Skip to content

ORBIT GALAXY 640CF0505Recalls & reported problems

Medos International Sàrl · ORBIT GALAXY Detachable Coil System with Tight Distal Loop technology 175cm d=2/3D D=5mm 5cm Contents: 1 detachable coil system and 1 luer valve. Coil Type: Complex Fill Note: ONLY use with TRUFILL DCS Syringe II. · Class 2

All models of ORBIT GALAXY (26)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886704030287), received between Jun 25, 2026 and Jun 25, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Separation Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 25, 2026Separation Failure

    A HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION, A COMPETITIVE MICROCATHETER (MC) WAS IN PLACE AN ORBIT GALAXY COIL (PRODUCT CODE: 640CF0505, LOT NUMBER: 31204208) WAS DELIVERED TO TARGET SITE. THE COIL WAS PACKED IN THE ANEURYSM AND TRIED TO DETACH, BUT THE COIL WAS UNABLE TO DETACH. THE DOCTOR ONLY RETRACTED THE COIL ALONE AND SWITCHED TO A SECOND COIL, ANOTHER ORBIT GALAXY (PRODUCT CODE: 640CF0510, LOT NUMBER: 31211764) AND TRIED TO DETACH IT, BUT THE SECOND COIL WAS ALSO UNABLE TO DETACH. THE DOCTOR ONLY RETRACTED THE SECOND COIL ALONE AND SWITCHED TO A THIRD COIL TO COMPLETE THE SURGERY. THE MICROCATHETER WAS NOT REPLACED. THERE WAS NO PATIENT INJURY REPORTED. ADDITIONAL EVENT INFORMATION RECEIVED ON 24-JUN-2026 INDICATED THAT A PRE-DEPLOYMENT ELECTRICAL TESTING WAS NOT PERFORMED. THE SAME DCB AND CONTROL CABLES WERE NOT USED WITH THE SUBSEQUENT COILS. THE COILS WERE ATTACHED TO THE COIL DELIVERY SYSTEMS WHEN REMOVED FROM THE PATIENT. THE COILS WERE NOT STRETCHED OR DAMAGED WHEN REMOVED FROM THE PATIENT. THERE WAS NO PROCEDURE PROLONGATION/DELAY DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25623147 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page