CERELINK 826851Recalls & reported problems

Medos International Sàrl · CERELINK ICP SENSOR Metal Skull Bolt Kit For Intraparenchymal Procedure · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10886704075431), received between Dec 2, 2021 and Aug 9, 2024 — classified as 6 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (4 reports).

1 other recall cover devices cleared under the same FDA submission (K173192); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings4
  • Failure to Zero1
  • Insufficient Device Problem Information1
  • Low Readings1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Failure of Implant1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 9, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    8 OF 12 REPORTS (DIFFERENT PATIENTS, SAME FAILURE). OTHER MFG REPORT NUMBERS: 3013886523-2024-00228, 3013886523-2024-00229, 3013886523-2024-00230, 3013886523-2024-00231, 3013886523-2024-00232, 3013886523-2024-00233, 3013886523-2024-00234, 3013886523-2024-00236, 3013886523-2024-00237, 3013886523-2024-00238, 3013886523-2024-00239. CASE 4: A FACILITY REPORTED A DIRECTLINK MODULE (ID 826828) WITH INCONSISTENT VALUES. THE MODULE WAS USED WITH CERELINK SENSOR (ID 826851). THE SENSOR WAS EXPLANTED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19950168 ↗
  • InjuryAug 9, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    4 OF 12 REPORTS (DIFFERENT PATIENTS, SAME FAILURE). OTHER MFG REPORT NUMBERS: 3013886523-2024-00228, 3013886523-2024-00229, 3013886523-2024-00230, 3013886523-2024-00232, 3013886523-2024-00233, 3013886523-2024-00234, 3013886523-2024-00235, 3013886523-2024-00236, 3013886523-2024-00237, 3013886523-2024-00238, 3013886523-2024-00239. CASE 2: A FACILITY REPORTED A DIRECTLINK MODULE (ID 826828) WITH INCONSISTENT VALUES. THE MODULE WAS USED WITH CERELINK SENSOR (ID 826851). THE SENSOR WAS EXPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19950165 ↗
  • InjuryAug 9, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    8 OF 12 REPORTS (DIFFERENT PATIENTS, SAME FAILURE). OTHER MFG REPORT NUMBERS: 3013886523-2024-00228, 3013886523-2024-00229, 3013886523-2024-00230, 3013886523-2024-00231, 3013886523-2024-00232, 3013886523-2024-00233, 3013886523-2024-00234, 3013886523-2024-00235, 3013886523-2024-00236, 3013886523-2024-00238, 3013886523-2024-00239. CASE 5: A FACILITY REPORTED A DIRECTLINK MODULE (ID 826828) WITH INCONSISTENT VALUES. THE MODULE WAS USED WITH CERELINK SENSOR (ID 826851). THE SENSOR WAS EXPLANTED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19950095 ↗
  • MalfunctionJun 8, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THIS IS 1 OF 2 REPORTS LINKED TO MFG REPORT NUMBERS: A FACILITY REPORTED A CERELINK SENSOR (ID 826851) WAS PLACED IN THE PATIENT ON (B)(6) 2022 FOR PSEUDOTUMOR EVALUATION. IN THE EVENING IT WAS REPORTED THAT THE ICPS WERE -3. THE NURSE STATED THAT SHE FELT THERE WAS NO CLINICAL REASON FOR THE NEGATIVE NUMBERS. THE NURSE DID REPORT THAT THE PATIENT WAS SITTING UP IN BED, ON HER PHONE, IN NO DISTRESS. THE PHYSICIAN ASSISTANT ALSO STATED THAT, " IT COULD BE ATTRIBUTED TO THE PATIENT SITTING UP IN BED, BUT DID FEEL IT WAS A LARGE DISCREPANCY." THE CERELINK MONITOR HAD A GOOD AND RESPONSIVE WAVEFORM, HOWEVER, THERE WAS A CONCERN THAT THE INFORMATION THE MONITOR WAS GIVING THE PROVIDERS WASN'T VALID. IT IS UNKNOWN IF THE SENSOR WAS REMOVED OR IF THE PATIENT WAS STILL MONITORED. CLINICAL QUESTIONNAIRE PROVIDED BY THE CUSTOMER: TIME/DATE OF FAILURE: (B)(6) 2022, REPORTED AT 1919. DESCRIPTION OF THE FAILURE: "PT HAD BOLT PLACED IN OR IN THE LATE AFTERNOON. THE MONITOR WAS ON BATTERY POWER UNTIL THE PATIENT WENT TO THE ICU WHEN IT WAS PLUGGED IN AND CONNECTED TO THE PATIENT MONITOR. ICP AT PLACEMENT WAS 12. NIGHT RN CALLED CLINICAL SPECIALIST AT 1919 REGARDING ICP BEING -3. NIGHT RN DID NOT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14634281 ↗
  • InjuryMay 10, 2022Failure to Zero; Failure to Zero

