, so check the product and lot details in each record.

  • Z-0536-2026 · Sep 5, 2025 · Open, Classified
    Brand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 46mm Model/Catalog Number: 321-59-46 Software Version: N/A Product Description: rTSA Reamer, 46mm Compone
  • Z-0535-2026 · Sep 5, 2025 · Open, Classified
    Brand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 42mm Model/Catalog Number: 321-59-42 Software Version: N/A Product Description: rTSA Reamer, 42mm Compone
  • Z-0537-2026 · Sep 5, 2025 · Open, Classified
    Brand Name: Equinoxe Ergo Product Name: Ergo Augments Pilot Tip Converter Model/Catalog Number: 301-11-11 Software Version: N/A Product Description: Pilot Tip Converter Piece
  • Z-1404-2024 · Mar 6, 2024 · Open, Classified
    Exactech Equinoxe GLENOID,KEELED, CEMENTED, mates with 47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-02-02, b) Medium, Item Number 314-02-03, c) Large, Item Nu
  • Z-1402-2024 · Mar 6, 2024 · Open, Classified
    Exactech Equinoxe GLENOID,KEELED, CEMENTED, Mates with 38, 41, 44 head, alpha curvature, Sizes: a), Small, Item Number 314-01-02; b) Medium, Item number 314-01-03; c) Large, Item
  • Z-1403-2024 · Mar 6, 2024 · Open, Classified
    Exactech Equinoxe GLENOID,KEELED, CEMENTED, mates with 47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-01-12, b) Medium, Item Number 314-01-13, c) Large, Item Nu
  • Z-1405-2024 · Mar 6, 2024 · Open, Classified
    Exactech Equinoxe GLENOID,KEELED, CEMENTED,47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-02-12, b) Medium, Item Number 314-02-13, c) Large, Item Number 314-02
  • Z-0884-2021 · Nov 9, 2020 · Open, Classified
    Equinoxe Humeral Stem, Primary, Press Fit, 11mm
  • Z-1459-2015 · Feb 13, 2015 · Terminated
    Equinoxe 44mm Short Humeral Head Catalog Number 310-01-44. Orthopedic shoulder implant component.
  • Z-2115-2012 · Jun 15, 2012 · Terminated
    Equinoxe Anatomic Replicator Plate Intended for Primary and Reverse Total Shoulder Arthroplasty

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page