MICRUSFRAME MFR100305Recalls & reported problems
Cerenovus, Inc. · MICRUSFRAME S Microcoil 10 Stretch Resistant Coil and Detachable Coil System Contents: 1 Detachable Coil System Coil Type: MICRUSFRAME S 10 Microcoil 3mm (coil diameter) x 5.4cm (coil length) 190cm (working length) · Class 2 · Implant
All models of MICRUSFRAME (47)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886704077800), received between Aug 16, 2018 and Nov 4, 2023 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Entrapment of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Entrapment of Device2
- Positioning Problem2
- Adverse Event Without Identified Device or Use Problem1
- Device Difficult to Setup or Prepare1
- Migration1
- Separation Failure1
- Stretched1
Patient problems
- No Consequences Or Impact To Patient3
- No Clinical Signs, Symptoms or Conditions2
- Hemorrhage, Subarachnoid1
- No Code Available1
- Rupture1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 4, 2023Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, THE COMPLAINT COIL, A 3MM X 5.4CM MICRUSFRAME 10 COIL (MFR100305 / 30810134) FAILED TO DETACH. IT WAS REPORTED THAT THE COIL WAS IN THE ANEURYSM BUT THE DETACHMENT SYSTEM FAILED SO THE COIL COULD NOT SEPARATE FROM THE DELIVERY WIRE. THEY CHANGED THREE (3) CABLES AND TWO (2) DETACHMENT BOXES, BUT IT DID NOT DETACH. THEY HAD TO REMOVE THE COIL. THE PROCEDURE WAS DELAYED; HOWEVER, THE DURATION WAS NOT REPORTED. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18070330 ↗MalfunctionJan 26, 2021Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION AND DURING THE RESHEATHING OF A 3MM X 5.4CM MICRUSFRAME10 COIL (MFR100305, L16030), THE INTRODUCER FAILED TO BE REZIPPED. THERE WAS NO PATIENT INJURY REPORTED. BASED ON THE DEVICE ANALYSIS OF THE PRODUCT RECEIVED, THE EMBOLIC COIL WAS FOUND STRETCHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11228791 ↗MalfunctionAug 18, 2020Positioning Problem; Positioning Problem
AS REPORTED BY THE FILED, DURING A COIL EMBOLIZATION, THE PHYSICIAN DIDN'T LIKE THE SHAPE OF THE COIL DEPLOYMENT OF A 3MM X 5.4CM MICRUSFRAME10 (MFR 100305, L15956) AND DECIDED TO PULL IT OUT. DURING RESHEATHING OF THE COIL, THE INTRODUCER FAILED TO BE RE-ZIPPED. THE PHYSICIAN USED A SAME LIKE PRODUCT. THE PROCEDURE WAS SUCCESSFULLY FINISHED. THERE WAS NOT PATIENT INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10419763 ↗MalfunctionAug 6, 2019Stretched; Stretched
AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING THE PREPPING OF A 3MM X 5.4CM MICRUSFRAME10 COIL (MFR100305, L14989), THE SHEATH SPLIT. THERE WAS NO PATIENT INJURY REPORTED. THE PROCEDURE WAS COMPLETED BY A COMPETITOR¿S COIL. THE PRODUCT WAS STORED ACCORDING TO LABELING INSTRUCTIONS. THE DEVICE WAS ADEQUATELY USED. THERE WAS NO EXCESSIVE FORCE APPLIED AT ANY TIME DURING THE PROCEDURE. THE COIL DEVICE WAS ATTEMPTED TO BE INSERTED INTO THE SL10 MICROCATHETER (MC). THERE WAS NO RESISTANCE/FRICTION ENCOUNTERED WHEN INSERTING THE COIL DELIVERY SYSTEM INTO THE MICROCATHETER. THERE WERE NO KINKS ON THE MC THAT MAY HAVE CONTRIBUTED TO THE EVENT. THERE WAS NO EVIDENCE OF DAMAGE TO THE ACTUAL COIL (I.E. STRETCHING, KINKED, DETACHED, SEPARATED). BASED ON THE ANALYSIS OF THE DEVICE RECEIVED, THE EMBOLIC COIL WAS STRETCHED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8865393 ↗InjuryJun 5, 2019Entrapment of Device; Migration; Entrapment of Device; Migration
