CEREPAK UNIFORM FCX180714Recalls & reported problems

Cerenovus, Inc. · CEREPAK UNIFORM XL Detachable Coil Delivery System Length - 190 cm Coil Diameter - 7 mm Coil Length - 14.3 cm · Class 2 · Implant

All models of CEREPAK UNIFORM (58)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886704084488), received between Dec 14, 2023 and May 28, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Separation Failure” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 2, 2025Class IOpen, Classified

    Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.

    Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Separation Failure2
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 28, 2024Separation Failure; Separation Failure

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE WHERE THE PHYSICIAN ACCESSED A PREVIOUSLY TREATED, RECANALIZED ANEURYSM WITH AN SL-10® MICROCATHETER (STRYKER). THE PHYSICIAN THEN INSERTED THE COMPLAINT COIL, A 7MM X 14.3CM CEREPAK UNIFORM XL (FCX180714 / 31173115). IT WAS REPORTED THAT ¿FELLOW WENT SLIGHTLY PASSED THE INVERTED T AND WENT TO USE MANUAL BREAK BEFORE BEING TOLD TO PULL BACK A LITTLE. WHEN HE REMOVED THE DELIVERY SYSTEM, THE COIL WAS STILL ATTACHED. [THE] PULL WIRE WAS BROKEN AT THE MANUAL BREAK SITE.¿ AN ATTEMPT TO PULL THE PULL WIRE WITH A HEMOSTAT WAS MADE, BUT ¿UNABLE TO GRASP.¿ THE COIL AND DELIVERY SYSTEM WAS REMOVED FROM THE PATIENT. IT WAS REPORTED THAT THE PROCEDURE WAS NOT DELAYED, IT WAS SUCCESSFULLY COMPLETED. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT. ON 28-MAY-2024, ADDITIONAL INFORMATION WAS RECEIVED. PER THE INFORMATION, THE INTENDED PROCEDURE WAS COILING OF THE PREVIOUSLY TREATED RIGHT MIDDLE CEREBRAL ARTERY (MCA) ANEURYSM. THE ANEURYSM WAS PREVIOUSLY TREATED IN FEBRUARY 2024 (EXACT DATE WAS NOT PROVIDED) AT THE SAME HOSPITAL BY THE SAME PHYSICIAN. THE INFORMATION INDICATED THAT WHEN THE 7MM X 14.3CM CER

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19411816 ↗
  • MalfunctionDec 14, 2023Separation Failure; Premature Separation; Separation Failure; Premature Separation

    AS REPORTED BY THE FIELD, THIS WAS A VERY TORTUOUS PIPELINE/COILING CASE OF THE LEFT DISTAL INTERNAL CAROTID ARTERY (ICA) OVER A 9MM IN SIZE, NECK MEASURING ALMOST 8MM. A PIPELINE EMBOLIZATION DEVICE WAS DEPLOYED. A HEADWAY DUO 156CM MICROCATHETERWAS JAILED INTO THE ANEURYSM PRIOR TO THE DEPLOYMENT OF THE PIPELINE. THE FIRST OF THE UNIFORM XL 7MM X 14.3 CM COIL ((B)(6)) WAS ADVANCED THROUGH THE MICROCATHETER (MC) WITH MINIMAL RESISTANCE AROUND THE SECOND MANUAL BREAK MARKER. THE COIL WAS INSERTED INTO THE ANEURYSM WITHOUT DIFFICULTY. THE MANUAL BREAK TECHNIQUE WAS ATTEMPTED WITHOUT SUCCESS. AS THE DOCTOR ATTEMPTED TO THEN PULL THE COIL OUT OF THE ANEURYSM, THE COIL SEEMED TO PREMATURELY DETACH IN THE MICROCATHETER. HE WAS ABLE TO PUSH THE COIL BACK INTO THE ANEURYSM. THE REST OF THE CASE WAS COILED WITH THE DP2-3 CASHMEERE COILS WITHOUT ANY ISSUE. NO NEGATIVE PATIENT OUTCOME AND MINIMAL DELAY OF TREATMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18327626 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →