- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886704084778), received between Apr 3, 2024 and Apr 3, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Premature Separation” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Oct 2, 2025Open, Classified
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
Z-0434-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Premature Separation1
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 3, 2024Separation Failure; Premature Separation; Separation Failure; Premature Separation
THE HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT PRESENTED TO (B)(6) HOSPITAL WITH AN INTRACRANIAL HEMORRHAGE. AFTER A VERY DIFFICULT ACCESS, THE PHYSICIAN WAS ABLE TO NAVIGATE THE ANEURYSM THROUGH TORTUOUS ANATOMY. USING A HEADWAY® DUO MICROCATHETER (MICROVENTION), THE PHYSICIAN ATTEMPTED TO DEPLOY THE FIRST COIL, A 5MM X 15CM CEREPAK FREEFORM XTRASOFT (XCR120515 / 31111683). THE COIL FAILED TO DETACH WITH THE CEREPAK TM MECHANICAL DETACHMENT HANDLE (MDH1 / 102109430). MANUAL BREAK WAS ATTEMPTED WITHOUT SUCCESS. THE COIL WAS REPLACED WITH ANOTHER 5MM X 15CM CEREPAK FREEFORM XTRASOFT (XCR120515) FROM THE SAME LOT (31111683) HERE MANUAL BREAK AS THE FIRST METHOD OF DETACHMENT, AND IT WAS ALSO UNSUCCESSFUL. IT WAS REPORTED THAT AT THIS POINT, A 2MM X 2CM CEREPAK FREEFORM MINI (MCR092020 / 31177313) WAS ADVANCED. ONCE AGAIN, MANUAL BREAK WAS CHOSEN AT THE PRIMARY DETACHMENT METHOD, AND AGAIN, THE COIL DID NOT DETACH. AT THIS POINT, THE COIL WAS REMOVED AND THE PHYSICIAN NOTICED WHAT APPEARED TO BE ¿FILAMENT¿ RETAINED WITHIN THE MICROCATHETER. THE PHYSICIAN WAS ABLE TO REMOVE [THE FILAMENT] WITHOUT INCIDENT AND THE PROCEDURE WAS MOVED FORWARD USING MICROVENTION COILS. THERE WAS NO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19038169 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.