Skip to content

Equinoxe 314-13-14 (UDI 10885862657954)Recalls & reported problems

Exactech, Inc. · Glenoid, Cage Pegged, Beta · Class 2 · Implant

All models of Equinoxe (305)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885862657954), received between Apr 21, 2026 and Jun 12, 2026 — classified as 3 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Detachment of Device or Device Component1
  • Naturally Worn1

Patient problems

  • Pain2
  • Metal Related Pathology1
  • Muscle Weakness/Atrophy1
  • Radiolucent Lines1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 12, 2026Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED VIA CLINICAL STUDY THAT PATIENT UNDERWENT A LEFT SHOULDER REPLACEMENT. SUBSEQUENTLY, THE PATIENT DEVELOPED ASEPTIC GLENOID LOOSENING, PRESENTING WITH SHOULDER PAIN. X-RAYS DEMONSTRATE STABLE ALIGNMENT WITH SLIGHT LUCENCY AT THE INFERIOR GLENOID PEGS. THE DEVICE REMAINS IMPLANTED AT THIS TIME AS A REVISION IS PLANNED FOR ON (B)(6) 2026. THE OUTCOME IS CONSIDERED CONTINUING. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Pain; Radiolucent Lines

    MAUDE report 25485589 ↗
  • InjuryMay 4, 2026Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A LEFT SHOULDER REPLACEMENT AND WAS SUBSEQUENTLY REVISED. DURING THE REVISION PROCEDURE, THE GLENOID EXHIBITED METALLOSIS. THE GLENOID CAGE WAS NOTED TO BE LOOSE AND DID NOT REQUIRE THE USE OF PEG EXTRACTORS FOR REMOVAL. THERE WAS NOTED WEAR. ALL IMPLANTED COMPONENTS WERE EXPLANTED. THE SURGEON IMPLANTED A COMPETITOR¿S HUMERAL COMPONENTS ALONG WITH A CUSTOM GLENOID COMPONENT. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Metal Related Pathology

    MAUDE report 25072085 ↗
  • InjuryApr 21, 2026Detachment of Device or Device Component

    AS REPORTED BY THE EQUINOXE SHOULDER STUDY, APPROXIMATELY 10 YEARS AND 2 MONTHS POST THE INITIAL RIGHT TOTAL SHOULDER ARTHROPLASTY, THE PATIENT WAS REVISED DUE TO DISASSOCIATION OF THE POLYETHYLENE. THE PATIENT HAD BEEN EXPERIENCING PAIN AND WEAKNESS SINCE 2024 WITHOUT A SPECIFIC MECHANISM OF INJURY/TRAUMA. TOTAL SHOULDER ARTHROPLASTY GLENOID POLY SHEAR FAILURE WAS FOUND ON X-RAY. THE REPLICATOR PLATE, TORQUE SCREW, HUMERAL HEAD AND GLENOID WERE REMOVED AND THE PATIENT WAS REVISED TO AN EXACTECH HYBRID IMPLANT. THE OUTCOME IS CONSIDERED TO BE RESOLVED.

    Patient problems: Muscle Weakness/Atrophy; Pain

    MAUDE report 24945325 ↗

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates