CEREPAK FREEFORM MCR091525Recalls & reported problems
Cerenovus, Inc. · CEREPAK FREEFORM Mini Detachable Coil Delivery System Length - 190 cm Coil Diameter - 1.5 mm Coil Length - 2.5 cm · Class 2 · Implant
All models of CEREPAK FREEFORM (60)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886704082965), received between May 8, 2024 and Nov 3, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 2, 2025Class IOpen, Classified
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Separation Failure2
- Failure to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 3, 2025Failure to Advance; Failure to Advance
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE ON 03-SEP-2025, AS EACH COIL LISTED: A 1MM X 2CM CEREPAK FREEFORM MINI (MCR091020 / 31345668), A 1.5MM X 2.5CM CEREPAK FREEFORM MINI (MCR091525 / 31216617), 1MM X 2CM CEREPAK FREEFORM MINI (MCR091020 / 31258390), AND 1MM X 3CM CEREPAK FREEFORM MINI (MCR091030 / 31585342) WAS BEING ADVANCED INTO THE CONCOMITANT ECHELON¿ 10 MICROCATHETER (MEDTRONIC), THE PHYSICIAN ENCOUNTERED RESISTANCE AS THE COIL ENTERED THE MICROCATHETER. THE PHYSICIAN RE-SHEATHED THE COIL, CHECKED TO ENSURE THE INTRODUCER WAS ¿HUBBED¿ CORRECTLY WITHIN THE MICROCATHETER, CONFIRMED THE SYSTEM WAS BEING FLUSHED AND TRIED ADVANCE THE COIL AGAIN. IT WAS REPORTED THAT THE COIL WOULD NOT ADVANCE MORE THAN APPROXIMATELY 5 CM INTO THE MICROCATHETER. EACH COIL WAS PLACED ON THE BACK TABLE FOR INVESTIGATION FOLLOWING THE CASE. THE TWO COILS THAT WERE AVAILABLE FOR RETURN: 1MM X 2CM CEREPAK FREEFORM MINI (MCR091020) AND 1.5MM X 2.5CM CEREPAK FREEFORM MINI (MCR091525) WOULD NOT ADVANCE OUTSIDE OF THE INTRODUCER. NONE OF THE COILS WERE INTRODUCED INTO THE PATIENT¿S BODY. THE INABILITY OF THE COILS TO BE ADVANCED THROUGH THE MICROCATHETER DI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23444656 ↗MalfunctionJun 18, 2024Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, EACH OF THE THREE COILS USED A 1.5MM X 4CM CEREPAK FREEFORM MINI (MCR091540 / 31207073), A 1MM X 4CM CEREPAK FREEFORM MINI (MCR091040 / 31045747), AND A 1.5MM X 2.5CM CEREPAK FREEFORM MINI (MCR091525 / 31149187) DID NOT DETACH USING BOTH MANUAL BREAK AND WITH THE DETACHMENT HANDLE. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. ON 14-JUN-2024, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE PROCEDURE WAS EMBOLIZATION OF THE MIDDLE MENINGEAL ARTERY. THE CATALOG AND LOT NUMBER OF THE DETACHMENT HANDLE USED WAS MDH1 / 102109430. THE SAME DETACHMENT HANDLE WAS USED WITH ALL THREE COILS; IT WAS INDICATED AS NOT AVAILABLE FOR RETURN. MANUAL BREAK WAS ATTEMPTED ON THE FIRST COIL IMPLANTED. THE INFORMATION INDICATED THAT ¿THIS IS WHEN THE SYSTEM APPEARED TO BE BROKEN INTO TWO PIECES. THE PULL WIRE WAS NOT ATTACHED. MECHANICAL [DETACHMENT] WAS ATTEMPTED ON THE THIRD COIL (SECOND ATTEMPT WAS SUCCESSFUL) FIRST, THIS WAS UNSUCCESSFUL, WE MOVED TO MANUAL BREAK IMMEDIATELY WITHOUT SUCCESS. WE WRAPPED THE PULL WIRE AROUND FINGER AND PULLED AND THE PULL WIRE APPEARED TO HAVE BROKEN. THE FOUR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19563948 ↗MalfunctionMay 8, 2024Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING COIL EMBOLIZATION OF A MIDDLE MENINGEAL ARTERY (MMA), TWO COILS, A 1MM X 3CM CEREPAK FREEFORM MINI (MCR091030 / 31177297) AND A 1.5MM X 2.5CM CEREPAK FREEFORM MINI (MCR091525 / 31149187) DID NOT DETACH CORRECTLY. THE CEREPAK TM DETACHMENT HANDLE (MDH1 / 102109430) AS WELL AS THE MANUAL BREAK WERE ATTEMPTS AND NEITHER METHOD WORKED. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19272109 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page