Skip to content

CEREPAK UNIFORM FCX181447Recalls & reported problems

Cerenovus, Inc. · CEREPAK UNIFORM XL Detachable Coil Delivery System Length - 190 cm Coil Diameter - 14 mm Coil Length - 47 cm · Class 2 · Implant

All models of CEREPAK UNIFORM (58)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886704084402), received between Mar 13, 2025 and Mar 13, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Separation Failure” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 2, 2025Class IOpen, Classified

    Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.

    Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 13, 2025Separation Failure; Separation Failure

    AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION OF A LARGE CAVERNOUS ANEURYSM 21X18, FLOW DIVERTER WAS PLACED IN RIGHT INTERNAL CAROTID ARTERY (RICA). THE SL-10 MICROCATHETER WAS JAILED IN ANEURYSM. A UNIFORM XL 20MM X 50 CM COIL (FCX182050, 3126862) DEPLOYED AND ATTEMPT TO DETACH WITH MECHANICAL DETACHMENT HANDLE (MDH1, LOT UNKNOWN) BUT FAILED TWICE. ATTEMPTED MANUAL BREAK, HOWEVER, IT ALSO FAILED. THE COIL WAS PULLED OUT WITH NO PROBLEM AND A NEW 18MM X 46CM CEREPAK UNIFORM DETACHED WITH MECHANICAL HANDLE. NEXT, A COIL 16MM X47CM WAS ALSO PLACED AND DETACHED WITH HANDLE. THEN, A UNIFORM XL 14MM X 47 CM COIL (FCX181447, 31177725) WAS PLACED AND DETACHMENT FAILED WITH TWO ATTEMPTS. MANUAL BREAK WAS ATTEMPTED BUT ALSO FAILED AND COIL HAD TO REMOVE WITHOUT INCIDENT. THE DOCTOR USED A NON J&J COIL AND THE PROCEDURE WAS COMPLETED. THE PATIENT WAS FINE IN POST-OP. ADDITIONAL EVENT INFORMATION RECEIVED ON 05-MAR-2025 INDICATED THAT THERE WAS NO RESISTANCE DURING ADVANCEMENT OF THE DEVICE THROUGH THE MICROCATHETER OR POSITIONING DIFFICULTY IN THE ANEURYSM. THERE WAS NO DAMAGE TO THE EMBOLIC COILS OR ANY DEVICE COMPONENT. THERE WAS SOME VESSEL TORTUOSITY BUT NOT SEVERE. THE PROCEDURE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21596416 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page