Skip to content

GALAXY G3 GLY121030Recalls & reported problems

Cerenovus, Inc. · GALAXY G3 Microcoil 12 Stretch Resistant Coil and Detachable Coil System Contents: 1 Detachable Coil System Coil Type: GALAXY G3 12 Stretch Resistant Coil 10mm (coil diameter) x 30cm (coil length) 190cm (working length) · Class 2 · Implant

All models of GALAXY G3 (58)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886704077381), received between Apr 1, 2020 and Oct 6, 2023 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Deformation Due to Compressive Stress1
  • Stretched1

Patient problems

  • Aneurysm1
  • No Clinical Signs, Symptoms or Conditions1
  • No Code Available1
  • Thromboembolism1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 6, 2023Stretched; Stretched

    THE HEALTHCARE PROFESSIONAL REPORTED THAT PRIOR TO THE START OF THE PROCEDURE, THE COMPLAINT DEVICE, A 10MM X 30CM GALAXY G3 COIL (GLY121030/30386790) WAS STRETCHED. THE DEVICE WAS NOT USED IN THE PATIENT. IT WAS DISCARDED. THERE WAS NO NEGATIVE PATIENT IMPACT. ON 27-SEP-2023, ADDITIONAL INFORMATION WAS RECEIVED. PER THE ADDITIONAL INFORMATION, THE PHYSICIAN DOES NOT REMEMBER THE EXACT DATE OF THE PROCEDURE. WHEN THE PRODUCT WAS REMOVED FROM THE RESPECTIVE PACKAGING, THE ISSUE WAS IMMEDIATELY VISIBLE, THE COIL WAS STRETCHED. THERE WAS NO DAMAGES NOTED ON THE PACKAGING OF THE DEVICE. ASIDE FROM THE STRETCHED CONDITION NOTED, NO OTHER ANOMALY WAS OBSERVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17892172 ↗
  • InjuryOct 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE EVENT WAS REPORTED VIA THE STERLING STUDY, THE 64-YEAR-OLD FEMALE PATIENT WITH A HISTORY OF CONTROLLED HYPERLIPIDEMIA, CONTROLLED HYPERTENSION, SMOKING (CURRENT), AND INTRACRANIAL ANEURYSMS (TOTAL OF 3) PRESENTED WITH A SUBARACHNOID HEMORRHAGE (SAH) SECONDARY TO RUPTURED ANEURYSM (HUNT & HESS GRADE 2) AND SUBSEQUENTLY UNDERWENT BALLOON-ASSISTED COIL EMBOLIZATION ON (B)(6) 2020. IMMEDIATE PRE-PROCEDURE ANGIOGRAPHY REVEALED A RUPTURED ANEURYSM ON THE LEFT SIDEWALL OF THE INTERNAL CAROTID ARTERY (ICA): HEIGHT: 10MM, DOME: 12MM, MAXIMUM ANEURYSM DIAMETER: 17MM, NECK SIZE: 7.8MM, AND DOME-TO-NECK RATIO: 1.5MM. THE PARENT VESSEL DIAMETER WAS 5.8MM. PLATELET REACTIVITY TESTING WAS NOT PERFORMED PRE-PROCEDURE. BALLOON-ASSISTED COIL EMBOLIZATION WAS PERFORMED WITH THE IMPLANTATION OF THE FOLLOWING COILS: A 14MM X 47CM MICRUSFRAME18 COIL (MFR181447 / L10314), A 11MM X 37CM MICRUSFRAME18 COIL (MFR181137 / L10312), A 10MM X 30CM GALAXY G3 COIL (GLY121030 / L11083), AND A 9MM X 25CM GALAXY G3 COIL (GLY120925 / L10732) VIA AN EXCELSIOR® SL-10® MICROCATHETER (STRYKER). AN 8MM X 24CM GALAXY G3 COIL (GLY120824 / L15747) WAS USED BUT NOT IMPLANTED. HEPARIN WAS ADMINISTERED DURING THE PROCEDURE.

    Patient problems: Aneurysm

    MAUDE report 17857130 ↗
  • InjuryDec 27, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE EVENT WAS REPORTED VIA THE (B)(6) STUDY, A (B)(6) YEAR OLD FEMALE PATIENT WITH A HISTORY OF CONGESTIVE HEART FAILURE (CHF), DIABETES, CONTROLLED HYPERTENSION, HYPOTHYROIDISM, GASTROESOPHAGEAL REFLUX DISEASE (GERD), SLEEP APNEA, LIVER CIRRHOSIS, ESOPHAGEAL VARICES, OSTEOPOROSIS, IDIOPATHIC PULMONARY FIBROSIS (IPF), CHRONIC OBSTRUCTIVE PULMONARY DISEASE (COPD), AND CLASS 1 OBESITY (BMI 30 KG/M2) UNDERWENT STENT-ASSISTED COIL EMBOLIZATION OF AN UNRUPTURED ANTERIOR COMMUNICATING ARTERY (ACOM) ANEURYSM ON (B)(6) 2021 AND EXPERIENCED A THROMBOEMBOLIC EVENT DURING THE PROCEDURE. MEDICATION WAS ADMINISTERED AS TREATMENT. THE PATIENT WAS ASYMPTOMATIC; THE THROMBOEMBOLIC EVENT WAS ONLY OBSERVED ANGIOGRAPHICALLY. THE PATIENT WAS DISCHARGED HOME WITH SELF-CARE ON THE FOLLOWING DAY WITH MODIFIED RANKIN SCALE (MRS) SCORE OF 2, UNCHANGED FROM BASELINE. PLATELET REACTIVITY TESTING WAS NOT PERFORMED. THE DIMENSIONS OF THE ACOM BIFURCATION ANEURYSM WERE AS FOLLOWS: ANEURYSM HEIGHT 9.7MM, DOME 8.4MM, MAXIMUM ANEURYSM DIAMETER 6.6MM, NECK SIZE 3.6MM, AND DOME-TO-NECK RATIO OF 2.3MM. THE PARENT VESSEL DIAMETER WAS 3.7MM. STENT-ASSISTED COIL EMBOLIZATION WAS PERFORMED WITH SIX CERENOVUS COILS: A 10M

    Patient problems: Thromboembolism

    MAUDE report 13090773 ↗
  • InjuryApr 1, 2020Deformation Due to Compressive Stress; Adverse Event Without Identified Device or Use Problem; Deformation Due to Compressive Stress; Adverse Event Without Identified Device or Use Problem

    AS REPORTED BY THE (B)(4) STUDY, A (B)(6)-YEAR OLD FEMALE (SUBJECT (B)(6)) WITH A HISTORY OF CARDIAC ARRHYTHMIA, DIABETES, CONTROLLED HYPERLIPIDEMIA, AND CONTROLLED HYPERTENSION, UNDERWENT COIL EMBOLIZATION OF A RUPTURED POSTERIOR COMMUNICATING ARTERY ANEURYSM ON (B)(6) 2018 AND EXPERIENCED ANEURYSM RECANALIZATION ON (B)(6) 2019 REQUIRING RETREATMENT ON (B)(6) 2019. IMMEDIATE PRE-PROCEDURE ANGIOGRAPHY ON (B)(6) 2018 REVEALED AN ANTERIOR CIRCULATION ANEURYSM ON RIGHT SIDEWALL OF THE POSTERIOR COMMUNICATING ARTERY. THE RUPTURED ANEURYSM HAD THE FOLLOWING DIMENSIONS: HEIGHT 11.3MM, DOME 11.3MM, MAXIMUM ANEURYSM DIAMETER 11.9MM, NECK SIZE 4.4MM, AND DOME-TO-NECK RATIO 2.6MM. PARENT VESSEL DIAMETER WAS 3.9MM. COIL EMBOLIZATION WAS SUBSEQUENTLY PERFORMED WITH THE IMPLANTATION OF ONE 9MM X 14CM MICRUSFRAME C (S11389), ONE 10MM X 30CM GALAXY G3 (L10166), ONE 9MM X 25CM GALAXY G3 (L10165), TWO 5MM X 15CM GALAXY G3 (L12691), TWO 4MM X 10CM GALAXY G3 (L12837), AND ONE 2MM X 6CM GALAXY G3 MINI (L12517) VIA A SL-10 (STRYKER) MICROCATHETER. TWO COILS FAILED TO UNSHEATHE PROPERLY (ONE 2MM X 6CM GALAXY G3 MINI - L12517 AND ONE 10MM X 16CM MICRUSFRAME14 ¿ S11391) AND WERE THEREFORE NOT IMPLANTED. H

    Patient problems: No Code Available

    MAUDE report 9911167 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →