Cerenovus, Inc. · CEREPAK FREEFORM Detachable Coil Delivery System Length - 190 cm Coil Diameter - 3 mm Coil Length - 8 cm · Class 2 · Implant

All models of CEREPAK FREEFORM (60)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886704083108), received between Sep 3, 2024 and Sep 3, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Activation Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 2, 2025Class IOpen, Classified

    Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.

    Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Activation Problem1
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 3, 2024Separation Failure; Activation Problem; Separation Failure; Activation Problem

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A STENT-ASSISTED ENDOVASCULAR EMBOLIZATION PROCEDURE TARGETING AN ANTERIOR COMMUNICATING ARTERY (ACOM) ANEURYSM, AFTER ¿JAILING¿ THE SL-10® MICROCATHETER (STRYKER) IN THE ANEURYSM WITH A NEUROFORM ATLAS® STENT SYSTEM (STRYKER), COIL EMBOLIZATION WAS COMMENCED. THE FIRST TWO CEREPAK FREEFORM COILS, A 5MM X 15CM CEREPAK FREEFORM AND A 4MM X 10CM CEREPAK FREEFORM, WERE SUCCESSFULLY DEPLOYED VIA MANUAL BREAK. A THIRD CEREPAK COIL, 3MM X 8CM CEREPAK FREEFORM (SCR120308 / 31149188) WAS POSITIONED AFTER THE FIRST TWO COILS. AFTER SNAPPING THE DETACHMENT ZONE AT THE ENGINEERED FRACTURE POINT AND WIRE WAS EXPOSED, THE PHYSICIAN REPORTED FEELING A LOT OF TENSION COMPARED TO THE FIRST TWO COILS. THE PHYSICIAN WAS ADVISED TO STOP PULLING AND THE SYSTEM WAS ASSESSED TO REDUCE ANY PROXIMAL FRICTION OUTSIDE AND TO REDUCE ANY POSSIBLE FORWARD TENSION ON THE INSIDE, IN THE PATIENT¿S ANATOMY. DETACHMENT WAS RE-ATTEMPTED WITH THE PULL WIRE BREAKING. NO PULL WIRE WAS LEFT TO TRANSLATE AND THE 3MM X 8CM CEREPAK FREEFORM COIL WAS REMOVED. IT WAS REPORTED THAT THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING STRYKER 360 COILS. THE PROCEDURE WAS PROLONGED BY

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20130793 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →