Equinoxe 321-09-05Recalls & reported problems
Exactech, Inc. · MODULAR IMPACTOR HANDLE · Class 1
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885862557674), received between Oct 4, 2022 and Jan 9, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Feb 3, 2026Open, Classified
Impactor handle may be missing cross-pin
Z-1509-2026 · Root cause: Under Investigation by firm · FDA record ↗
- Jun 5, 2023Open, Classified
The ball and spring housed within the locking mechanism of the handle can dislodge during disassembly or during impaction.
Z-2051-2023 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Break3
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 9, 2024Break; Break
AS REPORTED, DURING IMPACTION, THE BALL BEARING IS MISSING AND WILL NO LONGER HOLD IMPACTOR TIPS. THERE WERE NO FRAGMENTS OR PIECES THAT FELL INTO THE PATIENT. THERE WAS NO SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. SALES REP WAS UNABLE TO OBTAIN PHOTOS/X-RAYS. THE DEVICE WILL BE RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18476705 ↗InjuryMay 9, 2023Break; Break
AS REPORTED, THE 78 Y/O MALE PATIENT RECEIVED A TOTAL SHOULDER REPLACEMENT. SURGEON NOTICED A BB AND SMALL SPRING ON PATIENT X-RAYS AND SENT THE IMAGES TO SALES REP. SALES REP RESEARCHED THE INSTRUMENTS AND NOTICED THE CATCH BUTTON MISSING ON THE MODULAR IMPACTOR HANDLE. THERE WAS NO SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. IMAGES RECEIVED. THE DEVICE IS AVAILABLE FOR EVALUATION. ADDITIONAL INFORMATION RECEIVED INDICATES THAT THE TIP BROKE DURING THE CASE, AND NO ONE NOTICED IT UNTIL AFTER SALES REP HAD ALREADY LEFT WITH THE CLEANED TRAYS. THE PATIENT IS OK, THE SURGEON STATES THAT HE AND THE PATIENT DECIDED TO LEAVE THE PIECES WHERE THEY ARE UNLESS THE PATIENT STARTS HAVING ISSUES. SURGEON CALLED SALES REP AND SAID HE WOULD KEEP WATCHING IT AS WELL, HE DOESN¿T EXPECT IT TO HURT OR IRRITATE ANYTHING. FURTHERMORE, THE DEVICE WAS USED LATER AT A DIFFERENT FACILITY AND NO ONE THERE NOTICED IT WAS BROKEN, BUT SOMEHOW THEY MADE IT WORK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16895059 ↗MalfunctionOct 4, 2022Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
DURING SURGICAL USE, WHEN THE BASEPLATE IMPACTOR TIP WAS REMOVED, A BEARING AND SECOND SMALL METAL PIECE FELL OUT OF THE END. ASSUMING THE CONSTRUCT OF THE HANDLE, IS A BEARING ON EACH END AND A METAL PIECE BETWEEN THEM, THE SECOND BEARING WAS NOT ACCOUNTED FOR. THE, PER DR PARSONS, CONCLUDED THAT IT WAS HIGHLY UNLIKELY THAT THE SECOND BEARING MADE IT INTO THE SURGICAL SITE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15537216 ↗
Other IMPACTOR devices
- PINNACLE 2217-50-041Depuy (Ireland) · IMPACTOR4,295 reports
- ATTUNE 2544-01-006Depuy (Ireland) · IMPACTOR1 recall2,539 reports
- ATTUNE 2544-01-003Depuy (Ireland) · IMPACTOR1 recall1,498 reports
- ATTUNE 2544-01-017Depuy (Ireland) · IMPACTOR1 recall1,456 reports
- ATTUNE 2544-01-004Depuy (Ireland) · IMPACTOR1 recall875 reports
- DURALOC 2360-71-000Depuy Orthopaedics, Inc. · IMPACTOR716 reports
- SPECIALIST 2581-11-000Depuy (Ireland) · IMPACTOR704 reports
- INHANCE 6205-10-103Depuy (Ireland) · IMPACTOR337 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.