Skip to content

MICRUSFRAME MFR140711Recalls & reported problems

Cerenovus, Inc. · MICRUSFRAME C Microcoil 14 Stretch Resistant Coil and Detachable Coil System Contents: 1 Detachable Coil System Coil Type: MICRUSFRAME C 14 Stretch Resistant Coil 7mm (coil diameter) x 11cm (coil length) 190cm (working length) · Class 2 · Implant

All models of MICRUSFRAME (47)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704079347), received between Jan 12, 2022 and Aug 23, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Failure to Advance” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Advance1
  • Premature Separation1
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 23, 2024Separation Failure; Premature Separation; Separation Failure; Premature Separation

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, AFTER THE INTRODUCTION OF THE COIL, A 7MM X 11CM MICRUSFRAME 14 (MFR140711 / 31136249), IT COULD NOT BE RELEASED. IT WAS REPORTED THAT THE PHYSICIAN WANTED TO PULL THE COIL OUT OF THE ANEURYSM, BUT IT BECAME PREMATURELY DETACHED. THE PROCEDURE WAS DELAYED BY 20 MINUTES. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20053413 ↗
  • MalfunctionJan 12, 2022Failure to Advance; Failure to Advance

    AS REPORTED BY THE FIELD, DURING AN INTRACRANIAL COIL EMBOLIZATION, A CERENOVUS COIL WAS PLACED WITHOUT TROUBLE AND THEN A MICRUSFRAME14 7MMX11CM COIL (MFR140711, LOT UNKNOWN) WAS BEING USED AS THE SECOND COIL. HOWEVER, THE COIL COULD NOT BE ADVANCED WITHIN PRESHAPED PROWLER14 MICROCATHETER (606151FX, LOT UNKNOWN). THUS, THE COIL AND THE MC HAD TO BE WITHDRAWN. A NEW CERENOVUS COIL WAS USED WITH A NEW MC AND THE PROCEDURE WAS COMPLETED. ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE COIL DID NOT APPEAR DAMAGED, IT COULD SIMPLY NOT BE PUSHED ANY FURTHER WITHIN THE MICROCATHETER. AFTER ADVANCEMENT WAS NOT POSSIBLE, WITHDRAWAL WAS TRIED BUT ALSO IMPOSSIBLE, THE COIL GOT STUCK. NOTHING WAS NOTED TO BE OBSTRUCTING THE MICROCATHETER. A CONTINUOUS FLUSH ON THE MICROCATHETER WAS MAINTAINED. THE INTRODUCER WAS FLUSHED UNTIL LIQUID WAS VISIBLE AT THE DISTAL END OF THE SLIT IN THE CLEAR TUBE. THE TIP OF THE INTRODUCER WAS FIRMLY INSTALLED INTO THE HUB OF THE MICROCATHETER AND LOCKED WITH THE RHV DURING DEVICE ADVANCEMENT. NO EXCESSIVE FORCE WAS APPLIED TO THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13235372 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page