Skip to content

CEREPAK FREEFORM MCR091030Recalls & reported problems

Cerenovus, Inc. · CEREPAK FREEFORM Mini Detachable Coil Delivery System Length - 190 cm Coil Diameter - 1 mm Coil Length - 3 cm · Class 2 · Implant

All models of CEREPAK FREEFORM (60)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886704082934), received between Jun 28, 2023 and Aug 6, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Separation Failure” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 2, 2025Class IOpen, Classified

    Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.

    Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Separation Failure5
  • Fracture1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2025Separation Failure; Separation Failure

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, THE PHYSICIAN WAS PERFORMING A PROCEDURE TO EMBOLIZE THE MIDDLE MENINGEAL ARTERY (MMA). THE DOCTOR WAS UTILIZING CEREPAK COILS FOR THE EMBOLIZATION. UPON DEPLOYING A FREEFORM MINI 1MM X 3CM COIL (MCR091030, 31140702) INTO THE ARTERY, THE FIRST ATTEMPT TO DETACH THE COIL WAS WITH A MECHANICAL DETACHMENT HANDLE (MDH1,102109430) WHICH WAS UNSUCCESSFUL. AT WHICH POINT THE DOCTOR WAS INSTRUCTED TO PERFORM A MANUAL BREAK TECHNIQUE WHICH WAS ALSO UNSUCCESSFUL. THE DOCTOR REMOVED THE COIL FROM THE MICROCATHETER SUCCESSFULLY AND THE COIL WAS SAVED FOR ANALYSIS. AFTER DEPLOYING 3 OTHER COILS SUCCESSFULLY, ANOTHER FREEFORM MINI 1MM X 3CM COIL (MCR091030, 31140702) WAS DEPLOYED AND UPON MECHANICAL HANDLE DETACHMENT BEING UNSUCCESSFUL, AGAIN, THE MANUAL BREAK TECHNIQUE WAS PERFORMED AT WHICH THE PULL WIRE FRACTURED AND COULD NOT BE RETRIEVED THEREFORE RESULTING IN A FAILED DETACHMENT. THE COIL WAS SUCCESSFULLY RETRIEVED THROUGH THE MICROCATHETER, BUT THE DEVICE WAS DESTROYED AND UNABLE TO BE SAVED FOR ANALYSIS. ADDITIONAL EVENT INFORMATION RECEIVED ON 25-JUL-2025 CONFIRMED THAT TWO FREEFORM MINI 1MM X 3CM COILS FAILED TO DETACH. A TERUMO HEADWAY DUO 156

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22727225 ↗
  • MalfunctionAug 6, 2025Fracture; Separation Failure; Fracture; Separation Failure

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, THE PHYSICIAN WAS PERFORMING A PROCEDURE TO EMBOLIZE THE MIDDLE MENINGIAL ARTERY (MMA). THE DOCTOR WAS UTILIZING CEREPAK COILS FOR THE EMBOLIZATION. UPON DEPLOYING A FREEFORM MINI 1MM X 3CM COIL (MCR091030, 31140702) INTO THE ARTERY, THE FIRST ATTEMPT TO DETACH THE COIL WAS WITH A MECHANICAL DETACHMENT HANDLE (MDH1,102109430) WHICH WAS UNSUCCESSFUL. AT WHICH POINT THE DOCTOR WAS INSTRUCTED TO PERFORM A MANUAL BREAK TECHNIQUE WHICH WAS ALSO UNSUCCESSFUL. THE DOCTOR REMOVED THE COIL FROM THE MICROCATHETER SUCCESSFULLY AND THE COIL WAS SAVED FOR ANALYSIS. AFTER DEPLOYING 3 OTHER COILS SUCCESSFULLY, ANOTHER FREEFORM MINI 1MM X 3CM COIL (MCR091030, 31140702) WAS DEPLOYED AND UPON MECHANICAL HANDLE DETACHMENT BEING UNSUCCESSFUL, AGAIN, THE MANUAL BREAK TECHNIQUE WAS PERFORMED AT WHICH THE PULL WIRE FRACTURED AND COULD NOT BE RETRIEVED THEREFORE RESULTING IN A FAILED DETACHMENT. THE COIL WAS SUCCESSFULLY RETRIEVED THROUGH THE MICROCATHETER, BUT THE DEVICE WAS DESTROYED AND UNABLE TO BE SAVED FOR ANALYSIS. ADDITIONAL EVENT INFORMATION RECEIVED ON 25-JUL-2025 CONFIRMED THAT TWO FREEFORM MINI 1MM X 3CM COILS FAILED TO DETACH. A TERUMO HEADWAY DUO 156

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22727028 ↗
  • MalfunctionJul 1, 2025Premature Separation; Premature Separation

    A HEALTHCARE PROFESSIONAL REPORTED THAT WHILE DEPLOYING A FREEFORM MINI 1MM X 3CM COIL (MCR091030, 31177297) IN THE ANEURYSM, THE COIL DETACHED PREMATURELY. THE USER WAS ABLE TO USE PUSHER WIRE TO PUSH THE REST OF THE COIL INTO THE ANEURYSM WITHOUT AN EVENT. THE DELIVERY SYSTEM FOR THE COIL IS AVAILABLE FOR RETURN, HOWEVER THE COIL ITSELF IS NOT, SINCE IT WAS SUCCESSFULLY IMPLANTED IN THE PATIENT. THERE WERE NO SURGERY DELAYS DUE TO THE REPORTED EVENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22375680 ↗
  • MalfunctionJul 8, 2024Separation Failure; Separation Failure

