CEREPAK UNIFORM FCX181846Recalls & reported problems
Cerenovus, Inc. · CEREPAK UNIFORM XL Detachable Coil Delivery System Length - 190 cm Coil Diameter - 18 mm Coil Length - 46 cm · Class 2 · Implant
All models of CEREPAK UNIFORM (58)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704084440), received between Dec 12, 2023 and Dec 12, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 2, 2025Class IOpen, Classified
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Separation Failure2
- Difficult to Advance1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 12, 2023Separation Failure; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A PIPELINE / COIL EMBOLIZATION CASE IN TARGETING A RIGHT DISTAL INTERNAL CAROTID ARTERY (ICA) ANEURYSM OVER 2CM IN SIZE WITH A NECK MEASURED AT ALMOST 20MM IN A VERY TORTUOUS VASCULATURE, AFTER MANY ATTEMPTS WITH MULTIPLE WIRES AND CATHETERS, ACCESS WAS OBTAINED INTO THE RIGHT MIDDLE CEREBRAL ARTERY (MCA) AND A 5MM X 35MM PIPELINE EMBOLIZATION DEVICE (PED) (MEDTRONIC) WAS DEPLOYED. A HEADWAY® DUO 156 MICROCATHETER (MICROVENTION) WAS JAILED INTO THE ANEURYSM PRIOR TO THE DEPLOYMENT OF THE PED. THE FIRST 18MM X 46CM CEREPAK UNIFORM XL COIL (FCX181846 / 31168338) WAS ADVANCED THROUGH THE MICROCATHETER MINIMAL RESISTANCE AROUND THE SECOND MANUAL BREAK MARKER. THE COIL WAS INSERTED INTO THE ANEURYSM WITHOUT DIFFICULT. THE MANUAL BREAK TECHNIQUE WAS ATTEMPTED WITHOUT SUCCESS. THE WIRE HAD FRACTURED AND THE PHYSICIAN ATTEMPTED TO REMOVE THE WIRE WITH A FORCEPS; THE PHYSICIAN ALSO ATTEMPT TO RETRACT THE EXPOSE WIRE WITHOUT SUCCESS. THE COIL WAS THEN REMOVED WITHOUT ISSUE AND WAS FOUND INTACT IN THE FIELD. A SECOND 18MM X 46CM CEREPAK UNIFORM XL COIL FROM THE SAME LOT (31168338) WAS ADVANCED WITHOUT ANY ISSUE. A CEREPAKTM DETACHER (MDH1 / 102
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18315587 ↗MalfunctionDec 12, 2023Material Separation; Separation Failure; Difficult to Advance; Material Separation; Separation Failure; Difficult to Advance
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A PIPELINE / COIL EMBOLIZATION CASE IN TARGETING A RIGHT DISTAL INTERNAL CAROTID ARTERY (ICA) ANEURYSM OVER 2CM IN SIZE WITH A NECK MEASURED AT ALMOST 20MM IN A VERY TORTUOUS VASCULATURE, AFTER MANY ATTEMPTS WITH MULTIPLE WIRES AND CATHETERS, ACCESS WAS OBTAINED INTO THE RIGHT MIDDLE CEREBRAL ARTERY (MCA) AND A 5MM X 35MM PIPELINE EMBOLIZATION DEVICE (PED) (MEDTRONIC) WAS DEPLOYED. A HEADWAY® DUO 156 MICROCATHETER (MICROVENTION) WAS JAILED INTO THE ANEURYSM PRIOR TO THE DEPLOYMENT OF THE PED. THE FIRST 18MM X 46CM CEREPAK UNIFORM XL COIL (FCX181846 / 31168338) WAS ADVANCED THROUGH THE MICROCATHETER MINIMAL RESISTANCE AROUND THE SECOND MANUAL BREAK MARKER. THE COIL WAS INSERTED INTO THE ANEURYSM WITHOUT DIFFICULT. THE MANUAL BREAK TECHNIQUE WAS ATTEMPTED WITHOUT SUCCESS. THE WIRE HAD FRACTURED AND THE PHYSICIAN ATTEMPTED TO REMOVE THE WIRE WITH A FORCEPS; THE PHYSICIAN ALSO ATTEMPT TO RETRACT THE EXPOSE WIRE WITHOUT SUCCESS. THE COIL WAS THEN REMOVED WITHOUT ISSUE AND WAS FOUND INTACT IN THE FIELD. A SECOND 18MM X 46CM CEREPAK UNIFORM XL COIL FROM THE SAME LOT (31168338) WAS ADVANCED WITHOUT ANY ISSUE. A CEREPAKTM DETACHER (MDH1 / 102
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18315585 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page