Skip to content

CEREPAK FREEFORM XCR120304Recalls & reported problems

Cerenovus, Inc. · CEREPAK FREEFORM XtraSoft Detachable Coil Delivery System Length - 190 cm Coil Diameter - 3 mm Coil Length - 4 cm · Class 2 · Implant

All models of CEREPAK FREEFORM (60)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886704083634), received between Jun 26, 2023 and Feb 13, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Separation Failure” (6 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 4 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 2, 2025Class IOpen, Classified

    Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.

    Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Separation Failure6
  • Difficult or Delayed Separation1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2024Separation Failure; Separation Failure

    AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION, TWO FREEFORM XSFT 3MM X 4 CM COIL (XCR120304, 31131259) AND A FREEFORM XSFT 4MM X 6 CM COIL (XCR120406, UNKNOWN LOT) WOULD NOT DETACH. THREE DETACHMENTS WERE ATTEMPTED MANUAL BREAK. NO COILS WOULD DETACH. NO PATIENT INJURY WAS REPORTED. ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2024 INDICATED THAT NO COILS WERE IMPLANTED PRIOR TO THE ATTEMPT OF THE CEREPAK COILS. THERE WAS NO DIFFICULTY ADVANCING THE COILS OR POSITIONING WITHIN THE ANEURYSM. NO ATTEMPTS WERE MADE USING THE DETACHMENT HANDLE. THERE WAS NO DAMAGE TO THE EMBOLIC COILS OR ANY DEVICE COMPONENT. THE TARGET VESSEL WAS THE LEFT INTERNAL CAROTID ARTERY (LICA) BIFURCATION WITH A TYPE 3 ARCH AND VERY TORTUOUS CAROTID. THE DELAY WAS ONLY SLIGHT AND NOT SIGNIFICANT ENOUGH FOR THE DOCTOR TO COMMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18698233 ↗
  • MalfunctionFeb 2, 2024Separation Failure; Separation Failure

    AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION, FREEFORM XSFT 3MM X 4 CM COIL (XCR120304, 31131259) AND A FREEFORM XSFT 4MM X 6 CM COIL (XCR120406, UNKNOWN LOT) WOULD NOT DETACH. THREE DETACHMENTS WERE ATTEMPTED MANUAL BREAK. NO COILS WOULD DETACH. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18634735 ↗
  • MalfunctionFeb 1, 2024Separation Failure; Separation Failure

    AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION TO THE ANTERIOR COMMUNICATING ARTERY (ACCOM), TWO FREEFORM XSFT 4MM X 8 CM COILS (XCR120408, 31126838), A FREEFORM XSFT 3MM X 6 CM COIL (XCR120306, 31136147) AND A TWO FREEFORM XSFT 3MM X 4 CM COILS (XCR120304, LOT UNKNOW) AND (XCR120304, 31131259) WERE ADVANCED INTO ANEURYSM AND MANUAL DETACHMENT WAS ATTEMPTED WITH NO SUCCESS. THE COILS WERE REMOVED FROM THE PATIENT. THE FREEFORM XSFT 3MM X 4 CM COILS WERE PLACED INTO THE ANEURYSM AND DETACHMENT WAS ATTEMPTED WITH A MECHANICAL DETACHMENT HANDLE (MDH1) WITH NO SUCCESS. THE COILS WERE REMOVED FROM THE PATIENT. ADDITIONAL EVENT INFORMATION WAS RECEIVED ON 22-JAN-2024 INDICATED THAT A STRYKER SL-10 MICROCATHETER WAS USED. THREE COILS WERE SUCCESSFULLY DETACHED WITH MANUAL DETACHMENT PRIOR TO THE FAILURE, THEN TWO FAILURES WITH 4MM X 8 CM COIL, THEN A SUCCESSFUL DETACHMENT OF TWO 3MM X 6 CM, THEN A FAILED 3MM X 6 CM , THEN TWO MORE FAILED WITH 3MM X 4 CM WITH DETACHMENT HANDLE. THERE WAS NO RESISTANCE DURING ADVANCEMENT OF THE DEVICE THROUGH THE MICROCATHETER OR POSITIONING DIFFICULTY IN THE ANEURYSM. THERE WAS NO DAMAGE TO THE EMBOLIC COILS OR ANY DEVICE COMPONENT. THERE WAS MILD TO

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18625689 ↗
  • MalfunctionFeb 1, 2024Separation Failure; Separation Failure

    AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION TO THE ANTERIOR COMMUNICATING ARTERY (ACCOM), TWO FREEFORM XSFT 4MM X 8 CM COILS (XCR120408, 31126838), A FREEFORM XSFT 3MM X 6 CM COIL (XCR120306, 31136147) AND A TWO FREEFORM XSFT 3MM X 4 CM COILS (XCR120304, LOT UNKNOW) AND (XCR120304, 31131259) WERE ADVANCED INTO ANEURYSM AND MANUAL DETACHMENT WAS ATTEMPTED WITH NO SUCCESS. THE COILS WERE REMOVED FROM THE PATIENT. THE FREEFORM XSFT 3MM X 4 CM COILS WERE PLACED INTO THE ANEURYSM AND DETACHMENT WAS ATTEMPTED WITH A MECHANICAL DETACHMENT HANDLE (MDH1) WITH NO SUCCESS. THE COILS WERE REMOVED FROM THE PATIENT. ADDITIONAL EVENT INFORMATION WAS RECEIVED ON 22-JAN-2024 INDICATED THAT A STRYKER SL-10 MICROCATHETER WAS USED. THREE COILS WERE SUCCESSFULLY DETACHED WITH MANUAL DETACHMENT PRIOR TO THE FAILURE, THEN TWO FAILURES WITH 4MM X 8 CM COIL, THEN A SUCCESSFUL DETACHMENT OF TWO 3MM X 6 CM, THEN A FAILED 3MM X 6 CM , THEN TWO MORE FAILED WITH 3MM X 4 CM WITH DETACHMENT HANDLE. THERE WAS NO RESISTANCE DURING ADVANCEMENT OF THE DEVICE THROUGH THE MICROCATHETER OR POSITIONING DIFFICULTY IN THE ANEURYSM. THERE WAS NO DAMAGE TO THE EMBOLIC COILS OR ANY DEVICE COMPONENT. THERE WAS MILD TO

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18625679 ↗
  • MalfunctionNov 6, 2023Separation Failure; Separation Failure

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A LEFT INTERNAL MAXILLARY ARTERY (IMAX) EMBOLIZATION WITH PARTICLES AND COILS TO SACRIFICE VESSEL; TORTUOSITY IS HIGH, A HEADWAY® 21 MICROCATHETER (MICROVENTION) IS USED. THE SECOND COIL USED WAS A 3MM X 4CM CEREPAK FREEFORM (XCR120304 / 31131259). THE COIL ADVANCED THROUGH THE MICROCATHETER WITHOUT ISSUE. AFTER THREE (3) ATTEMPTS WITH THE HANDLE, A CEREPAKTM DETACHER (MDH1 / 102109430) WITHOUT SUCCESS, THE MANUAL BREAK WAS ATTEMPTED AS SUGGESTED BY THE INSTRUCTIONS FOR USE (IFU) AND IT WAS UNSUCCESSFUL. THE COIL WAS REMOVED FROM THE VESSEL INTACT. IT WAS REPORTED THAT A THIRD COIL, A 3MM X 6CM CEREPAK FREEFORM MINI (MCR093060 / 31136064) WAS USED AND IT ADVANCED THROUGH THE MICROCATHETER WITHOUT ISSUE. AFTER THREE (3) UNSUCCESSFUL ATTEMPTS WITH THE HANDLE, THE MANUAL BREAK WAS ATTEMPTED AND THE COIL DID DETACH IN THE VESSEL. THE COIL IS BEING RETURNED BECAUSE THE HANDLE ATTEMPTS WERE WITHOUT SUCCESS. THE HANDLE WAS NOT RETAINED FOR ANALYSIS. THERE WAS NO NEGATIVE IMPACT TO THE PATIENT. THE REPORTED ISSUES RESULTED IN MINIMAL DELAY IN TREATMENT. ON 31-OCT-2023, ADDITIONAL INFORMATION WAS RECEIVED FROM THE CLINICAL ACCOUNT SPECIALIST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18080942 ↗
  • MalfunctionNov 6, 2023Material Separation; Difficult or Delayed Separation; Material Separation; Difficult or Delayed Separation

