1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704084006), received between Apr 1, 2024 and Dec 10, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Separation Failure” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 2, 2025Class IOpen, Classified

    Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.

    Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Separation Failure2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 10, 2024Separation Failure; Separation Failure

    AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION, IN A VERY LARGE BASILIAR ANEURYSM, THE PHYSICIAN WAS UNABLE TO DETACH A HELIFORM XL 4MM X 15 CM COIL (SHD180415, 31098462) AND A UNIFORM 4MM X 11.5 CM COIL (FCX100412, 31207070). EACH COIL WAS ATTEMPTED INITIALLY WITH A MECHANICAL DETACHMENT HANDLE (MDH1, LOT UNKNOWN). AFTER THAT WAS UNSUCCESSFUL, THE MANUAL BREAK WAS ATTEMPTED WITHOUT SUCCESS. EACH COIL WAS TAKEN OUT OF THE ANEURYSM WITHOUT ISSUES AND ANOTHER COIL WAS USED WITH THE SAME MICROCATHETER (PLATO 17 STRAIGHT). WITH EACH COIL, THE MICROCATHETER OUTSIDE THE BODY WAS STRAIGHT. NO NEGATIVE IMPACT ON PATIENT AND NO SIGNIFICANT DELAY IN TREATMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20895296 ↗
  • MalfunctionApr 1, 2024Separation Failure; Separation Failure

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE OF AN INTRACRANIAL ANEURYSM IN A 61-YEAR-OLD MALE PATIENT WITH A HISTORY OF HYPERTENSION, THE 4MM X 11.5CM CEREPAK UNIFORM (FCX100412 / 31126840) FAILED TO DETACH AFTER MULTIPLE ATTEMPTS. THE PROCEDURE WAS NOT DELAYED; IT WAS REPORTED TO HAVE SUCCESSFULLY COMPLETED WITHOUT ANY PATIENT CONSEQUENCE. ON 20-MAR-2024, ADDITIONAL INFORMATION WAS RECEIVED FROM THE CLINICAL ACCOUNT SPECIALIST. PER THE INFORMATION, THE PROCEDURE WAS TARGETING AN ANTERIOR COMMUNICATING ARTERY (ACOM) ANEURYSM. THE DETACHMENT ATTEMPTS WERE THREE MANUAL BREAKS AND ON THE THIRD ATTEMPT, THE FILAMENT DETACHMENT WIRE BROKE. THE COIL WAS NOT STRETCHED AND IT WAS STILL ATTACHED TO THE DELIVERY SYSTEM. THE MICROCATHETER USED WAS AN EXCELSIOR SL-10 45°(STRYKER). THE PROCEDURE WAS COMPLETED WITH A SHORTER CEREPAK UNIFORM COIL, A 4MM X 7.5CM CEREPAK UNIFORM (FCX100407).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19016317 ↗