CEREPAK FREEFORM XCR120306Recalls & reported problems
Cerenovus, Inc. · CEREPAK FREEFORM XtraSoft Detachable Coil Delivery System Length - 190 cm Coil Diameter - 3 mm Coil Length - 6 cm · Class 2 · Implant
All models of CEREPAK FREEFORM (60)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886704083641), received between Feb 1, 2024 and Mar 6, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Oct 2, 2025Class IOpen, Classified
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Separation Failure4
- Material Separation2
- Premature Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 6, 2025Material Separation; Material Separation
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION, THE HYPO TUBE OF A FREEFORM XSFT 3MM X 6 CM COIL (XCR120306, 31445669) SLIPPED OFF THE BACK WHEN MANUAL DETACHMENT HAPPENED. COIL STILL DETACHED. THERE WAS NO PATIENT INJURY REPORTED. ADDITIONAL EVENT INFORMATION RECEIVED ON 04-MAR-2025 INDICATING THAT THE UNRUPTURED ANEURYSM WAS AT MIDDLE CEREBRAL ARTERY, M1 SEGMENT. THE DEVICE WAS NOT INADVERTENTLY SUBJECTED TO FORCE DURING THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21537471 ↗MalfunctionDec 26, 2024Material Separation; Material Separation
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, THE 3MM X 6CM CEREPAK FREEFORM XTRASOFT (XCR120306 / 31092859) DETACHED ¿FLAWLESSLY.¿ IT WAS THEN REPORTED THAT ¿IT SEEMED LIKE THE COIL POSITIONING MARKER WAS LEFT BEHIND.¿ THE MICROCATHETER (UNSPECIFIED BRAND) WAS REMOVED AND REPLACED. THERE WAS NO NEGATIVE PATIENT IMPACT. THE PROCEDURE WAS REPORTED AS SUCCESSFULLY COMPLETED WITHOUT ANY DELAY. ON 18-DEC-2024, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE TARGET OF THE PROCEDURE WAS THE INTERNAL CAROTID ARTERY (ICA). RELATED TO THE POSITIONING MARKER THAT REPORTEDLY WAS LEFT BEHIND, THE INFORMATION STATES, ¿IT SEEMED TO BE ABOUT 3-4 INCHES OF "WIRE" WHEN SPEAKING TO THE TECH. NEAR WHERE THE POSITIONING MARKERS WOULD BE.¿ THE MARKER WAS IN THE MICROCATHETER AND THE MICROCATHETER WAS REMOVED AND REPLACED. THE CONCOMITANT MICROCATHETER USED WAS AN SL-10® MICROCATHETER (STRYKER); THE REPLACEMENT WAS ANOTHER SL-10® MICROCATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21019970 ↗MalfunctionMay 6, 2024Separation Failure; Premature Separation; Separation Failure; Premature Separation
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION TO THE POSTERIOR INFERIOR CEREBELLAR ARTERY (PICA), A FREEFORM XSFT 3MM X 6 CM COIL (XCR120306, 31111679) DID NOT DETACHED PROPERLY. REMOVED FROM ANEURYSM AND THE COIL DETACHED INSIDE THE MARKSMAN 160CM MICROCATHETER (MC). THE MICROCATHETER AND COIL WERE REMOVED FROM THE PATIENT¿S BODY. THE SURGERY WAS DELAYED DUE TO THE REPORTED EVENT BY 20 MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO FRAGMENTS WERE GENERATED. ADDITIONAL EVENT INFORMATION WAS RECEIVED ON 02-MAY-2024 INDICATING THAT ONE COIL HAD BEEN PREVIOUSLY IMPLANTED. THERE WAS NO RESISTANCE DURING ADVANCEMENT OF THE DEVICE THROUGH THE MICROCATHETER OR POSITIONING DIFFICULTY IN THE ANEURYSM. TWO ATTEMPTS WERE MADE USING THE MECHANICAL DETACHMENT HANDLE ( MDH1, LOT UNKNOWN). THE VESSEL WAS EXCESSIVELY TORTUOUS OR WITH ACUTE BENDS. THE PHYSICIAN DID NOT FEEL THAT THE REPORTED 20 MINUTES PROCEDURAL DELAY WAS CLINICALLY SIGNIFICANT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19248551 ↗MalfunctionApr 1, 2024Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION TARGETING AN ANTERIOR COMMUNICATING ARTERY (ACOM) ANEURYSM WITH NORMAL VESSEL, STRAIGHT WITH NO BENDS, THE PHYSICIAN PLACED AND DETACHED VIA A DETACHMENT HANDLE (MDH1 / LOT# 101456956) SEVERAL CEREPAK COILS VIA AN SL-10® MICROCATHETER (STRYKER) WITH FLUSH LINE. WHEN THE PHYSICIAN ATTEMPTED TO DEPLOY TWO (2) 3MM X 6CM CEREPAK FREEFORM XTRASOFT COILS (XCR120306) FROM THE SAME LOT (31092859) THE COILS FAILED TO DETACH. THE PHYSICIAN REMOVED COILS AND CONTINUED TO PLACE SIX (6) MORE COILS. THERE WAS NO NEGATIVE PATIENT IMPACT. ON 21-MAR-2024, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE ANEURYSM WAS AN OBLONG 12MM X 9MM IN SIZE. THE LOT NUMBER OF THE DETACHMENT HANDLE WAS 101456956. WHEN THE 3MM X 6CM CEREPAK FREEFORM XTRASOFT WERE REMOVED, THEY WERE NOT STRETCHED; THE COIL WERE BOTH STILL ATTACHED TO THEIR RESPECTIVE DELIVERY SYSTEMS. THE TWO COILS WERE REMOVED WITHOUT ISSUES. THE SUBSEQUENT SIX COILS WERE PLACED VIA THE SAME SL-10 MICROCATHETER; THE ADDITIONAL SIX COILS WERE NOT CEREPAK COILS, THEY WERE