GALAXY G3 GLY123575Recalls & reported problems
Cerenovus, Inc. · GALAXY G3 Microcoil 12 Stretch Resistant Coil and Detachable Coil System Contents: 1 Detachable Coil System Coil Type: GALAXY G3 12 Stretch Resistant Coil 3.5mm (coil diameter) x 7.5cm (coil length) 190cm (working length) · Class 2 · Implant
All models of GALAXY G3 (58)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886704077435), received between Oct 5, 2017 and Mar 28, 2023 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Difficult to Advance” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Difficult to Advance1
- Separation Failure1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Infarction, Cerebral1
- No Consequences Or Impact To Patient1
- Thromboembolism1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 28, 2023Separation Failure; Separation Failure
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION AT THE CEREBRAL ANEURYSM. A GALAXY G3 3.5MM X 7.5CM COIL (GLY123575, 30508877) WAS CONNECTED TO AN ENPOWER CONTROL CABLE (ECB00018200, 30960241). BUT, THE INDICATOR LED OF CDB2 DID NOT ILLUMINATE. THE CABLE WAS REPLACED WITH ANOTHER CABLE, BUT THE SAME ISSUE CONTINUED. THE COMPLAINT COIL WAS REPLACED WITH ANOTHER COIL AND THE INDICATOR LED WAS ILLUMINATED. THE COMPLAINT COIL WAS DELIVERED, AND THE DETACHMENT CYCLE WAS PERFORMED. THE PROCEDURE WAS COMPLETED. THERE WAS NO NEGATIVE IMPACT ON THE PATIENT. A CONTINUOUS FLUSH WAS DONE. THE LESION WAS THE INTERNAL CAROTID ARTERY. OTHER CONCOMITANT DEVICE IS A MICROCATHETER (HEADWAY).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16626604 ↗InjuryApr 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE HEALTHCARE PROFESSIONAL REPORTED THAT A 58-YEAR-OLD FEMALE UNDERWENT PULSERIDER ANEURYSM NECK RECONSTRUCTION DEVICE (ANRD)-ASSISTED COIL EMBOLIZATION OF A RECURRENT RIGHT MIDDLE CEREBRAL ARTERY (MCA) ANEURYSM AND EXPERIENCED INTRAOPERATIVE THROMBUS FORMATION AT THE ANEURYSM NECK. THE THROMBUS WAS NOTED TO BE PROTRUDING INTO THE PARENT VESSEL. USING A SCEPTER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON (MICROVENTION), THE COIL MASS WAS PUSHED UP TO THE ANEURYSM SIDE, AND THE DELIVERY TO THE ANEURYSM NECK WAS IMPROVED. THE THROMBUS ALSO DECREASED, AND BLOOD FLOW IMPROVED, SO ANTIPLATELET DRUGS WERE PRESCRIBED. THE PATIENT'S CLINICAL COURSE WAS MONITORED. POST-PROCEDURE, THERE WERE NO REMARKABLE FINDINGS. IT WAS LAST REPORTED THAT THE PATIENT IS HOSPITALIZED. THE ATTENDING PHYSICIAN'S OPINION ON THE RELATIONSHIP BETWEEN THIS EVENT AND THE PULSERIDER ANRD IS AS FOLLOWS: ¿IT WOULD HAVE BEEN BETTER IF PULSERIDER HAD BEEN PUSHED MORE FIRMLY INTO THE MOTHER'S VEIN; BECAUSE IT WAS SLIGHTLY SWEET, THE COIL WAS CLOGGED UP TO THE VERY END OF THE ANEURYSM NECK, AND AS A RESULT, PULSERIDER DOWN IN THE MOTHER'S VEIN. COMPARED TO THE STENT, THE COIL OF PULSERIDER WAS INSERTED MORE AND
Patient problems: Thromboembolism
MAUDE report 14048402 ↗MalfunctionJun 23, 2021Difficult to Advance; Difficult to Advance
