CEREPAK FREEFORM MCR092545Recalls & reported problems

Cerenovus, Inc. · CEREPAK FREEFORM Mini Detachable Coil Delivery System Length - 190 cm Coil Diameter -2.5 mm Coil Length - 4.5 cm · Class 2 · Implant

All models of CEREPAK FREEFORM (60)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886704083030), received between Apr 18, 2024 and Apr 18, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Failure to Advance” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 2, 2025Class IOpen, Classified

    Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.

    Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 18, 2024Failure to Advance; Failure to Advance

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A COIL EMBOLIZATION PROCEDURE TARGETING A RUPTURED LEFT MIDDLE CEREBRAL ARTERY (MCA) BIFURCATION ANEURYSM, THE 2.5MM X 4.5CM CEREPAK FREEFORM MINI (MCR092545 / 31201421) FAILED TO ADVANCE MORE THAN 2CM INTO THE CONCOMITANT MICROCATHETER. PRIOR TO THE USE OF THIS COIL, THREE (3) OTHER CEREPAK COILS WERE ADVANCED THROUGH THE SYSTEM AND WERE SUCCESSFULLY DEPLOYED. WHEN THE TECHNOLOGIST FELT RESISTANCE, THE COIL WAS RETRACTED BACK INTO THE INTRODUCER SHEATH AND THEN ADVANCED AGAIN WITH NO SUCCESS. THE COIL WAS REMOVED, AND THE SYSTEM CHECKED FOR SIGNS THAT WOULD HAVE CAUSED THE RESISTANCE. NOTHING WAS NOTED BY THE PHYSICIAN. THE COIL WAS INSPECTED FOR ANY DAMAGE BEFORE BEING RE-SHEATHED AND ATTEMPTED TO USE AGAIN. NO DAMAGE WAS NOTED, AND THE COIL MET THE SAME RESISTANCE WHEN IT WAS ADVANCED FOR THE THIRD TIME. THE 2.5MM X 4.5CM CEREPAK FREEFORM MINI WAS REMOVED AND A 2MM X 6CM CEREPAK FREEFORM MINI WERE USED WITH NO ISSUES. THE PROCEDURE CONTINUED WITH THREE (3) MORE SUCCESSFULLY DEPLOYED COILS THROUGH THE SAME MICROCATHETER. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. ON 11-MAR-2024, ADDITIONAL INFORMATION WAS RECEIVED. PER THE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19137726 ↗

Other Device, Vascular, for Promoting Embolization devices