CEREPAK FREEFORM SCR121230Recalls & reported problems
Cerenovus, Inc. · CEREPAK FREEFORM Detachable Coil Delivery System Length - 190 cm Coil Diameter - 12 mm Coil Length - 30 cm · Class 2 · Implant
All models of CEREPAK FREEFORM (60)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704083252), received between Jul 16, 2024 and Nov 15, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 2, 2025Class IOpen, Classified
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Separation Failure2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 15, 2024Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE TARGETING A LEFT POSTERIOR COMMUNICATING ARTERY (PCOM) ANEURYSM, EIGHT (8) COILS WERE IMPLANTED WITHOUT FAIL. TWO COILS FAILED TO DETACH: THE TWO COILS ARE A 12MM X 30CM CEREPAK FREEFORM (SCR121230 / 31173118) AND A 3MM X 6CM CEREPAK FREEFORM MINI (MCR093060 / 31207074). IT WAS REPORTED THAT THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT ANY DELAY. THERE WAS NO NEGATIVE PATIENT IMPACT. ON 30-OCT-2024, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE ANEURYSM WAS AN UNRUPTURED SACCULAR ANEURYSM WITH A SIZE OF 11MM. THE DETACHMENT HANDLE (MDH1 / 102109430) WAS USED ON THE FIRST COIL, THE 12MM X 30CM CEREPAK FREEFORM (SCR121230 / 31173118) FOR ONE DETACHMENT ATTEMPT, WHEN THE COIL DID NOT DETACH, THEY SWITCHED TO MANUAL BREAK. THE 12MM X 30CM CEREPAK FREEFORM COIL STILL DID NOT DETACH. PER THE INFORMATION, MANUAL BREAK WAS THE PRIMARY METHOD OF DETACHMENT FOR ALL REMAINING COILS. THE DETACHMENT HANDLE IS NOT AVAILABLE FOR RETURN. WHEN THE 12MM X 30CM CEREPAK FREEFORM COIL WAS REMOVED, IT WAS STILL ATTACHED TO THE DELIVERY SYSTEM. WHEN THE 3MM X 6CM CEREPAK FREEFORM MINI (MCR093060 /
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20700414 ↗MalfunctionJul 16, 2024Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, ATTEMPTS WERE MADE TO DETACH THE 12MM X 30CM CEREPAK FREEFORM COIL (SCR121230 / 31140716) WITH THE CEREPAK TM DETACHMENT HANDLE (MDH1 / LOT# UNKNOWN), THE COIL WOULD NOT RELEASE. THREE ATTEMPTS TO DETACH WERE MADE WITH THE DETACHMENT HANDLE FOLLOWED BY TWO ATTEMPTS TO DETACH VIA MANUAL BREAK TO NO AVAIL. THE COMPLAINT COIL WAS REMOVED AND ANOTHER COIL WAS USED AND IT ¿DETACHED FLAWLESSLY.¿ THERE WAS NO DELAY IN THE PROCEDURE AS A RESULT OF THE REPORTED ISSUE. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT; THE PROCEDURE WAS SUCCESSFULLY COMPLETED. ON 10-JUL-2024, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION THE PROCEDURE WAS TARGETING A 13 MM X 12 MM X 11MM RUPTURED ANEURYSM ON THE RIGHT POSTERIOR COMMUNICATING (PCOM) ARTERY. THE LOT NUMBER OF THE DETACHMENT HANDLE WAS PROVIDED (102109430); THE HANDLE IS NOT AVAILABLE FOR RETURN. WHEN THE 12MM X 30CM CEREPAK FREEFORM COIL WAS REMOVED, IT WAS NOT STRETCHED AND IT WAS STILL ATTACHED TO THE DELIVERY SYSTEM. THE MICROCATHETER USED WAS AN EXCELSIOR® SL-10® MICROCATHETER (STRYKER). THE REPLACEMENT COIL WAS A 12MM X 40CM CEREPAK UNIFOR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19757566 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page