    AS REPORTED BY THE FIELD, THIS WAS A COIL EMBOLIZATION TO A POSTERIOR COMMUNICATING ARTERY (PCOM) ANEURYSM. THE FIRST 4 COILS WERE DEPLOYED WITHOUT AN ISSUE. THE 5TH COIL, A FREEFORM MINI 1MM X 3CM ( MCR091030, 31140702) WAS SITUATED AND DETACHMENT WAS ATTEMPTED WITH A MECHANICAL DETACHMENT HANDLER (MDH1, 102109430) WHICH DID NOT DETACH. MANUAL BREAK METHOD WAS THEN ATTEMPTED, ONCE AGAIN FAILING TO DETACH AT BOTH 2CM PULL AND 4CM PULL. COIL WAS REMOVED AND A 6TH COIL, A FREEFORM MINI 1MM X 2CM (MCR091020, 31131257) WAS POSITIONED. AFTER ATTEMPTING AND FAILING DETACHMENT WITH HANDLE MANUAL BREAK METHOD WAS UTILIZED AGAIN. THE COIL STILL FAILED TO DETACH AND WAS THEREFORE REMOVED. CASE WAS COMPLETED WITH ONE COMPETITIVE COIL. THERE WAS A 10-MINUTE PROCEDURAL DELAYS DUE TO THE EVENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. ADDITIONAL EVENT INFORMATION WAS RECEIVED ON 24-JUN-2024 INDICATING THAT A PHENOM 17 150CM MICROCATHETER WAS USED. THERE WAS NO RESISTANCE DURING ADVANCEMENT OF THE DEVICES THROUGH THE MICROCATHETER OR POSITIONING DIFFICULTY IN THE ANEURYSM. THERE WAS NO DAMAGE TO THE EMBOLIC COIL OR ANY DEVICE COMPONENT. THE VESSEL WAS MODERATELY TORTUOUS, NOTHING ¿EXTREME¿ PER T

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19689482 ↗
  • MalfunctionMay 8, 2024Separation Failure; Separation Failure

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING COIL EMBOLIZATION OF A MIDDLE MENINGEAL ARTERY (MMA), TWO COILS, A 1MM X 3CM CEREPAK FREEFORM MINI (MCR091030 / 31177297) AND A 1.5MM X 2.5CM CEREPAK FREEFORM MINI (MCR091525 / 31149187) DID NOT DETACH CORRECTLY. THE CEREPAK TM DETACHMENT HANDLE (MDH1 / 102109430) AS WELL AS THE MANUAL BREAK WERE ATTEMPTS AND NEITHER METHOD WORKED. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19272090 ↗
  • MalfunctionJun 28, 2023Separation Failure; Separation Failure

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING A COIL EMBOLIZATION OF AN ANEURYSM TO THE MIDDLE MENINGEAL ARTERY, A FREEFORM MINI 1MM X 1.5CM COIL (MCR091030, 31045735) FAILED TO DETACH. NO PREVIOUS COILS HAD BEEN PREVIOUSLY IMPLANTED. THERE HAD NOT BEEN ANY RESISTANCE DURING ADVANCEMENT OF THE DEVICE THROUGH THE SL10 STRYKER MICROCATHETER (MC) OR POSITIONING DIFFICULTY IN THE ANEURYSM. THERE HAD BEEN THREE ATTEMPTS MADE USING THE DETACHMENT HANDLE PRIOR TO MANUAL DETACHMENT. THERE WAS NO DAMAGE TO THE EMBOLIC COIL OR ANY DEVICE COMPONENT. THERE WAS NO VESSEL ACUTE BENDS. THERE WAS NO PATIENT INJURY REPORTED. THERE WAS ABOUT A 20-MINUTE PROCEDURAL DELAY, WE HAD TO BRING IN THE BALT COIL CART.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17215908 ↗

FDA clearance and classification

Cleared through 510(k) K220040 on Jul 14, 2022 (decision: Substantially Equivalent). Submitted by Medos International SARL as “CEREPAK Uniform, Uniform XL, Uniform 3D, Heliform Soft, Heliform XtraSoft, Heliform XL, Heliform XtraSoft XL, Freeform, Freeform Mini, and Freeform XtraSoft Detachable Coil Systems”. The FDA review took 190 days. FDA record ↗

The FDA regulates product code KRD (device, vascular, for promoting embolization) under 21 CFR 870.3300 as a Class II device. 21 CFR 870.3300 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the CEREPAK FREEFORM MCR091030

Has the CEREPAK FREEFORM MCR091030 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0434-2026 (Class I), began on Oct 2, 2025: Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the CEREPAK FREEFORM MCR091030?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Jun 28, 2023 and Aug 6, 2025. In those reports, the most common specific device problems are “Separation Failure” (5), “Fracture” (1) and “Premature Separation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CEREPAK FREEFORM MCR091030?

None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CEREPAK FREEFORM MCR091030 FDA approved?

It was cleared through 510(k) K220040 on Jul 14, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the CEREPAK FREEFORM MCR091030 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the CEREPAK FREEFORM MCR091030, and is it still sold?

The FDA's device database lists Cerenovus, Inc. as the company responsible for it (brand CEREPAK FREEFORM, model MCR091030). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page