ENVOY 670-256-00Recalls & reported problems
Cerenovus, Inc. · ENVOY Guiding Catheter MPC 100 cm 6F (2.0 mm) .070 Inch (1.8 mm) · Class 2
All models of ENVOY (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886704031697), received between Aug 8, 2024 and Feb 12, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Contamination /Decontamination Problem2
- Crack1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 12, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
AS REPORTED BY THE FIELD, BOTH ENVOY 6F, MPC 100CM CATHETERS (67025600, 31027256) HAD PROTRUDING PLASTIC RESIDUES FROM PRODUCTION ON THE THREADED PIECE. THE PROFESSOR DID NOT WANT TO USE THEM DUE TO THE RISK OF EMBOLIZATION. WE NOTICED THIS PROBLEM WHEN UNPACKING, SO THERE IS NO RISK OF INFECTION. NO PATIENT INJURY AS THE DEVICES WERE NOT CLINICALLY USED. ADDITIONAL EVENT INFORMATION RECEIVED ON 12-FEB-2025 INDICATED THAT THE POUCHES WERE SEALED PRIOR TO BEING OPENED. THERE WAS NOT BEEN ANY PACKAGING ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21364576 ↗MalfunctionFeb 12, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
AS REPORTED BY THE FIELD, BOTH ENVOY 6F, MPC 100CM CATHETERS (67025600, 31027256) HAD PROTRUDING PLASTIC RESIDUES FROM PRODUCTION ON THE THREADED PIECE. THE PROFESSOR DID NOT WANT TO USE THEM DUE TO THE RISK OF EMBOLIZATION. WE NOTICED THIS PROBLEM WHEN UNPACKING, SO THERE IS NO RISK OF INFECTION. NO PATIENT INJURY AS THE DEVICES WERE NOT CLINICALLY USED. ADDITIONAL EVENT INFORMATION RECEIVED ON 12-FEB-2025 INDICATED THAT THE POUCHES WERE SEALED PRIOR TO BEING OPENED. THERE WAS NOT BEEN ANY PACKAGING ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21364536 ↗MalfunctionAug 8, 2024Crack; Crack
AS REPORTED BY THE FIELD, DURING TREATMENT OF AN INTERNAL CAROTID ARTERY STENOSIS WITH AN ENVOY 6F, MPC 100CM (67025600, 31258401) AND EMBOLIZATION. IN THE CASE OF THE ENVOY, THERE WAS A SPLINTER, WHICH IS WHY THE FILTER WIRE TWISTED. UNFORTUNATELY, THE FILTER WIRE EZ EMBOLIC PROTECTION SYSTEM COULD NO LONGER BE USED AND HAD TO BE REPLACED. AFTER CHANGING THE ENVOY AND FILTER WIRE, THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS A 10-MINUTE PROCEDURAL DELAY DUE TO THE EVENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. BASED ON THE (B)(6) 2024 INDICATED THAT THE 10 MINUTES PROCEDURAL DELAY WAS NOT CONSIDERED CLINICALLY SIGNIFICANT. IT WAS FURTHER CLARIFIED THAT THE REPORT OF ¿THERE WAS A SPLINTER¿ MEANS THAT THE CATHETER IS BROKEN. THERE WAS NO SEPARATION OF MATERIAL INTO 2 OR MORE PIECES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19939510 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page