Skip to content

CEREPAK FREEFORM MCR092535Recalls & reported problems

Cerenovus, Inc. · CEREPAK FREEFORM Mini Detachable Coil Delivery System Length - 190 cm Coil Diameter - 2.5 mm Coil Length - 3.5 cm · Class 2 · Implant

All models of CEREPAK FREEFORM (60)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886704083023), received between Aug 1, 2025 and Aug 1, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Separation Failure” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 2, 2025Class IOpen, Classified

    Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.

    Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 1, 2025Separation Failure; Separation Failure

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, THE 2.5MM X 3.5CM CEREPAK FREEFORM MINI (MCR092535 / 31212048) WAS POSITIONED AND THERE WERE TWO ATTEMPTS TO DETACH IT WITH THE DETACHMENT HANDLE (MDH1 / 102109430) AND ONCE VIA MANUAL BREAK. THE COIL FAILED TO DETACH. THE PROCEDURE WAS REPORTEDLY DELAYED BY 15 MINUTES, BUT IT WAS SUCCESSFULLY COMPLETED WITHOUT ANY NEGATIVE PATIENT IMPACT. ON 31-JUL-2025, ADDITIONAL INFORMATION WAS RECEIVED. THE INFORMATION INDICATED THAT THE PROCEDURE WAS TARGETING A 3.7MM X 3MM X 3MM UNRUPTURED ANEURYSM ON THE ANTERIOR COMMUNICATING ARTERY. WHEN THE COIL WAS REMOVED, IT WAS NOT STRETCHED; IT WAS STILL ATTACHED TO THE DELIVERY SYSTEM. THERE WAS NO EVIDENCE OF BLOOD FLOW INTERRUPTION AS A RESULT OF THE REPORTED ISSUE. THE CONCOMITANT MICROCATHETER WAS A 150CM X 50CM PROWLER 14 MICROCATHETER (606151FX). THE PROCEDURE WAS COMPLETED WITH 3MM X 6CM CEREPAK FREEFORM XTRASOFT (XCR120306) USING THE SAME 150CM X 50CM PROWLER 14 MICROCATHETER AND THE SAME DETACHMENT HANDLE (THE LOT NUMBER PROVIDED WAS 102109430). THE INFORMATION INDICATED THAT THE PHYSICIAN DID CONSIDER THE 15-MINUTE PROCEDURE EXTENSION TO BE SIGNIFICAN

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22688264 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates