CEREPAK UNIFORM FCX100412Recalls & reported problems
Cerenovus, Inc. · CEREPAK UNIFORM Detachable Coil Delivery System Length - 190 cm Coil Diameter - 4 mm Coil Length - 11.5 cm · Class 2 · Implant
All models of CEREPAK UNIFORM (58)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704084006), received between Apr 1, 2024 and Dec 10, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 2, 2025Class IOpen, Classified
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Separation Failure2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 10, 2024Separation Failure; Separation Failure
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION, IN A VERY LARGE BASILIAR ANEURYSM, THE PHYSICIAN WAS UNABLE TO DETACH A HELIFORM XL 4MM X 15 CM COIL (SHD180415, 31098462) AND A UNIFORM 4MM X 11.5 CM COIL (FCX100412, 31207070). EACH COIL WAS ATTEMPTED INITIALLY WITH A MECHANICAL DETACHMENT HANDLE (MDH1, LOT UNKNOWN). AFTER THAT WAS UNSUCCESSFUL, THE MANUAL BREAK WAS ATTEMPTED WITHOUT SUCCESS. EACH COIL WAS TAKEN OUT OF THE ANEURYSM WITHOUT ISSUES AND ANOTHER COIL WAS USED WITH THE SAME MICROCATHETER (PLATO 17 STRAIGHT). WITH EACH COIL, THE MICROCATHETER OUTSIDE THE BODY WAS STRAIGHT. NO NEGATIVE IMPACT ON PATIENT AND NO SIGNIFICANT DELAY IN TREATMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20895296 ↗MalfunctionApr 1, 2024Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE OF AN INTRACRANIAL ANEURYSM IN A 61-YEAR-OLD MALE PATIENT WITH A HISTORY OF HYPERTENSION, THE 4MM X 11.5CM CEREPAK UNIFORM (FCX100412 / 31126840) FAILED TO DETACH AFTER MULTIPLE ATTEMPTS. THE PROCEDURE WAS NOT DELAYED; IT WAS REPORTED TO HAVE SUCCESSFULLY COMPLETED WITHOUT ANY PATIENT CONSEQUENCE. ON 20-MAR-2024, ADDITIONAL INFORMATION WAS RECEIVED FROM THE CLINICAL ACCOUNT SPECIALIST. PER THE INFORMATION, THE PROCEDURE WAS TARGETING AN ANTERIOR COMMUNICATING ARTERY (ACOM) ANEURYSM. THE DETACHMENT ATTEMPTS WERE THREE MANUAL BREAKS AND ON THE THIRD ATTEMPT, THE FILAMENT DETACHMENT WIRE BROKE. THE COIL WAS NOT STRETCHED AND IT WAS STILL ATTACHED TO THE DELIVERY SYSTEM. THE MICROCATHETER USED WAS AN EXCELSIOR SL-10 45°(STRYKER). THE PROCEDURE WAS COMPLETED WITH A SHORTER CEREPAK UNIFORM COIL, A 4MM X 7.5CM CEREPAK UNIFORM (FCX100407).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19016317 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page