ENVOY 670-256-90Recalls & reported problems
Cerenovus, Inc. · ENVOY Guiding Catheter MPC 90 cm 6F (2.0 mm) .070 Inch (1.8 mm) · Class 2
All models of ENVOY (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704031710), received between Feb 28, 2018 and Oct 30, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak1
- Material Frayed1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 30, 2018Fluid/Blood Leak; Fluid/Blood Leak
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE DIAGNOSTIC ANGIOGRAM PROCEDURE, THE ENVOY® GUIDING CATHETER, 6F, 90 CM, MPC LEAKED DIAGNOSTIC CONTRAST. THE REPORTED LEAK WAS NOTED WHEN THE CATHETER WAS BEING FLUSHED. THE LEAK WAS DISCOVERED PRIOR TO THE CATHETER BEING INSERTED INTO THE PATIENT. IT WAS REPORTED THAT THE LOCATION OF THE LEAK WAS AT THE WELD JUNCTION BETWEEN THE PROXIMAL BLUE AND DISTAL YELLOW POLYMER SEGMENT OF THE CATHETER. THE ENVOY GUIDING CATHETER DID NOT APPEAR DAMAGED PRIOR TO THE LEAK BEING NOTICED; NO EXCESSIVE FORCE WAS APPLIED. ANOTHER ENVOY GUIDING CATHETER WAS USED TO CONTINUE AND COMPLETE THE PROCEDURE. THERE WAS NO REPORT OF PROCEDURAL DELAY, NO REPORT OF PATIENT INJURY OR ADVERSE EVENT ASSOCIATED WITH THE REPORTED ISSUE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8020157 ↗MalfunctionFeb 28, 2018Material Frayed; Material Frayed
THE HEALTH CARE PROFESSIONAL REPORTED THAT THE ENVOY 6F, MPC 90CM GUIDING CATHETER (67025690 / 17707474) DID NOT FIT INTO THE CORDIS 6F SHORT SHEATH. THE DEVICE WAS EXAMINED AND IT WAS REPORTED THAT THE DISTAL TIP OF THE GUIDING CATHETER APPEARED A LITTLE FRAYED, THE NOTED ISSUE IS NOT CONDUCIVE TO SMOOTH ACCESS. A NEW ENVOY 6F, MPC 90CM GUIDING CATHETER WAS USED AND THE NEW GUIDING CATHETER WAS REPORTED TO FITTED WELL. THERE WAS NO REPORT OF PATIENT INJURY OR COMPLICATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7304546 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page