CEREPAK HELIFORM SHD180312Recalls & reported problems
Cerenovus, Inc. · CEREPAK HELIFORM XL Detachable Coil Delivery System Length - 190 cm Coil Diameter - 3 mm Coil Length - 12 cm · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704083467), received between Mar 10, 2025 and Apr 21, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Oct 2, 2025Open, Classified
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
Z-0434-2026 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Separation Failure2
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 21, 2025Separation Failure; Separation Failure
AS REPORTED BY THE HEALTH CARE PROFESSIONAL, DURING A COIL EMBOLIZATION TO A VERTEBRAL ARTERY FISTULA, THE PHYSICIAN WAS COILING THROUGH A PLATO 017 160CM MICROCATHETER (MC). INSERTED 10 COILS PREVIOUSLY. INSERTED A HELIFORM XL 3MM X 12 CM COIL (SHD180312, 31168339), DURING DETACHMENT, TRIED THE MECHANICAL DETACHMENT HANDLE (MDH1, 102109430) BUT WAS UNSUCCESSFUL SO THE USER TURNED TO MANUAL BREAK METHOD. DURING MANUAL BREAK, PULL WIRE BROKE OFF INSIDE OF THE PUSHER TUBE, SO UNABLE TO GRASP WITH HEMOSTAT TO PULL ON THE PULL WIRE. THE COIL WAS UNABLE TO DETACH. 5 COILS WERE SUBSEQUENTLY IMPLANTED AND DETACHED WITH NO ISSUES. THERE WERE NO PROCEDURAL DELAYS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. ADDITIONAL EVENT INFORMATION RECEIVED ON 01-APR-2025 INDICATED THAT THERE WAS NO RESISTANCE DURING ADVANCEMENT OF THE DEVICE THROUGH THE MICROCATHETER OR POSITIONING DIFFICULTY IN THE ANEURYSM. THERE WERE NO VISUAL DEFORMITIES ON THE COIL. THE VESSEL WAS NOT EXCESSIVELY TORTUOUS OR WITH ACUTE BENDS. IN THE PHYSICIAN'S OPINION, THERE WERE NO FACTORS THAT MAY HAVE CONTRIBUTED TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21872791 ↗MalfunctionMar 10, 2025Stretched; Separation Failure; Stretched; Separation Failure
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION TO THE MIDDLE CEREBRAL ARTERY (MCA) BIFURCATION, A HELIFORM XL 3MM X 12 CM COIL (SHD180312, 31168339) DID NOT DETACH WITH THE MECHANICAL DETACHMENT HANDLE (MDH1, UNKNOWN LOT), IT ALSO DID NOT DETACH MANUALLY. THE DOCTOR RETRIEVED THE COIL WITH NO PROBLEMS. THE SURGERY WAS NOT DELAYED DUE TO THE REPORTED EVENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. ADDITIONAL EVENT INFORMATION RECEIVED ON 26-FEB-2025 INDICATED THAT THREE COILS WERE PREVIOUSLY IMPLANTED DURING PROCEDURE. THERE WAS NO RESISTANCE DURING ADVANCEMENT OF THE DEVICE THROUGH THE MICROCATHETER OR POSITIONING DIFFICULTY IN THE ANEURYSM. THREE ATTEMPTS WERE MADE WITH THE DETACHMENT HANDLE. IT WAS NOTED THAT THE PULL WIRE OF THE DEVICE WAS STRETCHED. THERE VESSEL WAS NOT EXCESSIVELY TORTUOUS OR WITH ACUTE BENDS. ADDITIONAL EVENT INFORMATION RECEIVED ON 27-FEB-2025 CONFIRMED THAT THE COIL WAS STRETCHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21561235 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page