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE STENT-ASSISTED COIL EMBOLIZATION PROCEDURE TARGETING THE ANEURYSM ON THE TERMINAL INTERNAL CAROTID ARTERY (ICA), A LOOP OF THE 3MM X 5.4CM MICRUSFRAME 10 COIL (MFR100305 / UNK LOT) IMPLANTED IN THE TARGET LESION WAS PUSHED OUT INTO THE PARENT VESSEL BY THE REPORTED STIFFNESS OF THE 2MM X 2CM DELTAPLUSH 10 COIL (DPL10020220/S10016). AS A RESULT, A NEUROFORM ATLAS® STENT SYSTEM (STRYKER) WAS SUBSEQUENTLY PLACED OVER THE PROTRUDING COIL LOOP TO SECURE THE COIL TO THE PARENT VESSEL WALL. A 2 MM X 2 CM BARRICADE¿ HELICAL COIL (BLOCKADE MEDICAL) WAS USED TO REPLACE THE DELTAPLUSH COIL AFTER THE STENT PLACEMENT WITHOUT INCIDENT; THE COIL WAS PLACED THROUGH THE STRUTS OF THE ATLAS STENT. THE COIL LOOP DID NOT CAUSE ANY RESTRICTION OR REDUCTION IN BLOOD FLOW; THERE WAS NO THROMBOSIS REPORTED AS A RESULT OF THE EVENT. THERE IS NO ADDITIONAL INTERVENTION PLANNED TO ADDRESS THE REMAINING COIL LOOP IN THE PARENT VESSEL; THE COIL LOOP WAS SECURED TO THE PARENT VESSEL WALL BY THE STENT. THERE WAS NO REPORT OF ANY PATIENT INJURY OR COMPLICATION AS A RESULT OF THE REPORTED EVENT. THE MICRUSFRAME COIL REMAINS IMPLANTED IN THE PATIENT AND IS THUS,
Patient problems: No Code Available
MAUDE report 8670595 ↗InjuryNov 14, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED VIA THE (B)(6) REGISTRY, A (B)(6) FEMALE (PATIENT (B)(6)), WITH A HISTORY OF CARDIAC ARRHYTHMIA, HEADACHES, SMOKING, AND CONTROLLED HYPERTENSION EXPERIENCED SEVERE INTRAOPERATIVE TARGET ANEURYSM RUPTURE DURING DEPLOYMENT OF A 3MM X 5.4CM MICRUSFRAME10 (MFR100305/L10115) THERMO-MECHANICAL COIL. LATER, A 1.5MM X 3CM GALAXY G3 MINI (GLM915030/L12447) THERMO-MECHANICAL COIL PARTIALLY PROTRUDED INTO THE ANEURYSM NECK AND PARENT VESSEL WITH CLOT, RESULTING IN NEAR VESSEL OCCLUSION FOR ONE MINUTE, BUT DID NOT RESULT IN PATIENT INJURY. AT THE TIME OF THE ANEURYSM RUPTURE, THE PATIENT¿S SYSTOLIC BLOOD PRESSURE IMMEDIATELY DROPPED TO BELOW 100 AND THE PATIENT WAS ADMINISTERED 50MG OF PROTAMINE FOR HEPARIN REVERSAL AND A 250CC BOLUS OF 3% HYPERTONIC SALINE. THE PATIENT WAS HYPERVENTILATED, AND NEUROSURGERY WAS CONTACTED IMMEDIATELY. THEY WERE ABLE TO FINISH DEPLOYING THE NEUROFORM ATLAS (STRYKER) STENT WITH DETACHMENT AND RAPID COILING OF THE REMAINING COILS. THE SL-10 (STRYKER) MICROCATHETER WAS ADVANCED OVER THE TRAXCESS 14 (MICROVENTION) MICROWIRE ACROSS THE ANEURYSM NECK, SPANNING INTO THE RIGHT PERICALLOSAL ARTERY AND THE SECOND NEUROFORM ATLAS STENT WAS ADVANCED AND DEPLOYED
Patient problems: Hemorrhage, Subarachnoid; Rupture
MAUDE report 8071647 ↗MalfunctionAug 16, 2018Entrapment of Device; Positioning Problem; Entrapment of Device; Positioning Problem
AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING A COIL EMBOLIZATION OF A 3.5 X 2.8 X 2.7MM RUPTURED ANEURYSM, THE 3MM X 5.4CM MICRUSFRAME10 (MFR100305/UNK LOT) THERMO-MECHANICAL FRAMING COIL APPEARED TO FIT PERFECTLY IN THE BEGINNING OF THE CASE, BUT THE PROXIMAL PORTION OF THE EMBOLIC COIL FAILED TO BREAK. THE COIL WAS SUPPLIED WITH A 3MM NON-CERENOVUS COIL, WHICH WAS CONSIDERED TOO SMALL. IT WAS FURTHER STATED THAT "THE LAST PART OF THE COIL DID NOT PERFORM LOOPS ANYMORE". THERE WAS NO REPORT OF CONSEQUENCE OF IMPACT TO THE PATIENT. IMAGING WAS PROVIDED BY THE CUSTOMER. THE PRODUCT WILL NOT BE RETURNED FOR ANALYSIS. IT WAS REPORTED THAT THE INFORMATION PROVIDED FOR THIS COMPLAINT WAS ALL OF THE DETAILS THAT WERE KNOWN/AVAILABLE REGARDING THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7789337 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.