    THIS IS 2 OF 3 REPORTS LINKED TO MFG REPORT NUMBER: 3013886523-2021-00537, 3013886523-2021-00538. A FACILITY REPORTED A PATIENT HAD A CERELINK SENSOR PLACED WITH GOOD WAVEFORM AND ICP VALUE. 25 TO 30 MINUTES LATER, THE SURGEON RECEIVED A CALL THAT THE ICP READING WAS -32. THEY CHANGED THE CABLES AND SHOWED A MESSAGE OF FAILED SENSOR. THE MONITOR READ OCCASIONALLY THEN WORKED AGAIN. IT STOPPED WORKING AGAIN AT 11PM. THE SURGEON DECIDED TO WAIT UNTIL MORNING BUT STILL DID NOT WORK. THEY TOOK THE PATIENT BACK TO THE OPERATIVE ROOM TO HAVE IT REPLACED. A SECOND SENSOR WOULD NOT ZERO WITH THE MONITOR WHICH STATED FAILED SENSOR AND IT WAS REPLACED WITH ANOTHER SENSOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14346306 ↗
  • MalfunctionApr 12, 2022Output Problem; Output Problem

    WHILE DEMONSTRATING HOW TO USE THE CERELINK MONITOR AN ERROR MESSAGE APPEARED SAYING THAT THE SYSTEM DETECTED A PROBLEM EITHER WITH THE EXTENSION CABLE OR SENSOR. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14085477 ↗
  • InjuryDec 2, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    A FACILITY REPORTED THE MICROSENSOR REGISTERED SENSOR FAILURE. THE SENSOR WAS EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 12920774 ↗
  • InjuryDec 2, 2021Low Readings; Low Readings

    A FACILITY REPORTED A PATIENT HAD A BOLT PLACED WITH GOOD WAVEFORM AND ICP VALUE. 25-30 MINUTES LATER, THE SURGEON RECEIVED A CALL THAT THE ICP READING WAS -32. THEY CHANGED THE CABLES AND SHOWED A MESSAGE OF FAILED SENSOR. THE MONITOR READ OCCASIONALLY THEN WORKED AGAIN. IT STOPPED WORKING AGAIN AT 11PM. THE SURGEON DECIDED TO WAIT UNTIL MORNING BUT STILL DID NOT WORK. THEY TOOK THE PATIENT BACK TO THE OR TO HAVE IT REPLACED. A SECOND SENSOR WOULD NOT ZERO WITH THE MONITOR WHICH STATED FAILED SENSOR AND IT WAS REPLACED WITH ANOTHER SENSOR.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12918545 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K173192. One submission often covers many models, so check the product and lot details in each record.

  • Z-1037-2026 · Dec 12, 2025 · Open, Classified
    Codman CereLink ICP Sensor Basic Kit. Catalog Number: 826850. The Codman CereLink ICP Sensor Basic Kit consists of the CereLink ICP Sensor (ICP Sensor) and a 14-gauge Tuohy needl

Other DEVICE, MONITORING, INTRACRANIAL PRESSURE devices

All DEVICE, MONITORING, INTRACRANIAL PRESSURE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page