STRYKER COILS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT ANY CLINICALLY SIGNIFICA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19016495 ↗MalfunctionApr 1, 2024Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION TARGETING AN ANTERIOR COMMUNICATING ARTERY (ACOM) ANEURYSM WITH NORMAL VESSEL, STRAIGHT WITH NO BENDS, THE PHYSICIAN PLACED AND DETACHED VIA A DETACHMENT HANDLE (MDH1 / LOT# 101456956) SEVERAL CEREPAK COILS VIA AN SL-10® MICROCATHETER (STRYKER) WITH FLUSH LINE. WHEN THE PHYSICIAN ATTEMPTED TO DEPLOY TWO (2) 3MM X 6CM CEREPAK FREEFORM XTRASOFT COILS (XCR120306) FROM THE SAME LOT (31092859) THE COILS FAILED TO DETACH. THE PHYSICIAN REMOVED COILS AND CONTINUED TO PLACE SIX (6) MORE COILS. THERE WAS NO NEGATIVE PATIENT IMPACT. ON 21-MAR-2024, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE ANEURYSM WAS AN OBLONG 12MM X 9MM IN SIZE. THE LOT NUMBER OF THE DETACHMENT HANDLE WAS 101456956. WHEN THE 3MM X 6CM CEREPAK FREEFORM XTRASOFT WERE REMOVED, THEY WERE NOT STRETCHED; THE COIL WERE BOTH STILL ATTACHED TO THEIR RESPECTIVE DELIVERY SYSTEMS. THE TWO COILS WERE REMOVED WITHOUT ISSUES. THE SUBSEQUENT SIX COILS WERE PLACED VIA THE SAME SL-10 MICROCATHETER; THE ADDITIONAL SIX COILS WERE NOT CEREPAK COILS, THEY WERE STRYKER COILS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT ANY CLINICALLY SIGNIFICA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19016493 ↗MalfunctionFeb 1, 2024Separation Failure; Separation Failure
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION TO THE ANTERIOR COMMUNICATING ARTERY (ACCOM), TWO FREEFORM XSFT 4MM X 8 CM COILS (XCR120408, 31126838), A FREEFORM XSFT 3MM X 6 CM COIL (XCR120306, 31136147) AND A TWO FREEFORM XSFT 3MM X 4 CM COILS (XCR120304, LOT UNKNOW) AND (XCR120304, 31131259) WERE ADVANCED INTO ANEURYSM AND MANUAL DETACHMENT WAS ATTEMPTED WITH NO SUCCESS. THE COILS WERE REMOVED FROM THE PATIENT. THE FREEFORM XSFT 3MM X 4 CM COILS WERE PLACED INTO THE ANEURYSM AND DETACHMENT WAS ATTEMPTED WITH A MECHANICAL DETACHMENT HANDLE (MDH1) WITH NO SUCCESS. THE COILS WERE REMOVED FROM THE PATIENT. ADDITIONAL EVENT INFORMATION WAS RECEIVED ON 22-JAN-2024 INDICATED THAT A STRYKER SL-10 MICROCATHETER WAS USED. THREE COILS WERE SUCCESSFULLY DETACHED WITH MANUAL DETACHMENT PRIOR TO THE FAILURE, THEN TWO FAILURES WITH 4MM X 8 CM COIL, THEN A SUCCESSFUL DETACHMENT OF TWO 3MM X 6 CM, THEN A FAILED 3MM X 6 CM , THEN TWO MORE FAILED WITH 3MM X 4 CM WITH DETACHMENT HANDLE. THERE WAS NO RESISTANCE DURING ADVANCEMENT OF THE DEVICE THROUGH THE MICROCATHETER OR POSITIONING DIFFICULTY IN THE ANEURYSM. THERE WAS NO DAMAGE TO THE EMBOLIC COILS OR ANY DEVICE COMPONENT. THERE WAS MILD TO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18625666 ↗
FDA clearance and classification
Cleared through 510(k) K220040 on Jul 14, 2022 (decision: Substantially Equivalent). Submitted by Medos International SARL as “CEREPAK Uniform, Uniform XL, Uniform 3D, Heliform Soft, Heliform XtraSoft, Heliform XL, Heliform XtraSoft XL, Freeform, Freeform Mini, and Freeform XtraSoft Detachable Coil Systems”. The FDA review took 190 days. FDA record ↗
The FDA regulates product code KRD (device, vascular, for promoting embolization) under 21 CFR 870.3300 as a Class II device. 21 CFR 870.3300 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the CEREPAK FREEFORM XCR120306
Has the CEREPAK FREEFORM XCR120306 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0434-2026 (Class I), began on Oct 2, 2025: Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the CEREPAK FREEFORM XCR120306?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Feb 1, 2024 and Mar 6, 2025. In those reports, the most common specific device problems are “Separation Failure” (4), “Material Separation” (2) and “Premature Separation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CEREPAK FREEFORM XCR120306?
None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the CEREPAK FREEFORM XCR120306 FDA approved?
It was cleared through 510(k) K220040 on Jul 14, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the CEREPAK FREEFORM XCR120306 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the CEREPAK FREEFORM XCR120306, and is it still sold?
The FDA's device database lists Cerenovus, Inc. as the company responsible for it (brand CEREPAK FREEFORM, model XCR120306). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization117 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page