CEREPAK UNIFORM FCX181447Recalls & reported problems
Cerenovus, Inc. · CEREPAK UNIFORM XL Detachable Coil Delivery System Length - 190 cm Coil Diameter - 14 mm Coil Length - 47 cm · Class 2 · Implant
All models of CEREPAK UNIFORM (58)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886704084402), received between Mar 13, 2025 and Mar 13, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Separation Failure” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 2, 2025Class IOpen, Classified
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 13, 2025Separation Failure; Separation Failure
AS REPORTED BY THE FIELD, DURING A COIL EMBOLIZATION OF A LARGE CAVERNOUS ANEURYSM 21X18, FLOW DIVERTER WAS PLACED IN RIGHT INTERNAL CAROTID ARTERY (RICA). THE SL-10 MICROCATHETER WAS JAILED IN ANEURYSM. A UNIFORM XL 20MM X 50 CM COIL (FCX182050, 3126862) DEPLOYED AND ATTEMPT TO DETACH WITH MECHANICAL DETACHMENT HANDLE (MDH1, LOT UNKNOWN) BUT FAILED TWICE. ATTEMPTED MANUAL BREAK, HOWEVER, IT ALSO FAILED. THE COIL WAS PULLED OUT WITH NO PROBLEM AND A NEW 18MM X 46CM CEREPAK UNIFORM DETACHED WITH MECHANICAL HANDLE. NEXT, A COIL 16MM X47CM WAS ALSO PLACED AND DETACHED WITH HANDLE. THEN, A UNIFORM XL 14MM X 47 CM COIL (FCX181447, 31177725) WAS PLACED AND DETACHMENT FAILED WITH TWO ATTEMPTS. MANUAL BREAK WAS ATTEMPTED BUT ALSO FAILED AND COIL HAD TO REMOVE WITHOUT INCIDENT. THE DOCTOR USED A NON J&J COIL AND THE PROCEDURE WAS COMPLETED. THE PATIENT WAS FINE IN POST-OP. ADDITIONAL EVENT INFORMATION RECEIVED ON 05-MAR-2025 INDICATED THAT THERE WAS NO RESISTANCE DURING ADVANCEMENT OF THE DEVICE THROUGH THE MICROCATHETER OR POSITIONING DIFFICULTY IN THE ANEURYSM. THERE WAS NO DAMAGE TO THE EMBOLIC COILS OR ANY DEVICE COMPONENT. THERE WAS SOME VESSEL TORTUOSITY BUT NOT SEVERE. THE PROCEDURE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21596416 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page