CEREPAK FREEFORM XCR120510Recalls & reported problems
Cerenovus, Inc. · CEREPAK FREEFORM XtraSoft Detachable Coil Delivery System Length - 190 cm Coil Diameter - 5 mm Coil Length - 10 cm · Class 2 · Implant
All models of CEREPAK FREEFORM (60)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886704083696), received between Jan 2, 2024 and Jun 24, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 2, 2025Class IOpen, Classified
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Twisted/Bent1
- Positioning Problem1
- Premature Separation1
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 24, 2025Material Twisted/Bent; Positioning Problem; Material Twisted/Bent; Positioning Problem
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A SUBARACHNOID HEMORRHAGE (SAH) CASE WHERE THE 40-YEAR-OLD FEMALE PATIENT WHO WAS SUFFERING FROM A MYCOTIC LEFT SUPERIOR CEREBELLAR ARTERY (SCA) ANEURYSM UNDERWENT ENDOVASCULAR EMBOLIZATION PROCEDURE USING CEREPAK COILS. RIGHT FEMORAL ACCESS WAS MADE AND A 6FR. SHORT SHEATH, BENCHMARK¿ GUIDE CATHETER (PENUMBRA), 5 FR. VERT CATHETER (MERIT MEDICAL) WAS INTRODUCED TO THE RIGHT VERTEBRAL ARTERY (RVA) WITH THE BENCHMARK IN THE RVA. THE 5 FR. VERT CATHETER WAS REMOVED. THE SL-10® 45° MICROCATHETER (STRYKER) AND SYNCHRO SOFT GUIDEWIRE (STRYKER) WERE ADVANCED TO THE LEFT SUPERIOR CEREBELLA ARTERY WHERE THE PHYSICIAN HAD TROUGH ACCESSING AND AS A RESULT, CHANGED THE SYNCHRO SOFT GUIDEWIRE TO THE ARISTOTLE® 14 GUIDEWIRE (SCIENTIA VASCULAR). THEN THE COMPLAINT COIL, A 5MM X 10CM CEREPAK FREEFORM XSFT (XCR120510 / 31154786) WAS ADVANCED. THE MICROCATHETER LOST ACCESS AND THE COIL WAS REMOVED TO RESHEATH, AND DURING THE RESHEATHING, IT KINKED TWICE. THE COIL WAS NOT USED. THE ANEURYSM WAS RE-ACCESSED AND FIVE (5) CEREPAK XSFT FREEFORM WERE SUCCESSFULLY USED. ON 23-JAN-2025, ADDITIONAL INFORMATION WAS RECEIVED. THE ADDITIONAL INFORMATION CLARIFIE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22323345 ↗MalfunctionJun 11, 2025Premature Separation; Premature Separation
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A PROCEDURE ON (B)(6) 2025 TO TREAT A FISTULA, THE WIRE ON THE 5MM X 10CM CEREPAK FREEFORM XSFT (XCR120510 / 31146337) SNAPPED WHILE THE DEVICE WAS BEING FED INTO THE MICROCATHETER. NO FRAGMENT WAS GENERATED. IT WAS REPORTED THAT THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT ANY DELAY AND WITHOUT ANY NEGATIVE IMPACT TO THE PATIENT. ON 11-JUN-2025, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE LOCATION OF THE FISTULA TARGET BY THE PROCEDURE WAS A CAVERNOUS FISTULA. THE INFORMATION INDICATED THAT THERE WAS ¿NO ATTEMPT MADE TO DETACH [THE COIL]. WHEN FEEDING COIL INTO THE MICROCATHETER, THE TECH STATED HE BROKE THE WIRE AT THE MANUAL BREAK SITE.¿ THE MICROCATHETER USED WAS A HEADWAY® DUO MICROCATHETER (TERUMO NEURO). THE COIL WAS REPLACED WITH ANOTHER 5MM X 10CM CEREPAK FREEFORM XSFT (XCR120510).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22185990 ↗MalfunctionJan 2, 2024Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT THE PHYSICIAN WAS EMBOLIZING A CAROTID-CAVERNOUS FISTULA (CCF) WITH MODERATE TORTUOSITY. THE PHYSICIAN DEPLOYED A TOTAL OF 12 COILS AD THEN USED ONYX¿ 34 LIQUID EMBOLIC AND ONYX¿ 18 LIQUID EMBOLIC (MEDTRONIC). THE FOLLOWING THREE COILS: 5MM X 10CM CEREPAK FREEFORM XSFT (XCR120510 / 31146337), 5MM X 15CM CEREPAK FREEFORM XSFT (XCR120515 / 31173107), AND 4MM X 12CM CEREPAK FREEFORM (SCR120412 / 31140706) DID NOT DETACH UTILIZING THE MANUAL DETACHMENT. IT WAS REPORTED THAT BREAKING OF THE MANUAL BREAK WAS DIFFICULT, BUT ULTIMATELY IT WAS ABLE TO EXPOSE THE PULL WIRE AND PULL BACK THE WIRE TO ATTEMPT TO DETACH THE COIL. THE DIFFICULTY WITH THE BREAKING OF THE MANUAL BREAK AT THE SORTING HAPPENED ON ALL COILS DOCUMENTED IN THIS COMPLAINT. ALL THREE COILS WERE ABLE TO BE REMOVED WITHOUT ANY NEGATIVE PATIENT IMPACT. ONE OF THE COILS REPORTEDLY DETACHED ON THE TABLE AFTER REMOVAL. THE REPORTED ISSUE RESULTED IN MINIMAL DELAY. THERE WAS NO NEGATIVE PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18432760 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page