CEREPAK UNIFORM FCX181118Recalls & reported problems
Cerenovus, Inc. · CEREPAK UNIFORM XL Detachable Coil Delivery System Length - 190 cm Coil Diameter - 11 mm Coil Length - 18.9 cm · Class 2 · Implant
All models of CEREPAK UNIFORM (58)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886704084358), received between Dec 2, 2024 and Dec 2, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Separation Failure” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 2, 2025Class IOpen, Classified
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 2, 2024Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT ON (B)(6) 2024, DURING A RETREATMENT OF A PREVIOUSLY COILED INTERNAL CAROTID ARTERY (ICA) TERMINUS ANEURYSM, THE PHYSICIAN ATTEMPTED TO DEPLOY THE FIRST COIL, A 11MM X 18.9CM CEREPAK UNIFORM XL (FCX181118 / 31168333), WHEN HE WAS READY TO DETACH THE COIL VIA MECHANICAL DETACHMENT, THE COIL DID NOT DETACH. THE PHYSICIAN MADE ANOTHER UNSUCCESSFUL ATTEMPT. THE PHYSICIAN THEN IMPLANTED TWO SIMILAR SIZED COMPETITOR COILS. HE THEN ATTEMPTED TO DEPLOY TWO MORE CEREPAK COILS, AND THEY WORKED FINE. THE PHYSICIAN THEN ATTEMPTED TO DEPLOY TWO MORE CEREPAK FREEFORM COILS, A 6MM X 20CM CEREPAK FREEFORM (SCR120620 / 31058141) AND A 6MM X 10CM CEREPAK FREEFORM (SCR120610 / 31146353) AND THEY DID NOT DETACH. THERE WAS NO REPORT OF NEGATIVE PATIENT IMPACT OR DELAY IN THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20821313 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page