Skip to content

CEREPAK UNIFORM FCX181846Recalls & reported problems

Cerenovus, Inc. · CEREPAK UNIFORM XL Detachable Coil Delivery System Length - 190 cm Coil Diameter - 18 mm Coil Length - 46 cm · Class 2 · Implant

All models of CEREPAK UNIFORM (58)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704084440), received between Dec 12, 2023 and Dec 12, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Separation Failure” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 2, 2025Class IOpen, Classified

    Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.

    Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Separation Failure2
  • Difficult to Advance1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 12, 2023Separation Failure; Separation Failure

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A PIPELINE / COIL EMBOLIZATION CASE IN TARGETING A RIGHT DISTAL INTERNAL CAROTID ARTERY (ICA) ANEURYSM OVER 2CM IN SIZE WITH A NECK MEASURED AT ALMOST 20MM IN A VERY TORTUOUS VASCULATURE, AFTER MANY ATTEMPTS WITH MULTIPLE WIRES AND CATHETERS, ACCESS WAS OBTAINED INTO THE RIGHT MIDDLE CEREBRAL ARTERY (MCA) AND A 5MM X 35MM PIPELINE EMBOLIZATION DEVICE (PED) (MEDTRONIC) WAS DEPLOYED. A HEADWAY® DUO 156 MICROCATHETER (MICROVENTION) WAS JAILED INTO THE ANEURYSM PRIOR TO THE DEPLOYMENT OF THE PED. THE FIRST 18MM X 46CM CEREPAK UNIFORM XL COIL (FCX181846 / 31168338) WAS ADVANCED THROUGH THE MICROCATHETER MINIMAL RESISTANCE AROUND THE SECOND MANUAL BREAK MARKER. THE COIL WAS INSERTED INTO THE ANEURYSM WITHOUT DIFFICULT. THE MANUAL BREAK TECHNIQUE WAS ATTEMPTED WITHOUT SUCCESS. THE WIRE HAD FRACTURED AND THE PHYSICIAN ATTEMPTED TO REMOVE THE WIRE WITH A FORCEPS; THE PHYSICIAN ALSO ATTEMPT TO RETRACT THE EXPOSE WIRE WITHOUT SUCCESS. THE COIL WAS THEN REMOVED WITHOUT ISSUE AND WAS FOUND INTACT IN THE FIELD. A SECOND 18MM X 46CM CEREPAK UNIFORM XL COIL FROM THE SAME LOT (31168338) WAS ADVANCED WITHOUT ANY ISSUE. A CEREPAKTM DETACHER (MDH1 / 102

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18315587 ↗
  • MalfunctionDec 12, 2023Material Separation; Separation Failure; Difficult to Advance; Material Separation; Separation Failure; Difficult to Advance

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A PIPELINE / COIL EMBOLIZATION CASE IN TARGETING A RIGHT DISTAL INTERNAL CAROTID ARTERY (ICA) ANEURYSM OVER 2CM IN SIZE WITH A NECK MEASURED AT ALMOST 20MM IN A VERY TORTUOUS VASCULATURE, AFTER MANY ATTEMPTS WITH MULTIPLE WIRES AND CATHETERS, ACCESS WAS OBTAINED INTO THE RIGHT MIDDLE CEREBRAL ARTERY (MCA) AND A 5MM X 35MM PIPELINE EMBOLIZATION DEVICE (PED) (MEDTRONIC) WAS DEPLOYED. A HEADWAY® DUO 156 MICROCATHETER (MICROVENTION) WAS JAILED INTO THE ANEURYSM PRIOR TO THE DEPLOYMENT OF THE PED. THE FIRST 18MM X 46CM CEREPAK UNIFORM XL COIL (FCX181846 / 31168338) WAS ADVANCED THROUGH THE MICROCATHETER MINIMAL RESISTANCE AROUND THE SECOND MANUAL BREAK MARKER. THE COIL WAS INSERTED INTO THE ANEURYSM WITHOUT DIFFICULT. THE MANUAL BREAK TECHNIQUE WAS ATTEMPTED WITHOUT SUCCESS. THE WIRE HAD FRACTURED AND THE PHYSICIAN ATTEMPTED TO REMOVE THE WIRE WITH A FORCEPS; THE PHYSICIAN ALSO ATTEMPT TO RETRACT THE EXPOSE WIRE WITHOUT SUCCESS. THE COIL WAS THEN REMOVED WITHOUT ISSUE AND WAS FOUND INTACT IN THE FIELD. A SECOND 18MM X 46CM CEREPAK UNIFORM XL COIL FROM THE SAME LOT (31168338) WAS ADVANCED WITHOUT ANY ISSUE. A CEREPAKTM DETACHER (MDH1 / 102

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18315585 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page