MICRUSFRAME MFR140717Recalls & reported problems
Cerenovus, Inc. · MICRUSFRAME C Microcoil 14 Stretch Resistant Coil and Detachable Coil System Contents: 1 Detachable Coil System Coil Type: MICRUSFRAME C 14 Stretch Resistant Coil 7mm (coil diameter) x 17cm (coil length) 190cm (working length) · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886704078012), received between Nov 11, 2019 and Mar 6, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Separation Failure2
- Break1
- Detachment of Device or Device Component1
- Positioning Problem1
- Premature Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 6, 2024Separation Failure; Premature Separation; Separation Failure; Premature Separation
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE TARGETING A LARGE 8MM ANEURYSM LOCATED ON THE INTERNAL CAROTID ARTERY (ICA) JUST AFTER SYPHON / MEDIAL OF OPHTHALMIC ARTERY IN A YOUNG PATIENT WITH VERY TORTUOUS ANATOMY, THE ECHELON¿ 14 MICROCATHETER (MEDTRONIC) WAS PLACED INTO THE ANEURYSM BUT THE POSITION WAS NOT VERY STABLE DUE TO ANATOMY AND ANGLES, THE COILS WERE NOT DETACHABLE. THE FIRST COIL, A 7MM X 30CM MICRUSFRAME 10 (MFR100730 / 30860269) WAS CHECKED IN SALINE FOR CONFIGURATION AND CHECKED ELECTRICALLY; NO PROBLEMS WERE NOTED. THE COIL WAS PUSHED THROUGH THE MICROCATHETER WITH A LITTLE MORE RESISTANCE THAN USUALLY NOTED. THE COIL WAS THEN POSITIONED INTO THE TARGET ANEURYSM PERFECTLY. DETACHMENT DID NOT WORK. THE PHYSICIAN MADE 10 TO 12 UNSUCCESSFUL ATTEMPTS TO DETACH THE COIL. THE DETACHMENT CONTROL BOX AND THE CONTROL CABLES WERE EXCHANGED BUT THE COIL WAS STILL NOT DETACHABLE. THE BALLOON CATHETER (UNSPECIFIED BRAND) WAS INFLATED TO PROTECT AGAINST SMALL EMBOLIC COMING FROM THE ANEURYSM. THE COIL WAS REMOVED. DURING THE REMOVAL PROCESS, ELECTRIC SIGNAL WAS LOST AND AFTER COMPLETE REMOVAL, NO ELECTRICAL CONTACT WAS POSSIBLE. THE SE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18853229 ↗MalfunctionMay 4, 2022Break; Positioning Problem; Break; Positioning Problem
THE HEALTHCARE PROFESSIONAL REPORTED DURING AN ENDOVASCULAR ANEURYSM REPAIR (EVAR) COIL EMBOLIZATION PROCEDURE AT THE INFERIOR MESENTERIC ARTERY (IMA), A CONCOMITANT PROWLER SELECT PLUS MICROCATHETER WAS INDUCED TO THE TARGET LESION AFTER THE PARENT CATHETER WAS INTRODUCED INTO THE IMA. THEN THE 7MM X 17CM MICRUSFRAME 14 COIL (MFR140717 / 30362732) WAS USED. THE PHYSICIAN ATTEMPTED TO MAKE A FRAME BUT THE MICRUSFRAME WAS NOT ABLE TO MAKE THE SHAPE AS THE PHYSICIAN INTENDED. THE MICRUSFRAME WAS RESHEATHED. THEN THE 8MM X 35CM DELTAFILL 18 COIL (DLF180835 / 30464336) WAS USED, BUT IT WAS ALSO RESHEATHED BECAUSE IT WAS NOT ABLE TO MAKE THE SHAPE AS THE PHYSICIAN INTENDED. THE SIZE OF THE COIL WAS CHANGED WITH A SMALLER COIL AND OTHER COILS WERE USED AND THE FRAMING WAS SUCCESSFULLY COMPLETED. AFTER THE SUCCESSFUL FRAMING, THE RESHEATHED 7MM X 17CM MICRUSFRAME 14 AND 8MM X 35CM DELTAFILL 18 WERE USED, BUT THEY WERE REMOVED FROM THE PATIENT¿S BODY BECAUSE THE SHEATHS ON THESE COILS WERE DAMAGED AND NOT ABLE TO BE GUIDED. COIL EMBOLIZATION WAS PERFORMED WITH OTHER COILS AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT ANY PATIENT ADVERSE EVENT OR COMPLICATION. CONTINUOUS FLUSH W
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14274526 ↗MalfunctionOct 12, 2021Separation Failure; Separation Failure
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION FOR THE RIGHT INTERNAL ILIAC ARTERY ANEURYSM, A CONCOMITANT MICROCATHETER, LEONISMOVA, SUMITOMO BAKELITE AND A MICROCATHETER, CARNELIAN MARVEL, TOKAI MEDICAL PRODUCTS WERE APPROACHED TO THE RIGHT SUPERIOR GLUTEAL ARTERY. A MICRUSFRAME14 7MM X 17CM COIL (MFR140717, 30375954) WAS USED AS THE FIRST COIL. THE TOTAL LENGTH OF THE COMPLAINT COIL CAME OUT OF THE MICROCATHETER WITHOUT ANY PROBLEM. THE DETACHMENT CYCLE WAS DONE AND THE DELIVERY WIRE WAS SLOWLY PULLED, HOWEVER, THE COMPLAINT COIL WAS NOT DETACHED. EVEN WHEN THE DETACHING OPERATION WAS PERFORMED AGAIN, THE COMPLAINT COIL WAS NOT DETACHED AS IN THE FIRST CHALLENGE. THE CONTROLLER BOX WAS TURNED ON AND OFF AGAIN. HOWEVER, THE DETACHMENT FAILURE CONTINUED. IT WAS CONFIRMED THE UPPER GREEN POWER LIGHT WAS ILLUMINATED DURING NORMAL OPERATION. THE COMPLAINT COIL WAS REPLACED WITH ANOTHER COIL OF THE SAME LOT AND THE REPLACEMENT COIL WAS DETACHED. AN ENPOWER CONTROL CABLE WAS USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12614893 ↗MalfunctionNov 11, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING A COIL EMBOLIZATION, A 7MM X 17CM MICRUSFRAME14 COIL (MFR140717, L12061) WAS BEING USED AS A THIRD COIL, HOWEVER, TRIED IT SEVERAL TIMES, BUT IT WAS NOT ¿DETAILED¿. THERE WAS NO PATIENT INJURY REPORTED. A PICTURE OF THE DEVICE WAS RECEIVED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9304935 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page