DELTAFILL DLF100612Recalls & reported problems
Cerenovus, Inc. · DELTAFILL Microcoil 10 Stretch Resistant Coil and Detachable Coil System Contents: 1 Detachable Coil System Coil Type: DELTAFILL 10 Stretch Resistant Coil 6mm (coil diameter) x 12cm (coil length) 190cm (working length) · Class 2 · Implant
All models of DELTAFILL (29)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704077091), received between Dec 11, 2017 and Mar 1, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Rupture1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 1, 2022Material Rupture; Material Rupture
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION, A DELTAFILL 10 6MM X 12CM COIL (DLF100612, 30474387) WAS CHECKED AS USUAL, AND NOTHING CONSPICUOUS WAS FOUND. WHEN TRYING TO PULL THE COIL BACK INTO THE SYSTEM, A DEFECT WAS FOUND IN THE RE-SHEETING SYSTEM. SINCE THE WIRE OF THE COIL WAS NOW OUTSIDE THE SYSTEM EXTRALUMINALLY, IT WAS NO LONGER POSSIBLE TO PULL THE COIL BACK INTO THE SYSTEM AND THUS TO USE IT. THERE WAS SURGERY DELAY BY 10 MINUTES DUE TO THE REPORTED EVENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. ADDITIONAL INFORMATION RECEIVED INDICATED THAT THERE WAS NO DIFFICULTY REMOVING THE DEVICE FROM THE PACKAGING. THERE WAS NO EXCESSIVE FORCE APPLIED. BASED ON THE ANALYSIS OF THE DEVICE RECEIVED, THE EMBOLIC COIL WAS FOUND TO BE STRETCHED AND KINKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13636992 ↗MalfunctionDec 11, 2017Stretched; Stretched
IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL THAT DURING A PROCEDURE THREE DELTAFILL10 (DLF100925/S12152 X2 & DLF100612/UNKNOWN LOT) WERE ATTEMPTED TO BE INTRODUCED THROUGH THE PROWLER SELECT LP MICROCATHETER (606-S155MX/17536302) BUT COULD NOT PASS THROUGH DESPITE BEING FLUSHED PRIOR TO USE. COMPETITOR PRODUCTS WERE FIRST USED WITH THE PROWLER SELECT LP MICROCATHETER, FOLLOWED BY SOME DELTAFILL18 COILS, AND NO ISSUES WERE EXPERIENCED. A TOTAL OF 8 COILS WERE IMPLANTED PRIOR TO THE COMPLAINT COILS. THE PHYSICIAN EXPERIENCED FRICTION, SIMILAR TO THE FRICTION EXPERIENCED WITH THE PREVIOUS GENERATION PRODUCT. THE PHYSICIAN FURTHER STATED THAT THE SHEATH IS MORE FRAGILE THAN THE ONE ON THE PREVIOUS GENERATION COILS AND PEELS MORE. THE PHYSICIAN REMOVED THE DELTAFILL10 COILS (COMPLAINT PRODUCTS) AND COMPLETED THE PROCEDURE WITH 18MM COILS. THE FIRST 9MM COIL WAS IMPOSSIBLE TO ADVANCE THROUGH THE HUB OF THE MICROCATHETER; SEVERE FRICTION WAS FELT. THE PHYSICIAN TRIED A SECOND ONE OF THE SAME DIAMETER AND THEN A THIRD COIL THAT WAS 6MM IN DIAMETER; ALL HAD THE SAME ISSUE. THE PHYSICIAN REPORTED THAT THE COILS COULD NOT BE ADVANCED MORE THAN A FEW CENTIMETERS IN THE MICROCATHETER. THERE WE
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7105566 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page