GPS 531-07-05Recalls & reported problems
Exactech, Inc. · SHOULDER GPS IMPACTOR HANDLE · Class 2
All models of GPS (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885862542038), received between Feb 24, 2021 and Feb 7, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K162567, K173372); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Jun 5, 2023Open, Classified
The ball and spring housed within the locking mechanism of the handle can dislodge during disassembly or during impaction.
Z-2050-2023 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Break2
Patient problems
- No Patient Involvement1
- Unintended Radiation Exposure1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 7, 2022Break; Break
IT WAS REPORTED THAT THE BALL BEARING AND SPRING FROM THE GPS GLENOID IMPACTOR HANDLE DISLODGED INSIDE OF PATIENT WHEN THE SURGEON IMPACTED THE GLENOID BASEPLATE. THE SURGEON BECAME AWARE WHEN THE SPRING WAS LOCATED DURING THE CASE. POST OP AP X RAY TAKEN SHOWS NO BALL BEARING LEFT IN THE PATIENT. THERE WAS LESS THAN A 5 MINUTE SURGICAL DELAY. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.
Patient problems: Unintended Radiation Exposure
MAUDE report 13476909 ↗MalfunctionFeb 24, 2021Break; Break
AS REPORTED, THE STERILE PROCESSING STAFF MEMBER RECEIVED THE BASEPLATE STILL ATTACHED TO THE HANDLE. WHEN ATTEMPTING TO SEPARATE THE 2 PIECES, THE BALL BEARING SECURING THE SPRING AND LOCKING MECHANISM OF THE BASEPLATE HAS DROPPED OUT. THIS WAS DONE TO WASH THE INSTRUMENTS. THERE WAS NO PATIENT INVOLVEMENT AND THE DEVICES ARE RETURNING.
Patient problems: No Patient Involvement
MAUDE report 11371607 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K162567, K173372. One submission often covers many models, so check the product and lot details in each record.
- Z-0603-2018 · Jul 11, 2017 · TerminatedExactech GPS Total Shoulder Application 3.2mm Vix Bit Orthopedic surgical tool
Other System, image processing, radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record