Skip to content

DELTAFILL DLF100410Recalls & reported problems

Cerenovus, Inc. · DELTAFILL Microcoil 10 Stretch Resistant Coil and Detachable Coil System Contents: 1 Detachable Coil System Coil Type: DELTAFILL 10 Microcoil 4mm (coil diameter) x 10cm (coil length) 190cm (working length) · Class 2 · Implant

All models of DELTAFILL (29)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704077053), received between Apr 4, 2024 and Oct 24, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Failure to Advance” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Advance2
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 24, 2024Failure to Advance; Premature Separation; Failure to Advance; Premature Separation

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR COIL EMBOLIZATION PROCEDURE TARGETING A LEFT POSTERIOR COMMUNICATING ARTERY (PCOM) ANEURYSM WITH NORMAL ANATOMY USING A CEREBASE (UNKNOWN CATALOG AND LOT NUMBER), A SOFIA¿ 6F INTERMEDIATE CATHETER (MICROVENTION) AND A TOTAL OF SIX COILS. IT WAS REPORTED THAT THREE OF THE SIX COILS ¿GOT STUCK IN THE MICROCATHETER AND COULD NOT BE REMOVED.¿ THE THREE COILS WERE: A 6MM X 25CM DELTAFILL 18 (DLF180625 / 30719513), A 5MM X 20CM DELTAFILL 18 (DLF180520 / 30942962), AND A 4MM X 10CM DELTAFILL 10 (DLF100410 / 30871956). IT WAS REPORTED THAT ¿WHEN TRYING TO PULL THE FIRST COIL [THE 6MM X 25CM DELTAFILL 18 (DLF180625 / 30719513)] OUT OF THE CONCOMITANT SL-10® MICROCATHETER (STRYKER), THE PUSHER WIRE TORE OFF AND THE COIL BECAME STUCK IN THE MICROCATHETER. THE MICROCATHETER THEREFORE, HAD TO BE REMOVED. IT WAS REPORTED THAT PRIOR TO THIS COIL, THE PREVIOUS THREE COILS ¿COULD BE PUSHED THROUGH THE SAME [MICROCATHETER].¿ THIS FIRST SL-10® MICROCATHETER (STRYKER) WAS NOT SAVED. THE SAME INCIDENT OCCURRED TWICE MORE DURING COIL EMBOLIZATION WITH THE 5MM X 20CM DELTAFILL 18 (DLF180520 / 30942962), AND A 4MM X 10CM DELTAFILL 10 (DLF100

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20526024 ↗
  • MalfunctionApr 4, 2024Failure to Advance; Failure to Advance

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE PRE-OP PREPARATION FOR AN ENDOVASCULAR EMBOLIZATION PROCEDURE, THE COMPLAINT DEVICE, A 4MM X 10CM DELTAFILL 10 (DLF100410 / 30784433) COULD NOT BE PUSHED OUT OF THE INTRODUCER SHEATH. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. ON 21-FEB-2024, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE COIL HAS NOT YET BEEN INTRODUCED INTO THE MICROCATHETER. NO DAMAGE WAS VISIBLE ON THE DEVICE. NOTHING WAS NOTED TO BE OBSTRUCTING THE INTRODUCER. THE REPORTED ISSUE OCCURRED WHEN THE COIL WAS BEING INSPECTED IN SALINE; THE ISSUE OCCURRED BEFORE CONNECTING TO THE MICROCATHETER. EXCESSIVE FORCE HAD NOT BEEN EXERTED ON THE DEVICE. THE 4MM X 10CM DELTAFILL 10 WAS RETURNED FOR EVALUATION AND ANALYSIS. BASED ON THE RESULT OF THE PRODUCT ANALYSIS COMPLETED ON 04-APR-2024, IN WHICH THE EMBOLIC COIL COMPONENT WAS OBSERVED TO BE STRETCHED AT THE PROXIMAL SECTION, THE REPORTED ISSUE MEETS USFDA REPORTING CRITERIA UNDER 21 CFR 803 AS A "MALFUNCTION."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19044790 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page