MICRUSFRAME MFR180813Recalls & reported problems
Cerenovus, Inc. · MICRUSFRAME S Microcoil 18 Stretch Resistant Coil and Detachable Coil System Contents: 1 Detachable Coil System Coil Type: MICRUSFRAME S 18 Stretch Resistant Coil 8mm (coil diameter) x 13.6cm (coil length) 190cm (working length) · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704078050), received between Oct 18, 2022 and Jan 17, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Deformation Due to Compressive Stress1
- Separation Failure1
Patient problems
- Aneurysm1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 17, 2023Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
AS REPORTED BY THE STERLING STUDY, A 54-YEAR-OLD FEMALE (SUBJECT (B)(6) WITH NO TARGETED MEDICAL HISTORY UNDERWENT BALLOON-ASSISTED COIL EMBOLIZATION OF A RUPTURED RIGHT ANTERIOR CEREBRAL ARTERY ANEURYSM ON (B)(6) 2022. THE PATIENT¿S BASELINE MODIFIED RANKIN SCALE SCORE WAS 0 AND THE HUNT & HESS GRADE WAS 2 - MODERATE TO SEVERE HEADACHE, NUCHAL RIGIDITY, AND NO NEUROLOGIC DEFICIT OTHER THAN CRANIAL NERVE PALSY. THE DIMENSIONS OF THE ANTERIOR CEREBRAL ARTERY ANEURYSM WERE AS FOLLOWS: ANEURYSM HEIGHT OF 8.5 MM, DOME 8.1 MM, MAXIMUM ANEURYSM DIAMETER OF 8.1 MM, NECK SIZE OF 2.6 MM, AND THE DOME-TO-NECK RATIO OF 3.1 MM. THE PARENT VESSEL DIAMETER WAS 1.7 MM. BALLOON-ASSISTED COIL EMBOLIZATION WAS PERFORMED WITH FOUR (4) CERENOVUS COILS VIA AN SL-10 MICROCATHETER (STRYKER). THE COILS WERE THE FOLLOWING: ONE 8MM X 13.6MM MICRUSFRAME S (MFR180813/K10699), ONE 6 MM X 15 CM GALAXY G3 (GLY120615/K10390), ONE 4 MM X 12 CM GALAXY G3 (GLY120412/30637578), AND ONE GALAXY G3 XTRASOFT (GLX120306/30603573). HEPARIN WAS ADMINISTERED. THE SUBJECT DID NOT EXPERIENCE INTRA-OPERATIVE RUPTURE OR THROMBOEMBOLIC EVENTS. IN THE OPINION OF THE INVESTIGATOR, TREATMENT OF THE TARGET ANEURYSM WAS CONSIDERED COM
Patient problems: Aneurysm
MAUDE report 16187039 ↗MalfunctionOct 18, 2022Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, THE PHYSICIAN ATTEMPTED TO DEPLOY AN 8MM X 13.6CM MICRUSFRAME 18 COIL (MFR180813 / K10699), BUT THE COIL WOULD NOT DETACH. MULTIPLE ATTEMPTS WERE MADE AND THE STAFF BELIEVES THAT THE ENPOWER DETACHMENT CONTROL BOX (DCB2000500) IS NOT FUNCTIONING CORRECTLY. THE PHYSICIAN HAD TO REMOVE THE COIL FROM THE PATIENT AND SWITCHED TO STRYKER COILS. THERE WAS NO DELAY REPORTED. THE COMPLAINT COIL WAS SUCCESSFULLY REMOVED FROM THE PATIENT AND THE REPLACEMENT STRYKER COIL WAS USED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO NEGATIVE PATIENT OUTCOME AS A RESULT OF THE REPORTED ISSUE. ON 06-OCT-2022, ADDITIONAL INFORMATION WAS RECEIVED. THE INFORMATION INDICATED THAT THE COIL EMBOLIZATION PROCEDURE WAS TARGETING AN ANTERIOR COMMUNICATING ARTERY (ACOM) ANEURYSM. THE COIL WAS NOT STRETCHED WHEN IT WAS REMOVED; IT WAS STILL ATTACHED TO THE DELIVERY SYSTEM WHEN IT WAS REMOVED FROM THE PATIENT. A PRE-DEPLOYMENT CHECK WAS PERFORMED. THE FAULT LIGHT WAS NOT SEEN DURING THE CASE. UPON PRESSING THE POWER BUTTON, ALL THE LIGHTS DID ILLUMINATE AND THE GREEN SYSTEM READY LIGHT DID ILLUMINATE. IT WAS NOT KNOWN IF
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15625932 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page