DELTAXSFT DLX100151Recalls & reported problems
Cerenovus, Inc. · DELTAXSFT Microcoil 10 Stretch Resistant Coil and Detachable Coil System Contents: 1 Detachable Coil System Coil Type: DELTAXSFT 10 Microcoil 1.5mm (coil diameter) x 1cm (coil length) 190cm (working length) · Class 2 · Implant
All models of DELTAXSFT (13)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886704077244), received between Feb 21, 2020 and Sep 22, 2021 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Separation Failure3
- Positioning Problem2
- Malposition of Device1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2021Separation Failure; Separation Failure
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION, A DELTAXSFT10 1.5MM X 1CM COIL (DLX100151, K10326) FAILED TO DETACH. WITH THIS PHENOMENON, IT WAS RETRACTED FULLY AND REPLACED WITH ANOTHER COIL. THERE WAS NO SURGERY DELAYED DUE TO THE REPORTED EVENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO FRAGMENTS WERE GENERATED. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12514945 ↗MalfunctionJan 13, 2021Separation Failure; Separation Failure
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION, A 1.5MM X 1CM DELTAXSFT10 COIL (DLX100151, L11766) FAILED TO DETACH AFTER DEPLOYING THE COIL INTO THE ANEURYSM. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11163118 ↗MalfunctionJan 12, 2021Separation Failure; Separation Failure
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION, A 1.5MM X 1CM DELTAXSFT10 COIL (DLX100151, L11766) FAILED TO DETACH AFTER DEPLOYING THE COIL INTO THE ANEURYSM. THERE WAS NO PATIENT INJURY REPORTED. NO ADDITIONAL INFORMATION WAS AVAILABLE AT THE TIME OF REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11156239 ↗MalfunctionSep 28, 2020Stretched; Positioning Problem; Stretched; Positioning Problem
THE HEALTHCARE PROFESSIONAL REPORTED THAT THE 1.5MM X 1CM DELTAXSFT 10 COIL (DLX100151 / L15456) WAS THE FIRST COIL CHOSEN FOR THE PROCEDURE; IT WAS REPORTED THAT THE PHYSICIAN DID NOT LIKE THE SHAPE OF THE COIL DEPLOYMENT AND DECIDED TO REMOVE THE COIL. DURING THE RESHEATHING ATTEMPT, THE COIL INTRODUCER FAILED TO BE RE-ZIPPED. THE PHYSICIAN USED ANOTHER 1.5MM X 1CM DELTAXSFT 10 COIL COIL AS REPLACEMENT AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO REPORT OF ANY PATIENT ADVERSE EVENT OR COMPLICATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10596960 ↗MalfunctionFeb 24, 2020Malposition of Device; Malposition of Device
AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING A COIL EMBOLIZATION, A 1.5MM X 1CM DELTAXSFT10 COIL (DELTAXSFT, L16085) WAS BEING USED AS THE FINAL FINISHING COIL. AFTER THE COIL DETACHED AND WAS REMOVED FROM THE EXCELSIOR MICROCATHETER (MC), THERE WAS ¿SOMETHING STICKING OUT¿ AT THE END OF THE MC. THE PROCEDURE WAS WELL COMPLETED. THERE WAS NO PATIENT INJURY REPORTED. THE DEVICE WAS REMOVED FROM THE PATIENT WITHOUT ADDITIONAL INTERVENTION. THE DEVICE(S) WERE USED AND PREPPED AS PER THE INSTRUCTIONS FOR USE. THE CONCOMITANT DEVICES FUNCTIONED AS EXPECTED. THERE WAS NO DEVICE FRAGMENTS THAT REMAINED IN THE PATIENT. THERE WAS NO EXCESSIVE FORCE APPLIED TO THE DEVICE. THERE WAS NO DIFFICULTY WITH DEVICE POSITIONING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9745203 ↗MalfunctionFeb 21, 2020Positioning Problem; Positioning Problem
AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING A COIL EMBOLIZATION, A 1.5MM X 1CM DELTAXSFT10 COIL (DLX100151, L14830) WAS BEING USED AS THE 5TH COIL, HOWEVER, THE PHYSICIAN DIDN'T LIKE THE SHAPE OF THE COIL DEPLOYMENT AND DECIDED TO PULL IT OUT. DURING RESHEATHING OF THE COIL, THE INTRODUCER FAILED TO BE RE-ZIPPED. THE PHYSICIAN USED A SAME LIKE PRODUCT. THE PROCEDURE WAS SUCCESSFULLY FINISHED. THERE WAS NO REPORT OF PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9738526 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page