THE EVENT WAS REPORTED VIA THE (B)(6) STUDY, THE (B)(6) YEAR-OLD FEMALE WITH A PAST MEDICAL HISTORY OF HEADACHES UNDERWENT STENT-ASSISTED COIL EMBOLIZATION OF AN UNRUPTURED LEFT INTERNAL CAROTID ARTERY (ICA) BIFURCATION ANEURYSM ON (B)(6) 2021. IMMEDIATE PRE-PROCEDURE ANGIOGRAPHY REVEALED AN ANTERIOR CIRCULATION ANEURYSM IN THE LEFT ICA. THE UNRUPTURED ANEURYSM HAD THE FOLLOWING DIMENSIONS: HEIGHT 4.0MM, DOME 3.5MM, MAXIMUM ANEURYSM DIAMETER 4.0MM, NECK SIZE 2.4MM, AND DOME-TO-NECK RATIO 1.5MM. THE PARENT VESSEL DIAMETER WAS 2.9MM. THE NEUROFORM ATLAS® STENT SYSTEM (STRYKER) STENT-ASSISTED COIL EMBOLIZATION WAS SUBSEQUENTLY PERFORMED WITH THE IMPLANTATION OF ONE 3MM X 6CM GALAXY G3 XSFT (GLX120306 / 30470347), TWO 2MM X 3CM GALAXY G3 MINI (GLM920030 / 30407098 & 30401363), AND ONE 1.5MM X 2CM GALAXY G3 MINI (GLM915020 / 30366403). A PHENOM¿ 17 MICROCATHETER (MEDTRONIC). PER THE INFORMATION DOCUMENTED IN THE CASE REPORT FORM (CRF), MICROCATHETER KICKBACK (LOSS OF ACCESS TO ANEURYSM) WAS EXPERIENCED ONCE DURING THE PROCEDURE. ADDITIONALLY, ONE 3.5MM X 7.5CM GALAXY G3 (GLY123575 / L17137) AND ONE 1.5MM X 2CM GALAXY G3 MINI (GLM915020 / 30392244) WERE USED BUT NOT IMPLANTED. HEPARIN WA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12051329 ↗InjuryJan 28, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2019, IT WAS REPORTED VIA THE STERLING STUDY THAT THE (B)(6) FEMALE PATIENT EXPERIENCED ¿TINY STROKES.¿ THE PATIENT IS A CURRENT SMOKER WITH A HISTORY OF DIABETES, HEADACHES, HYPERLIPEMIA, AND HYPERTENSION PRESENTED WITH A RUPTURED MIDLINE, BIFURCATION ANTERIOR COMMUNICATING ARTERY ANEURYSM WITH SUBARACHNOID HEMORRHAGE (SAH) ON (B)(6) 2019. THE ANEURYSM WAS 6.8 MM IN HEIGHT WITH A 3.8 MM DOME, MAXIMUM NECK SIZE WAS 6.8 MM, AND A NECK SIZE OF 5.1 MM. THE DOME TO NECK RATIO WAS 0.7 MM AND THE PARENT VESSEL WAS 2.1 MM IN DIAMETER. THE PATIENT¿S PRE-PROCEDURE MRS SCORE WAS 2 AND HUNT & HESS GRADE WAS 2. PRE-PROCEDURE SYMPTOMS INCLUDED HEADACHE WITH PHOTOPHOBIA AND DROWSINESS. TREATMENT WAS PERFORMED ON (B)(6) 2019, WITH THE SUCCESSFULLY IMPLANTATION OF THE FOLLOWING COILS USING THE NEURON¿ MICROCATHETER (PENUMBRA) IN THE FOLLOWING ORDER: 2 MM X 4 CM GALAXY G3 MINI (GLM920040 / L13705), 3 MM X 8 CM GALAXY G3 XSFT (GLX120308 / L12913), 2 MM X 4 CM GALAXY G3 MINI (GLM920040 / L12027), 3 MM X 8 CM GALAXY G3 XSFT (GLX120308 / L11777) AND 3.5 MM X 7.5 CM GALAXY G3 (GLY123575 / L12491). THERE WAS NO MICROCATHETER KICKBACK AND NO INTRA-OPERATIVE THROMBOEMBOLIC EVENTS. PACKING DENSITY
Patient problems: Infarction, Cerebral
MAUDE report 8283057 ↗MalfunctionOct 5, 2017Stretched; Stretched
AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING TREATMENT OF A SMALL, MALFORMED ANEURYSM, A GALAXY G3 (GLY123575/S11528) STRETCHED DURING REPOSITIONING OF THE COIL IN THE ANEURYSM. THE EVENT OCCURRED WHEN THE ANEURYSM WAS 29% FILLED. IT WAS REPORTED THAT THE FILLING WAS SUCCESSFUL BECAUSE THE ANEURYSM WAS SMALL AND MALFORMED. THE COIL WAS REMOVED STILL ATTACHED TO THE DPU. NECK REMODELING HAD BEEN PERFORMED WITH A DUO ECLIPSE BALLOON. A CONTINUOUS FLUSH HAD BEEN MAINTAINED THROUGH THE MICROCATHETER. ANOTHER COIL HAD PASSED THROUGH THE SAME MICROCATHETER PRIOR TO THIS EVENT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES OR PROCEDURE DELAY, AND THE PROCEDURE WAS COMPLETED WITH OTHER SPECTRA COILS. IT WAS REPORTED THAT THE DEVICE WOULD BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6915827 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page