Skip to content

Equinoxe® 322-38-03Recalls & reported problems

Exactech, Inc. · Humeral Liner · Class 2 · Implant

All models of Equinoxe (305)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885862593849), received between Aug 5, 2025 and Aug 5, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Unstable” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 15, 2025Open, Classified

    Reverse Shoulder humeral liners have an articular surface position outside of the dimensional specification within the labeling.

    Z-2562-2025 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Unstable1

Patient problems

  • Joint Laxity1
  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 5, 2025Unstable; Unstable

    IT WAS REPORTED THAT A 60 YO MALE PATIENT, WHO HAD A REVERSE SHOULDER REVISION PRIOR, UNDERWENT ANOTHER REVERSE REVISION. THE PATIENT COMPLAINED OF PAIN. SURGEON DETERMINED IT WAS INSTABILITY OF THE POLY. A POLY SWAP WAS DONE. THERE WERE NO DEVICE BREAKAGES OR SURGICAL DELAYS DURING THE PROCEDURE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO X-RAYS WERE PROVIDED. THE EXPLANTED DEVICES ARE NOT AVAILABLE FOR RETURN DUE TO CHAIN OF COMMAND. NO DEVICE IMAGES WERE PROVIDED. THIS IS 1 OF 2 CASES FOR THIS PATIENT.

    Patient problems: Joint Laxity; Pain

    MAUDE report 22723489 ↗

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page