    THE HEALTHCARE PROFESSIONAL REPORTED THAT THE 48-YEAR-OLD MALE PATIENT UNDERWENT A PHARYNGEAL-CAROTID ARTERY (CA) VESSEL SACRIFICE PROCEDURE TARGETING THE LEFT COMMON CAROTID ARTERY (LCCA), A 6FR SHEATH -RIGHT FEMORAL ARTERY, 6FR BENCHMARK¿ DISTAL ACCESS CATHETER (PENUMBRA) WITH 5FR VTK SELECT CATHETER (UNSPECIFIED BRAND) WERE INTRODUCED TO THE LCCA ALONG WITH AN EXCELSIOR® XT-27® MICROCATHETER (STRYKER) AND SYNCHRO2® 0.014¿ GUIDEWIRE (STRYKER). A 3MM-5MM MICRO VASCULAR PLUG (MVP) (MEDTRONIC) WAS DEPLOYED IN THE LCCA SEGMENT. THE EXCELSIOR® XT-27® MICROCATHETER WAS REMOVED. THE SL-10® 45° MICROCATHETER (STRYKER) AND THE SYNCHRO 14 GUIDEWIRE AT THE PROXIMAL TO WHERE THE PLUG WAS DEPLOYED WAS REMOVED. THE FOLLOWING CEREPAK FREEFORM COILS WERE PLACED: 4MM X 12CM CEREPAK FREEFORM (SCR120412 / 31140706), 5MM X 15CM CEREPAK FREEFORM (SCR120515 / 31090186), 6MM X 20CM CEREPAK FREEFORM (SCR120620 / 31058141), 6MM X 15CM CEREPAK FREEFORM (SCR120615 / 31136146), 4MM X 10CM CEREPAK FREEFORM (SCR120410 / 31140705), 6MM X 20CM CEREPAK FREEFORM XTRASOFT (XCR120620 / 31131261), 4MM X 12CM CEREPAK FREEFORM (SCR120515 / 31090186), 6MM X 20CM CEREPAK FREEFORM (SCR120620 / 31058141), 6MM X 25CM CEREP

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18080848 ↗
  • MalfunctionJun 26, 2023Separation Failure; Separation Failure

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, THIS WAS A RECANALIZATION TO THE BASILAR TIP ANEURYSM WITH STENT ALREADY IN PLACE. THE ISSUE HAPPENED WHEN THE FIRST COIL, A FREEFORM XSFT 3MM X 4 CM (XCR120304, 30982796) WOULD NOT DETACH USING THE HANDLE TWICE AND THE MECHANICAL DETACHMENT HANDLE (MDH1, 97988373) WAS ALSO UNSUCCESSFUL. THE COIL WAS ADVANCED THROUGH THE SHEATH AND THE HEADWAY DUO 156 CM MICROCATHETER (MC) WITH MINIMAL FRICTION. NO KINKING OF WIRE OF COIL. AFTER THE FIRST ATTEMPT OF THE HANDLE, THE COIL WAS ADVANCED A FEW MILLIMETERS FURTHER AND ATTEMPTED AGAIN WITH THE HANDLE. THE MANUAL DETACHMENT WAS ATTEMPTED AS SUGGESTED BY THE INSTRUCTIONS FOR USE (IFU) ALSO UNSUCCESSFUL. THE SYSTEM PULLED OUT COMPLETELY THROUGH MANUAL DETACHMENT ZONE. THE COIL WAS TAKEN OUT OF THE ANEURYSM. THE COIL APPEARS TO BE INTACT AND NOT STRETCHED. STRYKER COILS WERE USED IN THIS CASE AFTER INCIDENT. NO IMPACT TO PATIENT AND MINIMAL TIME DELAYS DUE TO THIS DETACHMENT ISSUE. ADDITIONAL INFORMATION RECEIVED INDICATED THAT THERE WAS MINIMAL DIFFICULTY POSITIONING THROUGH THE STENT. THERE WAS MODERATE VESSEL TORTUOSITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17199644 ↗

FDA clearance and classification

Cleared through 510(k) K220040 on Jul 14, 2022 (decision: Substantially Equivalent). Submitted by Medos International SARL as “CEREPAK Uniform, Uniform XL, Uniform 3D, Heliform Soft, Heliform XtraSoft, Heliform XL, Heliform XtraSoft XL, Freeform, Freeform Mini, and Freeform XtraSoft Detachable Coil Systems”. The FDA review took 190 days. FDA record ↗

The FDA regulates product code KRD (device, vascular, for promoting embolization) under 21 CFR 870.3300 as a Class II device. 21 CFR 870.3300 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the CEREPAK FREEFORM XCR120304

Has the CEREPAK FREEFORM XCR120304 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0434-2026 (Class I), began on Oct 2, 2025: Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the CEREPAK FREEFORM XCR120304?

The FDA's MAUDE database lists 7 adverse event reports for this device, received between Jun 26, 2023 and Feb 13, 2024. In those reports, the most common specific device problems are “Separation Failure” (6), “Difficult or Delayed Separation” (1) and “Material Separation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CEREPAK FREEFORM XCR120304?

None of the reports are classified as deaths, 0 as injuries and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CEREPAK FREEFORM XCR120304 FDA approved?

It was cleared through 510(k) K220040 on Jul 14, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the CEREPAK FREEFORM XCR120304 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the CEREPAK FREEFORM XCR120304, and is it still sold?

The FDA's device database lists Cerenovus, Inc. as the company responsible for it (brand CEREPAK FREEFORM, model XCR120304). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page