Skip to content

Equinoxe 321-09-05Recalls & reported problems

Exactech, Inc. · MODULAR IMPACTOR HANDLE · Class 1

All models of Equinoxe (303)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885862557674), received between Oct 4, 2022 and Jan 9, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 3, 2026Class IIOpen, Classified

    Impactor handle may be missing cross-pin

    Z-1509-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Jun 5, 2023Class IIOpen, Classified

    The ball and spring housed within the locking mechanism of the handle can dislodge during disassembly or during impaction.

    Z-2051-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break3
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 9, 2024Break; Break

    AS REPORTED, DURING IMPACTION, THE BALL BEARING IS MISSING AND WILL NO LONGER HOLD IMPACTOR TIPS. THERE WERE NO FRAGMENTS OR PIECES THAT FELL INTO THE PATIENT. THERE WAS NO SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. SALES REP WAS UNABLE TO OBTAIN PHOTOS/X-RAYS. THE DEVICE WILL BE RETURN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18476705 ↗
  • InjuryMay 9, 2023Break; Break

    AS REPORTED, THE 78 Y/O MALE PATIENT RECEIVED A TOTAL SHOULDER REPLACEMENT. SURGEON NOTICED A BB AND SMALL SPRING ON PATIENT X-RAYS AND SENT THE IMAGES TO SALES REP. SALES REP RESEARCHED THE INSTRUMENTS AND NOTICED THE CATCH BUTTON MISSING ON THE MODULAR IMPACTOR HANDLE. THERE WAS NO SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. IMAGES RECEIVED. THE DEVICE IS AVAILABLE FOR EVALUATION. ADDITIONAL INFORMATION RECEIVED INDICATES THAT THE TIP BROKE DURING THE CASE, AND NO ONE NOTICED IT UNTIL AFTER SALES REP HAD ALREADY LEFT WITH THE CLEANED TRAYS. THE PATIENT IS OK, THE SURGEON STATES THAT HE AND THE PATIENT DECIDED TO LEAVE THE PIECES WHERE THEY ARE UNLESS THE PATIENT STARTS HAVING ISSUES. SURGEON CALLED SALES REP AND SAID HE WOULD KEEP WATCHING IT AS WELL, HE DOESN¿T EXPECT IT TO HURT OR IRRITATE ANYTHING. FURTHERMORE, THE DEVICE WAS USED LATER AT A DIFFERENT FACILITY AND NO ONE THERE NOTICED IT WAS BROKEN, BUT SOMEHOW THEY MADE IT WORK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16895059 ↗
  • MalfunctionOct 4, 2022Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    DURING SURGICAL USE, WHEN THE BASEPLATE IMPACTOR TIP WAS REMOVED, A BEARING AND SECOND SMALL METAL PIECE FELL OUT OF THE END. ASSUMING THE CONSTRUCT OF THE HANDLE, IS A BEARING ON EACH END AND A METAL PIECE BETWEEN THEM, THE SECOND BEARING WAS NOT ACCOUNTED FOR. THE, PER DR PARSONS, CONCLUDED THAT IT WAS HIGHLY UNLIKELY THAT THE SECOND BEARING MADE IT INTO THE SURGICAL SITE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15537216 ↗

FDA clearance and classification

The FDA regulates product code HWA (impactor) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Equinoxe 321-09-05

Has the Equinoxe 321-09-05 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-1509-2026 (Class II), began on Feb 3, 2026: Impactor handle may be missing cross-pin. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Equinoxe 321-09-05?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Oct 4, 2022 and Jan 9, 2024. In those reports, the most common specific device problems are “Break” (3) and “Detachment of Device or Device Component” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Equinoxe 321-09-05?

None of the reports are classified as deaths, 1 as an injury and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Equinoxe 321-09-05 FDA approved?

Devices of this type (product code HWA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Exactech, Inc.

Who makes the Equinoxe 321-09-05, and is it still sold?

The FDA's device database lists Exactech, Inc. as the company responsible for it (brand Equinoxe, model 321-09-05). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other IMPACTOR devices

All IMPACTOR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page