CEREGLIDE NIC71137URecalls & reported problems
Cerenovus, Inc. · CEREGLIDE 71 Intermediate Catheter 137 cm 0.071 in (1.80 mm) 0.0820 in (2.08 mm) 0.0825 in (2.10 mm) · Class 2
All models of CEREGLIDE (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886704085386), received between May 19, 2025 and May 5, 2026 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Deformation Due to Compressive Stress2
- Adverse Event Without Identified Device or Use Problem1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Thromboembolism1
- Vascular Dissection1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2026Deformation Due to Compressive Stress
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A STROKE PROCEDURE, THE PHYSICIAN USED A 132CM CEREGLIDE 71 INTERMEDIATE CATHETER (NIC71132U / 31668075) THROUGH A ZOOM¿88 ASPIRATION CATHETER (IMPERATIVE CARE). BOTH CATHETERS WERE WELL-HYDRATED (FLUSHED), THE CEREGLIDE REPORTEDLY GOT STUCK IN THE ZOOM 88. DURING THE ADVANCEMENT OF THE CEREGLIDE 71 CATHETER, THE PHYSICIAN REPORTED THAT IT FELT TIGHT AND AS A RESULT, THE PHYSICIAN REMOVED IT AND THE CEREGLIDE 71 CATHETER STRETCHED. THE PHYSICIAN TRIED THE SECOND CEREGLIDE, A 137 CM CEREGLIDE 71 CATHETER (NIC71137U / 31684676). EVERYTHING FLUSHED WELL BUT THE SAME ISSUE WAS ENCOUNTERED. THERE WAS NO NEGATIVE IMPACT ON THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED ON 04-MAY-2026. PER THE INFORMATION, THE PROCEDURE WAS TARGETING THE M1 SEGMENT OF THE LEFT MIDDLE CEREBRAL ARTERY. THE TARGET VESSEL WAS NOT EXCESSIVELY TORTUOUS. THE SECOND CEREGLIDE CATHETER WAS ALSO USED WITH THE ZOOM 88. THE INFORMATION INDICATED THAT FOR BOTH CEREGLIDE CATHETERS, THE ISSUE OCCURRED ON THE FIRST PASS. AFTER THE SECOND CEREGLIDE CATHETER, THE PHYSICIAN USED A ZOOM 7MM X 137MM TO COMPLETE THE PROCEDURE. THE ZOOM 88 WAS ALSO USED TO COMPLETE THE PROCEDUR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25086331 ↗MalfunctionApr 27, 2026Deformation Due to Compressive Stress; Physical Resistance/Sticking
A HEALTHCARE PROFESSIONAL REPORTED THAT DURING A MECHANICAL THROMBECTOMY TO THE LEFT MIDDLE CEREBRAL ARTERY, SEGMENT M1, THE PHYSICIAN WAS TRYING TO INSERT THE 137CM CEREGLIDE 71 CATHETER (PRODUCT CODE: NIC71137U, LOT NUMBER: UNKNOWN) INTO A COPILOT TOUHEY THAT WAS CONNECTED TO A ZOOM 88 BASE CATHETER AND HE FELT RESISTANCE. HE REMOVED THE CEREGLIDE AND NOTICED IT WAS KINKED. ADDITIONAL EVENT INFORMATION RECEIVED ON 22-APR-2026 INDICATED THAT THE KINK OCCURRED IN THE PATIENT, PRIOR TO ASPIRATION OR DELIVERING STENT RETRIEVER. THE CATHETER TIP WAS INSIDE THE TOUEY WHEN THE DAMAGE WAS OBSERVED. THE DEVICE/SYSTEM WAS REMOVED OR REPLACED TO CONTINUE THE PROCEDURE. THE EVENT DID NOT RESULT IN ANY UNDUE PROCEDURAL DELAY THAT COULD BE CONSIDERED CLINICALLY SIGNIFICANT. THERE WAS NO DIFFICULTY IN REMOVING DEVICE FROM PACKAGING. A DIRECT ASPIRATION FIRST PASS TECHNIQUE (ADAPT) WAS USED. THE DEVICE WAS NOT SNAKED. THE ANARTOMOY WAS NOTED AS ¿NORMAL¿. THERE WAS NO BREAK IN MATERIAL INTEGRITY AT THE SITE OF THE DEFECT, SUCH AS CRACKING OR FRACTURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24999606 ↗InjuryMay 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A THROMBECTOMY PROCEDURE THAT WAS TARGETING A THROMBUS IN THE M1 SEGMENT OF THE MIDDLE CEREBRAL ARTERY (MCA), THE PHYSICIAN WAS USING CEREGLIDE 92 INNERGLIDE 9 122 CM CATHETER SYSTEM (SBC92122UG / 31485382) AND A 137 CM CEREGLIDE 71 CATHETER (NIC71137U / 31528444). IT WAS REPORTED THAT, ¿AFTER [THE] FIRST PASS, 2A AND DISTAL MIGRATION OF CLOT. [A SOLITAIRE¿ STENT RETRIEVER (MEDTRONIC)] WAS USED AND TICI3 WAS OBSERVED WITH SMALL DISSECTION. NO TREATMENT OF DISSECTION WAS REQUIRED, AND [THE] PATIENT WAS STABLE. THIS IS NOT A PARTICULAR DEVICE COMPLICATION, RATHER A CASE NOTATION.¿
Patient problems: Thromboembolism; Vascular Dissection
MAUDE report 22048776 ↗
Other Catheter, Percutaneous, Neurovasculature devices
- EMBOGUARD BG8795UNeuravi Limited · Catheter, Percutaneous, Neurovasculature2 deaths153 reports
- CEREBASE GS9090SDCerenovus, Inc. · Catheter, Percutaneous, Neurovasculature1 recall2 deaths97 reports
- CEREBASE GS9095SDCerenovus, Inc. · Catheter, Percutaneous, Neurovasculature1 recall1 death33 reports
- CEREBASE GS9080SDCerenovus, Inc. · Catheter, Percutaneous, Neurovasculature1 recall13 reports
- Balloon Guide Catheter BOB-895-USMicrovention, Inc. · Catheter, Percutaneous, Neurovasculature1 recall11 reports
- 070 Access System Gen 1Route 92 Medical · Catheter, Percutaneous, Neurovasculature1 recall0 reports
- EMBOGUARD BG8785UNeuravi Limited · Catheter, Percutaneous, Neurovasculature1 death13 reports
- CEREGLIDE SBC92122UGCerenovus, Inc. · Catheter, Percutaneous, Neurovasculature9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page