CEREPAK FREEFORM SCR120306Recalls & reported problems
Cerenovus, Inc. · CEREPAK FREEFORM Detachable Coil Delivery System Length - 190 cm Coil Diameter - 3 mm Coil Length - 6 cm · Class 2 · Implant
All models of CEREPAK FREEFORM (60)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886704083092), received between Nov 6, 2023 and Nov 6, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Separation” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Oct 2, 2025Class IOpen, Classified
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Separation1
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 6, 2023Material Separation; Separation Failure; Material Separation; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT THE 48-YEAR-OLD MALE PATIENT UNDERWENT A PHARYNGEAL-CAROTID ARTERY (CA) VESSEL SACRIFICE PROCEDURE TARGETING THE LEFT COMMON CAROTID ARTERY (LCCA), A 6FR SHEATH -RIGHT FEMORAL ARTERY, 6FR BENCHMARK¿ DISTAL ACCESS CATHETER (PENUMBRA) WITH 5FR VTK SELECT CATHETER (UNSPECIFIED BRAND) WERE INTRODUCED TO THE LCCA ALONG WITH AN EXCELSIOR® XT-27® MICROCATHETER (STRYKER) AND SYNCHRO2® 0.014¿ GUIDEWIRE (STRYKER). A 3MM-5MM MICRO VASCULAR PLUG (MVP) (MEDTRONIC) WAS DEPLOYED IN THE LCCA SEGMENT. THE EXCELSIOR® XT-27® MICROCATHETER WAS REMOVED. THE SL-10® 45° MICROCATHETER (STRYKER) AND THE SYNCHRO 14 GUIDEWIRE AT THE PROXIMAL TO WHERE THE PLUG WAS DEPLOYED WAS REMOVED. THE FOLLOWING CEREPAK FREEFORM COILS WERE PLACED: 4MM X 12CM CEREPAK FREEFORM (SCR120412 / 31140706), 5MM X 15CM CEREPAK FREEFORM (SCR120515 / 31090186), 6MM X 20CM CEREPAK FREEFORM (SCR120620 / 31058141), 6MM X 15CM CEREPAK FREEFORM (SCR120615 / 31136146), 4MM X 10CM CEREPAK FREEFORM (SCR120410 / 31140705), 6MM X 20CM CEREPAK FREEFORM XTRASOFT (XCR120620 / 31131261), 4MM X 12CM CEREPAK FREEFORM (SCR120515 / 31090186), 6MM X 20CM CEREPAK FREEFORM (SCR120620 / 31058141), 6MM X 25CM CEREP
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18080845 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page