Skip to content

ORBIT GALAXY 640CF1230Recalls & reported problems

Medos International Sàrl · ORBIT GALAXY Detachable Coil System with Tight Distal Loop technology 175cm d=2/3D D=12mm 30cm Contents: 1 detachable coil system and 1 luer valve. Coil Type: Complex Fill Note: ONLY use with TRUFILL DCS Syringe II. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886704030423), received between Jan 28, 2026 and Jan 28, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Separation Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 28, 2026Separation Failure

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A COIL EMBOLIZATION, A 12X30 ORBIT GALAXY COIL (PRODUCT CODE 640CF1230, LOT NUMBER 31613399) WAS DELIVERED TO TARGET SITE. THE COIL WAS FILLED IN THE ANEURYSM AND TRIED TO DETACH, BUT THE COIL WAS UNABLE TO DETACH. THE DOCTOR ONLY RETRACTED THE COIL ALONE AND SWITCHED A NEW ONE TO COMPLETE THE SURGERY. THE NON J&J MICROCATHETER (MC) WAS NOT REPLACED. ADDITIONAL EVENT INFORMATION RECEIVED ON 15-JAN-2026 INDICATED THAT THE SYRINGE USED WAS A JOHNSON & JOHNSON PRODUCT. THE SAME SYRINGE WAS USED TO DETACH SUBSEQUENT COILS. THE SAME SYRINGE WAS PREVIOUSLY EXCEEDED THE GREEN ZONE/POSITION 3. AFTER THE COIL DID NOT DETACH AT THE SYRINGE GREEN ZONE, IT WAS ATTEMPTED TO DEPLOY THE COIL AT THE SYRINGE RED ALTERNATIVE DETACHMENT ZONE. PRIOR TO ATTEMPTING TO DETACH, IT WAS VERIFIED UNDER FLUORO THAT THERE WAS NO STRESS AGAINST THE MICROCATHETER OR DELIVERY TUBE. THE SYRINGE PRESSURIZED AS INTENDED. NECK REMODELING WAS UTILIZED WITH A STENT. THE COIL WAS STILL ATTACHED TO THE COIL DELIVERY SYSTEM WHEN REMOVED FROM THE PATIENT. THERE WERE NO PROCEDURE DELAYS/PROLONGATION DUE TO THE EVENT.