Equinoxe 314-13-14 (UDI 10885862657954)Recalls & reported problems
Exactech, Inc. · Glenoid, Cage Pegged, Beta · Class 2 · Implant
All models of Equinoxe (305)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885862657954), received between Apr 21, 2026 and Jun 12, 2026 — classified as 3 injuries.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth2
- Detachment of Device or Device Component1
- Naturally Worn1
Patient problems
- Pain2
- Metal Related Pathology1
- Muscle Weakness/Atrophy1
- Radiolucent Lines1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJun 12, 2026Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED VIA CLINICAL STUDY THAT PATIENT UNDERWENT A LEFT SHOULDER REPLACEMENT. SUBSEQUENTLY, THE PATIENT DEVELOPED ASEPTIC GLENOID LOOSENING, PRESENTING WITH SHOULDER PAIN. X-RAYS DEMONSTRATE STABLE ALIGNMENT WITH SLIGHT LUCENCY AT THE INFERIOR GLENOID PEGS. THE DEVICE REMAINS IMPLANTED AT THIS TIME AS A REVISION IS PLANNED FOR ON (B)(6) 2026. THE OUTCOME IS CONSIDERED CONTINUING. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Pain; Radiolucent Lines
MAUDE report 25485589 ↗InjuryMay 4, 2026Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT THE PATIENT UNDERWENT A LEFT SHOULDER REPLACEMENT AND WAS SUBSEQUENTLY REVISED. DURING THE REVISION PROCEDURE, THE GLENOID EXHIBITED METALLOSIS. THE GLENOID CAGE WAS NOTED TO BE LOOSE AND DID NOT REQUIRE THE USE OF PEG EXTRACTORS FOR REMOVAL. THERE WAS NOTED WEAR. ALL IMPLANTED COMPONENTS WERE EXPLANTED. THE SURGEON IMPLANTED A COMPETITOR¿S HUMERAL COMPONENTS ALONG WITH A CUSTOM GLENOID COMPONENT. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Metal Related Pathology
MAUDE report 25072085 ↗InjuryApr 21, 2026Detachment of Device or Device Component
AS REPORTED BY THE EQUINOXE SHOULDER STUDY, APPROXIMATELY 10 YEARS AND 2 MONTHS POST THE INITIAL RIGHT TOTAL SHOULDER ARTHROPLASTY, THE PATIENT WAS REVISED DUE TO DISASSOCIATION OF THE POLYETHYLENE. THE PATIENT HAD BEEN EXPERIENCING PAIN AND WEAKNESS SINCE 2024 WITHOUT A SPECIFIC MECHANISM OF INJURY/TRAUMA. TOTAL SHOULDER ARTHROPLASTY GLENOID POLY SHEAR FAILURE WAS FOUND ON X-RAY. THE REPLICATOR PLATE, TORQUE SCREW, HUMERAL HEAD AND GLENOID WERE REMOVED AND THE PATIENT WAS REVISED TO AN EXACTECH HYBRID IMPLANT. THE OUTCOME IS CONSIDERED TO BE RESOLVED.
Patient problems: Muscle Weakness/Atrophy; Pain
MAUDE report 24945325 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page