CEREPAK FREEFORM XCR120408Recalls & reported problems
Cerenovus, Inc. · CEREPAK FREEFORM XtraSoft Detachable Coil Delivery System Length - 190 cm Coil Diameter - 4 mm Coil Length - 8 cm · Class 2 · Implant
All models of CEREPAK FREEFORM (60)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886704083672), received between Dec 12, 2023 and Sep 5, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 2, 2025Class IOpen, Classified
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Separation Failure4
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 5, 2024Separation Failure; Separation Failure
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION, CEREPAK COILS WERE BEING USED TO EMBOLIZE A VENUS FISTULA FEEDER. USING A CEREBASE GUIDE SHEATH, VECTA 46 CATHETER AND A SCEPTER XC BALLOON CATHETER TO INSERT COILS. WHEN USING THE MECHANICAL DETACHMENT HANDLE FOR A FREEFORM XSFT 3MM X 8 CM COIL (XCR120308, 31112709), THE HANDLE WAS MOVED TOO FAR LATERALLY/TOWARDS THE PHYSICIAN WHICH CAUSED THE MANUAL BREAK POINT AND PULL WIRE TO FRACTURE. THE PHYSICIAN BELIEVED IT WAS USER ERROR FOR MOVING THE PUSHER TUB/HANDLE. THE COIL WAS THEN REMOVED. NO DELAY OR PATIENT HARM OCCURRED. A FREEFORM XSFT 4MM X 8 CM COIL (XCR120408, LOT TO BE CONFIRMED) WAS PLACED BUT FAILED TO DETACH WITH MANUAL BREAK. THE MANUAL BREAK WAS INITIATED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). AFTER INITIAL BREAK AND PULL, THE COIL DIDN¿T DETACH. THE PHYSICIAN THEN PULLED A SECOND TIME, PULLING MANY CM OF WIRE. THE COIL STILL DIDN¿T DETACH. THE COIL WAS THEN REMOVED. NO PATIENT HARM OCCURRED AND NO SIGNIFICANT DELAYED OCCURRED. ADDITIONAL EVENT INFORMATION WAS RECEIVED ON 19-AUG-2024 INDICATING THAT THE FREEFORM XSFT 3MM X 8 CM WAS THE FIRST COIL USED. WHEN THE FREEFORM XSFT 4MM X 8 CM COIL WAS USED, THERE HA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20144072 ↗MalfunctionFeb 1, 2024Separation Failure; Separation Failure
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION TO THE ANTERIOR COMMUNICATING ARTERY (ACCOM), TWO FREEFORM XSFT 4MM X 8 CM COILS ((B)(6)), A FREEFORM XSFT 3MM X 6 CM COIL ((B)(6)) AND A TWO FREEFORM XSFT 3MM X 4 CM COILS ((B)(6), LOT UNKNOW) AND ((B)(6)) WERE ADVANCED INTO ANEURYSM AND MANUAL DETACHMENT WAS ATTEMPTED WITH NO SUCCESS. THE COILS WERE REMOVED FROM THE PATIENT. THE FREEFORM XSFT 3MM X 4 CM COILS WERE PLACED INTO THE ANEURYSM AND DETACHMENT WAS ATTEMPTED WITH A MECHANICAL DETACHMENT HANDLE (MDH1) WITH NO SUCCESS. THE COILS WERE REMOVED FROM THE PATIENT. ADDITIONAL EVENT INFORMATION WAS RECEIVED ON 22-JAN-2024 INDICATED THAT A STRYKER SL-10 MICROCATHETER WAS USED. THREE COILS WERE SUCCESSFULLY DETACHED WITH MANUAL DETACHMENT PRIOR TO THE FAILURE, THEN TWO FAILURES WITH 4MM X 8 CM COIL, THEN A SUCCESSFUL DETACHMENT OF TWO 3MM X 6 CM, THEN A FAILED 3MM X 6 CM , THEN TWO MORE FAILED WITH 3MM X 4 CM WITH DETACHMENT HANDLE. THERE WAS NO RESISTANCE DURING ADVANCEMENT OF THE DEVICE THROUGH THE MICROCATHETER OR POSITIONING DIFFICULTY IN THE ANEURYSM. THERE WAS NO DAMAGE TO THE EMBOLIC COILS OR ANY DEVICE COMPONENT. THERE WAS MILD TORTUOSITY. THE EVENT DID NOT RESULT IN ANY
