CEREPAK HELIFORM XHE120204Recalls & reported problems
Cerenovus, Inc. · CEREPAK HELIFORM XtraSoft XL Detachable Coil Delivery System Length - 190 cm Coil Diameter - 2 mm Coil Length - 4 cm · Class 2 · Implant
All models of CEREPAK HELIFORM (48)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704083931), received between Nov 18, 2024 and Jun 18, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 2, 2025Class IOpen, Classified
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Separation Failure2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 18, 2025Separation Failure; Separation Failure
AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING A COIL EMBOLIZATION OF THE MIDDLE MENINGEAL ARTERY (MMA), A HELIFORM XSFT XL 2MM X 4 CM COIL (XHE120204, 31434959) WAS USED THROUGH A PLATO 17 160CM. THE MECHANICAL DETACHMENT HANDLE (MDH1, LOT UNKNOWN) WAS USED FOR THE FIRST DETACHMENT ATTEMPT. AFTER THE FAILED DETACHMENT ATTEMPT WITH THE HANDLE, THE MANUAL BREAK WAS USED WHICH ALSO FAILED. THE COIL WAS REMOVED SUCCESSFULLY. NO PATIENT HARM OR DELAY OF CARE CAME FROM THIS DETACHMENT FAILURE. THE LEFT MMA WAS SUCCESSFULLY EMBOLIZED WITH THE FIRST TWO CEREPAK COILS. MILD TORTUOSITY WAS NOTED IN THE LEFT EXTRACRANIAL INTERNAL CAROTID ARTERY (LECA) AND THE LEFT MIDDLE MENINGEAL ARTERY. ADDITIONAL EVENT INFORMATION RECEIVED ON 02-JUN-2025 INDICATED THAT THERE WAS NO RESISTANCE DURING ADVANCEMENT OF THE DEVICE THROUGH THE MICROCATHETER OR POSITIONING DIFFICULTY IN THE ANEURYSM. THERE WAS NO DAMAGE NOTED TO THE EMBOLIC COIL OR ANY DEVICE COMPONENT. THERE WERE NO PROCEDURAL ELAY/PROLONGATION DUE TO THE EVENT. IT WAS NOTED THAT THE CAUSE OF THE FAILURE TO DETACH IS UNKNOWN. THERE WERE TWO SUCCESSFUL COIL DETACHMENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22282663 ↗MalfunctionNov 18, 2024Separation Failure; Separation Failure
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION OF A VERY TORTOUS ANTERIOR CEREBRAL ARTERY (ACA), THE DOCTOR WAS UNABLE TO DETACH 3 COILS: FREEFORM MINI 1.5MM X 3CM (MCR091530, 31218790), FREEFORM MINI 1.5MM X 3CM (MCR091530, 31136144) AND A HELIFORM XSFT XL 2MM X 4 CM (XHE120204, 31317932). EACH COIL WAS ATTEMPTED INITIALLY WITH THE MECHANICAL DETACHMENT HANDLE (MDH1, LOT UNKNOWN). AFTER THAT WAS UNSUCCESSFUL, THE MANUAL BREAK WAS ATTEMPTED WITHOUT SUCCESS. EACH COIL WAS TAKEN OUT OF THE ANEURYSM WITHOUT ISSUES AND ANOTHER COIL WAS USED WITH THE SAME MICROCATHETER (PLATO 17 STRAIGHT). WITH EACH COIL, THE MICROCATHETER OUTSIDE THE BODY WAS STRAIGHT. NO NEGATIVE IMPACT ON PATIENT AND NO SIGNIFICANT DELAY IN TREATMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20706693 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page