Skip to content

CEREPAK UNIFORM FCX181017Recalls & reported problems

Cerenovus, Inc. · CEREPAK UNIFORM XL Detachable Coil Delivery System Length - 190 cm Coil Diameter - 10 mm Coil Length - 17.1 cm · Class 2 · Implant

All models of CEREPAK UNIFORM (58)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886704084334), received between Apr 29, 2024 and Apr 29, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Separation Failure” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 2, 2025Class IOpen, Classified

    Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.

    Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2024Separation Failure; Separation Failure

    AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION TO THE RIGHT CAROTID TERMINUS, THE PHYSICIAN HAD DETACHMENT ISSUES WITH A UNIFORM XL 10MM X 17.1 CM COIL (FCX181017, 31243967). THE MECHANICAL DETACHMENT HANDLE (MDH1, 102109430) DID NOT BREAK. HE WAS ABLE TO GET IT TO DETACH BY PUSHING THE COIL PUSHER RAPIDLY IN AND OUT. THE NEXT COIL WAS A FREEFORM 9MM X 25 CM (SCR120925, 31173116), THE HANDLE DETACHMENT FAILED, AND HE THEN TRIED THE MANUAL BREAK, THIS ONE DETACHED. THEN A FREEFORM 6MM X 20 CM COIL (SCR120620, 31058141) WAS USED. THE HANDLE FAILED AND SO DID THE MANUAL BREAK. HE TRIED TO MAKE IT DETACH BY PUSHING THE COIL PUSHER IN AND OUT IN HOPES IT WOULD RELEASE. THE PHYSICIAN THEN TRIED TO REMOVE THE COIL, WHILE PULLING IT OUT OF THE ANEURYSM THE COIL DETACHED IN THE ECHELON 10 MICROCATHETER (MC). HE WAS ABLE TO PUSH THE COIL BACK INTO THE ANEURYSM WITH A MICROWIRE. ADDITIONAL EVENT INFORMATION WAS RECEIVED ON 16-APR-2024 INDICATING THAT THE UNIFORM XL 10MM X 17.1 CM COIL WAS THE FIRST COIL TO BE USED. THERE WAS NO RESISTANCE WHILE PUSHING THE COILS THROUGH THE MICROCATHETER; THERE WAS NO DIFFICULTY POSITIONING THE MICROCATHETER INTO THE ANEURYSM HOWEVER THERE WAS DIFFICULT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19201395 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page