Skip to content

CERELINK 826852Recalls & reported problems

Medos International Sàrl · CERELINK ICP SENSOR Plastic Skull Bolt Kit For Intraparenchymal Procedure · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704075455), received between Apr 21, 2022 and Jul 19, 2022 — classified as 2 injuries.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).

1 other recall cover devices cleared under the same FDA submission (K173192); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Low Readings1
  • No Device Output1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 19, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A FACILITY REPORTED A CERELINK MICROSENSOR WAS IMPLANTED AND CONNECTED TO A CERELINK MONITOR. THE CERELINK MONITOR STARTED DISPLAYING NEGATIVE VALUES THAT WERE NOT COMPATIBLE WITH THE PATIENT STATUS (THE PATIENT WAS FINE). THE VALUE WITH CERELINK MONITOR WAS -9. DUE TO THE NOT RELIABLE VALUES, THEY DECIDED TO MEASURE ICP WITH DVE AND WITH DVE THE ICP VALUE WAS 6.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15059148 ↗
  • InjuryApr 21, 2022No Device Output; Low Readings; No Device Output; Low Readings

    A FACILITY REPORTED A CERELINK MICROSENSOR WAS INSERTED WITH ICP EXPRESS AND AFTER A FEW MINUTES WITH POSITIVE ICP VALUES, THE ICP WENT DOWN TO -50 AND STOP MONITORING. EVEN SWITCHING ICP EXPRESS, AND REPLACING THE ZERO, VALUES WERE NOT DISPLAYED. THE SENSOR WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14167035 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K173192. One submission often covers many models, so check the product and lot details in each record.

  • Z-1037-2026 · Dec 12, 2025 · Open, Classified
    Codman CereLink ICP Sensor Basic Kit. Catalog Number: 826850. The Codman CereLink ICP Sensor Basic Kit consists of the CereLink ICP Sensor (ICP Sensor) and a 14-gauge Tuohy needl

Other DEVICE, MONITORING, INTRACRANIAL PRESSURE devices

All DEVICE, MONITORING, INTRACRANIAL PRESSURE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page