CEREPAK FREEFORM MCR091525Recalls & reported problems

Cerenovus, Inc. · CEREPAK FREEFORM Mini Detachable Coil Delivery System Length - 190 cm Coil Diameter - 1.5 mm Coil Length - 2.5 cm · Class 2 · Implant

All models of CEREPAK FREEFORM (60)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886704082965), received between May 8, 2024 and Nov 3, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Separation Failure” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 2, 2025Class IOpen, Classified

    Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.

    Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Separation Failure2
  • Failure to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 3, 2025Failure to Advance; Failure to Advance

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE ON 03-SEP-2025, AS EACH COIL LISTED: A 1MM X 2CM CEREPAK FREEFORM MINI (MCR091020 / 31345668), A 1.5MM X 2.5CM CEREPAK FREEFORM MINI (MCR091525 / 31216617), 1MM X 2CM CEREPAK FREEFORM MINI (MCR091020 / 31258390), AND 1MM X 3CM CEREPAK FREEFORM MINI (MCR091030 / 31585342) WAS BEING ADVANCED INTO THE CONCOMITANT ECHELON¿ 10 MICROCATHETER (MEDTRONIC), THE PHYSICIAN ENCOUNTERED RESISTANCE AS THE COIL ENTERED THE MICROCATHETER. THE PHYSICIAN RE-SHEATHED THE COIL, CHECKED TO ENSURE THE INTRODUCER WAS ¿HUBBED¿ CORRECTLY WITHIN THE MICROCATHETER, CONFIRMED THE SYSTEM WAS BEING FLUSHED AND TRIED ADVANCE THE COIL AGAIN. IT WAS REPORTED THAT THE COIL WOULD NOT ADVANCE MORE THAN APPROXIMATELY 5 CM INTO THE MICROCATHETER. EACH COIL WAS PLACED ON THE BACK TABLE FOR INVESTIGATION FOLLOWING THE CASE. THE TWO COILS THAT WERE AVAILABLE FOR RETURN: 1MM X 2CM CEREPAK FREEFORM MINI (MCR091020) AND 1.5MM X 2.5CM CEREPAK FREEFORM MINI (MCR091525) WOULD NOT ADVANCE OUTSIDE OF THE INTRODUCER. NONE OF THE COILS WERE INTRODUCED INTO THE PATIENT¿S BODY. THE INABILITY OF THE COILS TO BE ADVANCED THROUGH THE MICROCATHETER DI

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23444656 ↗
  • MalfunctionJun 18, 2024Separation Failure; Separation Failure

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, EACH OF THE THREE COILS USED A 1.5MM X 4CM CEREPAK FREEFORM MINI (MCR091540 / 31207073), A 1MM X 4CM CEREPAK FREEFORM MINI (MCR091040 / 31045747), AND A 1.5MM X 2.5CM CEREPAK FREEFORM MINI (MCR091525 / 31149187) DID NOT DETACH USING BOTH MANUAL BREAK AND WITH THE DETACHMENT HANDLE. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. ON 14-JUN-2024, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE PROCEDURE WAS EMBOLIZATION OF THE MIDDLE MENINGEAL ARTERY. THE CATALOG AND LOT NUMBER OF THE DETACHMENT HANDLE USED WAS MDH1 / 102109430. THE SAME DETACHMENT HANDLE WAS USED WITH ALL THREE COILS; IT WAS INDICATED AS NOT AVAILABLE FOR RETURN. MANUAL BREAK WAS ATTEMPTED ON THE FIRST COIL IMPLANTED. THE INFORMATION INDICATED THAT ¿THIS IS WHEN THE SYSTEM APPEARED TO BE BROKEN INTO TWO PIECES. THE PULL WIRE WAS NOT ATTACHED. MECHANICAL [DETACHMENT] WAS ATTEMPTED ON THE THIRD COIL (SECOND ATTEMPT WAS SUCCESSFUL) FIRST, THIS WAS UNSUCCESSFUL, WE MOVED TO MANUAL BREAK IMMEDIATELY WITHOUT SUCCESS. WE WRAPPED THE PULL WIRE AROUND FINGER AND PULLED AND THE PULL WIRE APPEARED TO HAVE BROKEN. THE FOUR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19563948 ↗
  • MalfunctionMay 8, 2024Separation Failure; Separation Failure

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING COIL EMBOLIZATION OF A MIDDLE MENINGEAL ARTERY (MMA), TWO COILS, A 1MM X 3CM CEREPAK FREEFORM MINI (MCR091030 / 31177297) AND A 1.5MM X 2.5CM CEREPAK FREEFORM MINI (MCR091525 / 31149187) DID NOT DETACH CORRECTLY. THE CEREPAK TM DETACHMENT HANDLE (MDH1 / 102109430) AS WELL AS THE MANUAL BREAK WERE ATTEMPTS AND NEITHER METHOD WORKED. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19272109 ↗