Skip to content

CEREPAK HELIFORM XHE120206Recalls & reported problems

Cerenovus, Inc. · CEREPAK HELIFORM XtraSoft XL Detachable Coil Delivery System Length - 190 cm Coil Diameter - 2 mm Coil Length - 6 cm · Class 2 · Implant

All models of CEREPAK HELIFORM (48)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704083948), received between Dec 27, 2024 and May 2, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Separation Failure” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 2, 2025Class IOpen, Classified

    Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.

    Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Separation Failure2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 2, 2025Separation Failure; Separation Failure

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A CAROTID ARTERY PROCEDURE, THE FOLLOWING COILS WERE ADVANCED THROUGH THE EXCELSIOR SL-10® MICROCATHETER (STRYKER) INTO THE FACIAL ARTERY WITHOUT ISSUES: 4MM X 10CM CEREPAK FREEFORM XSFT (XCR120410 / 31107820), 4MM X 10CM CEREPAK FREEFORM XSFT (XCR120410 / 31112708), 4MM X 10CM CEREPAK FREEFORM (SCR120410 / 31177726), 2MM X 6CM CEREPAK FREEFORM MINI (MCR092060 / 31459691), 2.5MM X 5CM CEREPAK FREEFORM XTRASOFT (XCR122505 / 31077167), 2MM X 6CM CEREPAK HELIFORM XTRASOFT XL (XHE120206 / 31216620), AND 2MM X 8CM CEREPAK HELIFORM XSFT (XHE120208 / 31361855). FLUSH LINE WAS RUNNING. THE DISTAL ANATOMY WAS EXTREMELY TORTUOUS. THE PROXIMAL PORTION OF THE SYSTEM WAS STRAIGHT, AND MANUAL DETACHMENT WAS ATTEMPTED FIRST WITH THE DETACHMENT HANDLE (MDH1 / 102109430) AND THEN WAS SHIFTED TO MANUAL BREAK WITHOUT ANY SUCCESSFUL DETACHMENT. IT WAS REPORTED THAT OTHER CEREPAK COILS AND STRYKER COILS WERE BOTH PLACED BEFORE AND AFTER THESE COILS WITHOUT ANY ISSUE. THERE WAS NO REPORT OF ANY NEGATIVE IMPACT ON THE PATIENT. ON 30-APR-2025, RESPONSES TO ADDITIONAL INFORMATION REQUESTS WERE RECEIVED. PER THE ADDITIONAL INFORMATION, TWO DETACHMENT HANDLES

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21953744 ↗
  • MalfunctionDec 27, 2024Separation Failure; Separation Failure

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION OF THE MIDDLE MENINGEAL ARTERY (MMA) WITH MILD TORTUOSITY NOTED IN THE LATERAL EXTERNAL CAROTID ARTERY (LECA) AND THE LEFT MMA, A 2MM X 6CM CEREPAK HELIFORM XTRASOFT XL (XHE120206 / 31393529) WAS USED THROUGH A PLATO® 17 MICROCATHETER (SCIENTIA VASCULAR). THE DETACHMENT HANDLE (MDH1 / LOT# UNKNOWN) WAS USED FOR THE FIRST TWO DETACHMENT ATTEMPTS. AFTER THE FAILED DETACHMENT ATTEMPTS WITH THE HANDLE, THE MANUAL BREAK METHOD WAS USED, WHICH ALSO FAILED. THE COIL WAS REMOVED, AND ANOTHER COIL OF THE SAME SIZE WAS USED WITH SUCCESS. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT NOR ANY DELAY IN THE PROCEDURE. THE LEFT MMA WAS SUCCESSFULLY EMBOLIZED WITH TWO MORE CEREPAK COILS. ON 27-DEC-2024, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION THE LOT NUMBER OF THE DETACHMENT HANDLE IS NOT KNOWN. WHEN THE COIL WAS REMOVED FROM THE PATIENT, IT WAS STILL ATTACHED TO THE DELIVERY SYSTEM. ¿THE COIL WAS INSPECTED ON THE BACK TABLE AND THERE WAS NO VISIBLE DAMAGE TO THE DELIVERY SYSTEM THAT COULD HAVE CONTRIBUTED TO THE DETACHMENT FAILURE.¿ THE COIL WAS NOT STRETCHED. THERE WAS NO EVIDENCE OF BLOOD FLOW IN

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21027290 ↗

FDA clearance and classification

Cleared through 510(k) K220040 on Jul 14, 2022 (decision: Substantially Equivalent). Submitted by Medos International SARL as “CEREPAK Uniform, Uniform XL, Uniform 3D, Heliform Soft, Heliform XtraSoft, Heliform XL, Heliform XtraSoft XL, Freeform, Freeform Mini, and Freeform XtraSoft Detachable Coil Systems”. The FDA review took 190 days. FDA record ↗

The FDA regulates product code HCG (device, neurovascular embolization) under 21 CFR 882.5950 as a Class II device. 21 CFR 882.5950 ↗

Common questions about the CEREPAK HELIFORM XHE120206

Has the CEREPAK HELIFORM XHE120206 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0434-2026 (Class I), began on Oct 2, 2025: Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the CEREPAK HELIFORM XHE120206?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Dec 27, 2024 and May 2, 2025. In those reports, the most common specific device problem is “Separation Failure” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CEREPAK HELIFORM XHE120206?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CEREPAK HELIFORM XHE120206 FDA approved?

It was cleared through 510(k) K220040 on Jul 14, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the CEREPAK HELIFORM XHE120206 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the CEREPAK HELIFORM XHE120206, and is it still sold?

The FDA's device database lists Cerenovus, Inc. as the company responsible for it (brand CEREPAK HELIFORM, model XHE120206). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page