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18625806 ↗MalfunctionFeb 1, 2024Separation Failure; Separation Failure
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION TO THE ANTERIOR COMMUNICATING ARTERY (ACCOM), TWO FREEFORM XSFT 4MM X 8 CM COILS (XCR120408, 31126838), A FREEFORM XSFT 3MM X 6 CM COIL (XCR120306, 31136147) AND A TWO FREEFORM XSFT 3MM X 4 CM COILS (XCR120304, LOT UNKNOW) AND (XCR120304, 31131259) WERE ADVANCED INTO ANEURYSM AND MANUAL DETACHMENT WAS ATTEMPTED WITH NO SUCCESS. THE COILS WERE REMOVED FROM THE PATIENT. THE FREEFORM XSFT 3MM X 4 CM COILS WERE PLACED INTO THE ANEURYSM AND DETACHMENT WAS ATTEMPTED WITH A MECHANICAL DETACHMENT HANDLE (MDH1) WITH NO SUCCESS. THE COILS WERE REMOVED FROM THE PATIENT. ADDITIONAL EVENT INFORMATION WAS RECEIVED ON 22-JAN-2024 INDICATED THAT A STRYKER SL-10 MICROCATHETER WAS USED. THREE COILS WERE SUCCESSFULLY DETACHED WITH MANUAL DETACHMENT PRIOR TO THE FAILURE, THEN TWO FAILURES WITH 4MM X 8 CM COIL, THEN A SUCCESSFUL DETACHMENT OF TWO 3MM X 6 CM, THEN A FAILED 3MM X 6 CM , THEN TWO MORE FAILED WITH 3MM X 4 CM WITH DETACHMENT HANDLE. THERE WAS NO RESISTANCE DURING ADVANCEMENT OF THE DEVICE THROUGH THE MICROCATHETER OR POSITIONING DIFFICULTY IN THE ANEURYSM. THERE WAS NO DAMAGE TO THE EMBOLIC COILS OR ANY DEVICE COMPONENT. THERE WAS MILD TO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18625647 ↗MalfunctionDec 12, 2023Break; Separation Failure; Break; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE TARGETING AN INTERNAL CAROTID ARTERY (ICA)-OPHTHALMIC SEGMENT SIDE WALL ANEURYSM, THE FIRST COIL USED WAS A 4MM X 8CM CEREPAK FREEFORM XTRASOFT (XCR120408 / 31126838). IT WAS SUCCESSFULLY NAVIGATED AND INSERTED INTO THE ANEURYSM. DURING MANUAL DETACHMENT, THE PULL WIRE WAS BROKEN IN HALF. THE PHYSICIAN ADMITTED THAT IT WAS USER ERROR BECAUSE THE DISTAL PORTION OF THE HYPOTUBE WAS NOT ¿IN-LINE¿ WITH THE PROXIMAL PORTION WHEN PULLING TO DETACH THE COIL. THE DISTAL PORTION OF THE HYPOTUBE WAS AT AN ANGLE WHEN THE PULLING OCCURRED ¿ WHICH CAUSED THE PULL WIRE TO SNAP. THE COIL WAS SUCCESSFULLY REMOVED. TWO ADDITIONAL COILS WERE THEN PLACED AND SUCCESSFULLY DETACHED. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. ON 04-DEC-2023, ADDITIONAL INFORMATION WAS RECEIVED. THE INFORMATION INDICATED THAT NO DETACHMENT HANDLE WAS USED; MANUAL BREAK WAS USED. THE MICROCATHETER USED WAS SL-10® MICROCATHETER (STRYKER); THE SAME MICROCATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO SIGNIFICANT DELAY AS A RESULT OF THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18315597 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page