CEREPAK FREEFORM XCR122505Recalls & reported problems
Cerenovus, Inc. · CEREPAK FREEFORM XtraSoft Detachable Coil Delivery System Length - 190 cm Coil Diameter - 2.5 mm Coil Length - 5 cm · Class 2 · Implant
All models of CEREPAK FREEFORM (60)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886704083733), received between Jun 3, 2024 and May 2, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 2, 2025Class IOpen, Classified
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Separation Failure3
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 2, 2025Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A CAROTID ARTERY PROCEDURE, THE FOLLOWING COILS WERE ADVANCED THROUGH THE EXCELSIOR SL-10® MICROCATHETER (STRYKER) INTO THE FACIAL ARTERY WITHOUT ISSUES: 4MM X 10CM CEREPAK FREEFORM XSFT (XCR120410 / 31107820), 4MM X 10CM CEREPAK FREEFORM XSFT (XCR120410 / 31112708), 4MM X 10CM CEREPAK FREEFORM (SCR120410 / 31177726), 2MM X 6CM CEREPAK FREEFORM MINI (MCR092060 / 31459691), 2.5MM X 5CM CEREPAK FREEFORM XTRASOFT (XCR122505 / 31077167), 2MM X 6CM CEREPAK HELIFORM XTRASOFT XL (XHE120206 / 31216620), AND 2MM X 8CM CEREPAK HELIFORM XSFT (XHE120208 / 31361855). FLUSH LINE WAS RUNNING. THE DISTAL ANATOMY WAS EXTREMELY TORTUOUS. THE PROXIMAL PORTION OF THE SYSTEM WAS STRAIGHT, AND MANUAL DETACHMENT WAS ATTEMPTED FIRST WITH THE DETACHMENT HANDLE (MDH1 / 102109430) AND THEN WAS SHIFTED TO MANUAL BREAK WITHOUT ANY SUCCESSFUL DETACHMENT. IT WAS REPORTED THAT OTHER CEREPAK COILS AND STRYKER COILS WERE BOTH PLACED BEFORE AND AFTER THESE COILS WITHOUT ANY ISSUE. THERE WAS NO REPORT OF ANY NEGATIVE IMPACT ON THE PATIENT. ON 30-APR-2025, RESPONSES TO ADDITIONAL INFORMATION REQUESTS WERE RECEIVED. PER THE ADDITIONAL INFORMATION, TWO DETACHMENT HANDLES
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21953814 ↗MalfunctionNov 5, 2024Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING COIL EMBOLIZATION PROCEDURE TARGETING A RUPTURED BASILAR TIP ANEURYSM, THE PHYSICIAN EXPERIENCED DETACHMENT ISSUES WITH THREE CEREPAK COILS: A 5MM X 17CM CEREPAK UNIFORM (FCX100517 / 31127138), THIS WAS THE 7TH COIL USED IN THE PROCEDURE, A 2.5MM X 5CM CEREPAK FREEFORM XTRASOFT (XCR122505 / 31077167), THIS WAS THE 15TH COIL USED IN THE PROCEDURE, AND THE 16TH COIL, ANOTHER 2.5MM X 5CM CEREPAK FREEFORM XTRASOFT (XCR122505 / 31077167). IT WAS REPORTED THAT THE PHYSICIAN ATTEMPTED TO PLACE THE COILS THROUGH AN SL-10® MICROCATHETER (STRYKER) MICROCATHETER. THEN ONCE THE COILS WERE ADVANCED INTO POSITION, THE PHYSICIAN HOOKED UP TO THE DETACHMENT HANDLE (MDH1 / LOT# UNKNOWN) AND ATTEMPT TO DETACH. FOR EACH OF THE THREE COILS, AFTER THE FIRST ATTEMPTED DETACHMENT, THE PHYSICIAN NOTED THAT THE COILS HAD NOT DETACH AND HE REATTACHED THE COILS TO THE DETACHMENT HANDLE AND MADE SECOND UNSUCCESSFUL ATTEMPT TO DETACH THE COILS. IT WAS REPORTED THAT THE COILS WERE REMOVED, AND THE PHYSICIAN CONTINUED WITH THE PROCEDURE. THERE WAS NO REPORT OF ANY PATIENT INJURY. ON 24-OCT-2024, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE TARGET
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20605959 ↗MalfunctionJun 3, 2024Separation Failure; Material Twisted/Bent; Separation Failure; Material Twisted/Bent
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION OF A RIGHT (RT) SUPERIOR CEREBELLAR ARTERY (SCA) ANEURYSM, THE FIRST COIL, A FREEFORM XSFT 2.5MM X 5 CM COIL (XCR122505, 31077167) FAILED TO DETACH USING THE MANUAL BREAK TECHNIQUE. THE TECHNOLOGIST RESPONSIBLE FOR THE DETACHMENT PERFORMED THE MANUAL BREAK TECHNIQUE AS PER THE IN-SERVICE THAT WAS CONDUCTED FOR THE PHYSICIANS AND STAFF. ENSURED THAT THE DELIVERY SYSTEM WAS STRAIGHT DURING THE DETACHMENT PROCESS. IT WAS OBSERVED THAT THE COIL FAILED TO DETACH WHEN REMOVING THE DELIVERY SYSTEM UNDER FLUORO. THE COIL WAS ADVANCED INTO THE ANEURYSM FOR A SECOND DETACHMENT ATTEMPT. THE COIL FAILED TO DETACH FOR A SECOND TIME AND THE COIL WAS REMOVED BUT DURING THE REMOVAL OF THE COIL IT DETACHED WHILE A PORTION OF THE COIL WAS IN THE CATHETER AND THE OTHER EXTENDING OUTSIDE OF THE MICROCATHETER (MC). BOTH MICROCATHETER AND COIL WERE REMOVED. NO PATIENT HARM OCCURRED DURING THE EMBOLIZATION OF THE RT SCA ANEURYSM. THE PHYSICIAN WAS ABLE TO EMBOLIZE THE ANEURYSM WITH ANOTHER CEREPAK FREEFORM XSFT 2.5MM X 5CM. THERE WAS MILD TO MODERATE TORTUOSITY. LEFT FEMORAL ACCESS, BMX 81, 5F SOFIA, AND SL10 MICROCATHETER. ADDITIONAL EVENT INFORMATION
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19446927 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page