Skip to content

DELTAXSFT DLX100304Recalls & reported problems

Cerenovus, Inc. · DELTAXSFT Microcoil 10 Stretch Resistant Coil and Detachable Coil System Contents: 1 Detachable Coil System Coil Type: DELTAXSFT 10 Microcoil 3mm (coil diameter) x 4cm (coil length) 190cm (working length) · Class 2 · Implant

All models of DELTAXSFT (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886704077343), received between Jun 6, 2019 and Jun 6, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Separation Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Separation Failure1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 6, 2019Separation Failure; Separation Failure

    AS REPORTED BY AN AFFILIATE, DURING A COIL EMBOLIZATION THERE WAS DIFFICULTIES DETACHING A DELTAXSFT10 3MM X 4CM COIL (DLX100304, L12179), REQUIRING ADDITIONAL MANIPULATION. THE COIL DID FINALLY DETACH. THERE WERE NO CONSEQUENCES TO THE PATIENT. THE EVENT DID NOT RESULT IN A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. PRE-DEPLOYMENT ELECTRICAL TESTING WAS PERFORMED BEFORE USE AND THE RESULTS WERE OK. DURING THE DETACHMENT CYCLE, THE DETACHMENT LIGHT DID ILLUMINATE AND THERE WAS AUDIBLE SIGNAL BEEP. ALL CONNECTIONS APPEAR TO FIT PROPERLY WITHOUT APPLICATION OF EXCESSIVE FORCE. ALTHOUGH THERE WERE NO ALLEGED PRODUCT MALFUNCTIONS WITH THE ENPOWER DETACHABLE CONTROL BOX (DCB), THE DCB WAS NOT USED WITH SUBSEQUENT COILS. THE TARGET SITE WAS A CEREBRAL RIGHT MIDDLE ANEURYSM. NECK REMODELING WAS UTILIZED WITH A NON-CERENOVUS STENT. THE DCB WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8674808 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices