Skip to content

GALAXY G3 GLY120202Recalls & reported problems

Cerenovus, Inc. · GALAXY G3 Microcoil 12 Stretch Resistant Coil and Detachable Coil System Contents: 1 Detachable Coil System Coil Type: GALAXY G3 12 Stretch Resistant Coil 2mm (coil diameter) x 2cm (coil length) 190cm (working length) · Class 2 · Implant

All models of GALAXY G3 (58)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886704077428), received between Jan 27, 2021 and Feb 20, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Separation Failure” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Separation Failure2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Aneurysm1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 20, 2026Separation Failure

    DURING A STENT-ASSISTED ENDOVASCULAR EMBOLIZATION PROCEDURE ON (B)(6) 2026, THE HEALTHCARE PROFESSIONAL REPORTED THAT THE 4MM X 16MM ENTERPRISE 2 VASCULAR RECONSTRUCTION DEVICE (VRD) (ENCR401600 / 9540635) WAS IMPEDED IN THE INTRODUCER AND COULD NOT BE PUSHED INTO THE CONCOMITANT MICROCATHETER (UNSPECIFIED BRAND). THE PHYSICIAN RETRACTED ONLY THE STENT AND REPLACED IT TO CONTINUE THE PROCEDURE. ANOTHER MICROCATHETER (UNSPECIFIED BRAND) WAS IN PLACE AND A 2MM X 2CM GALAXY G3 (GLY120202 / 1042546S) WAS DELIVERED TO THE TARGET SITE. THE COIL FILLED THE ANEURYSM AND WHEN THE PHYSICIAN TRIED TO DETACH THE COIL, THE COIL WAS UNABLE TO DETACH. THE PHYSICIAN RETRACTED THE COIL AND REPLACED IT WITH NEW COILS, A SECOND AND THIRD COIL SUCCESSIVELY, BOTH WERE 1.00MM X 1.00CM GALAXY G3 MINI (GLM910010) FROM THE SAME LOT (1112512S). IT WAS REPORTED THAT BOTH OF THESE COILS WERE ALSO UNABLE TO DETACH. THE PHYSICIAN RETRACTED THE SECOND COIL AND THE THIRD COIL IN SUCCESSION AND REPLACED IT WITH ANOTHER COIL FROM A DIFFERENT MANUFACTURE AND COMPLETE THE PROCEDURE. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. THE PROCEDURE WAS REPORTEDLY PROLONGED BY ABOUT 10 MINUTES. ON 12-FEB-2026, ADDITION

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24403203 ↗
  • MalfunctionJul 14, 2025Separation Failure; Separation Failure

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, THE PHYSICIAN PUT THE MICROCATHETER (UNSPECIFIED COMPETITOR BRAND) IN PLACE AND THE FIRST COIL, A 3MM X 6CM GALAXY G3 (GLY120306 / 3542402S) WAS DELIVERED TO THE TARGET SITE. THE COIL FILLED IN THE ANEURYSM AND THE PHYSICIAN ATTEMPTED TO DETACH THE COIL, BUT THE COIL WAS UNABLE TO BE DETACHED. THE PHYSICIAN REMOVED THE COIL AND THE MICROCATHETER FROM THE PATIENT AND REPLACED THE COIL WITH A 2MM X 2CM GALAXY G3 (GLY120202 / 0512508S) TO BE DELIVERED WITH THE SAME MICROCATHETER AND THE SECOND COIL HAD THE SAME ISSUE. THE PHYSICIAN REMOVED THE SECOND COIL AND THE MICROCATHETER FROM THE PATIENT AND REPLACED IT WITH A THIRD COIL AND SECOND MICROCATHETER (UNSPECIFIED COMPETITOR BRAND) TO COMPLETE THE PROCEDURE, WHICH WAS PROLONGED BY APPROXIMATELY 10 MINUTES. THERE WAS NO NEGATIVE IMPACT ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22503261 ↗
  • InjuryJan 27, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED VIA THE (B)(6) STUDY, A (B)(6) YEAR OLD FEMALE (SUBJECT (B)(6)) WITH A HISTORY OF BEING A CURRENT SMOKER UNDERWENT COIL EMBOLIZATION OF A LEFT SIDE WALL POSTERIOR COMMUNICATING ARTERY (PCA) (ANTERIOR CIRCULATION) ANEURYSM ON (B)(6) 2019. IMMEDIATE PRE-PROCEDURE ANGIOGRAPHY REVEALED A RUPTURED LEFT PCA ANEURYSM WITH THE FOLLOWING DIMENSIONS: HEIGHT 5.1MM, DOME 3.3MM, MAXIMUM ANEURYSM DIAMETER 5.1MM, NECK SIZE 2.0MM, AND DOME-TO-NECK RATIO 1.7MM. THE PARENT VESSEL DIAMETER WAS 3.2MM. COIL EMBOLIZATION WAS SUBSEQUENTLY PERFORMED WITH THE IMPLANTATION OF ONE 4MM X 6CM GALAXY G3 XSFT (GLX120406/L16293), ONE 2MM X 2CM GALAXY G3 (GLY120202/L12097), ONE 3MM X 8CM GALAXY G3 MINI (GLM930080/L12616), ONE 1.5MM X 2CM GALAXY G3 MINI (GLM915020/L11355), AND ONE 1MM X 1CM GALAXY G3 MINI (GLM910010/L14466) VIA AN EXCELSIOR SL-10 (STRYKER) MICROCATHETER. THE STUDY COILS WERE SUCCESSFULLY IMPLANTED AT THE TARGET SITE WITH 24.09% ANGIOSUITE PACKING DENSITY. ACCORDING TO THE CASE REPORT FORM (CRF), THERE WAS NO REPORTED MICROCATHETER KICKBACK. IN THE OPINION OF THE INVESTIGATOR, TREATMENT OF THE TARGET ANEURYSM WAS CONSIDERED COMPLETE WITH MODIFIED RAYMOND-ROY SCORE OF CLASS I: COMPLETE =

    Patient problems: Aneurysm

    MAUDE report 11237031 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page