RIGIDFIX 213708Recalls & reported problems
Medos International Sàrl · RIGIDFIX Femoral Rod Soft Tissue Only 8mm, 30-120mm (Plus or Minus 6 Percent) · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886705003020), received between Jul 17, 2018 and Aug 8, 2022 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Mechanical Jam2
- Break1
- Device-Device Incompatibility1
Patient problems
- No Consequences Or Impact To Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 8, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED BY THE SALES REP THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE RIGIDFIX FEMORAL ROD 8 MM ST 1PK DEVICE DID NOT FIT INTO THE GUIDE. THERE WAS A SMALL DELAY TO ATTACH ROD TO THE GUIDE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15187119 ↗MalfunctionJan 22, 2019Break; Break
THIS IS REPORT 2 OF 2 FOR THE SAME EVENT. IT WAS REPORTED BY THE SALES REP THAT DURING AN ANTERIOR CRUCIATE LIGAMENT ACL RECONSTRUCTION SURGICAL PROCEDURE, AFTER REMOVING THE SOFT TISSUE U-SHAPED GUIDE FROM THE PATIENT'S KNEE, THE TECH IN THE BACK TABLE TRIED TO UNSCREW THE 8MM ROD FROM THE SOFT TISSUE GUIDE AND THE BOLT THAT HOLDS BOTH GUIDES TOGETHER BROKE OFF; LEAVING THE 8MM GUIDE ATTACHED TO THE SOFT TISSUE GUIDE. TOURNIQUET TIME WAS NOT AFFECTED FROM THE NORM. THE GUIDE WAS NO LONGER NEEDED DURING THAT CASE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8266721 ↗MalfunctionNov 26, 2018Mechanical Jam; Mechanical Jam
THE SALES REPRESENTATIVE REPORTED THAT THE CUSTOMERS RIGIDFIX FEMORAL RODD 8MM IS WORN AT THE TIP AREA, AND THE 7MM RIGIDFIX FEMORAL ROD WAS WARPED AT THE TIP. THE SURGEON WAS ABLE TO COMPLETE THE PROCEDURE USING OTHER LIKE DEVICES, WITH NO DELAY AND NO PATIENT CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8102424 ↗MalfunctionJul 17, 2018Mechanical Jam; Mechanical Jam
IT WAS REPORTED BY SALES REPRESENTATIVE THAT THE SURGEON USED A RIGIDFIX FEMORAL ST GUIDE FRAME, WITH AN 8MM ST FEMORAL ROD, ON A PATIENT IN AN ACL PROCEDURE. AFTER THE SURGEON WAS FINISHED USING THE DEVICE IN THE PROCEDURE, THE CUSTOMER WAS HAVING SOME DIFFICULTY TRYING TO UNSCREW THE THUMBSCREW ON THE DEVICE WHEN THE THUMBSCREW BROKE OFF, PREVENTING THE FEMORAL ROD FROM BEING REMOVED FROM THE GUIDE FRAME. POST USE, NO PATIENT CONSEQUENCES. PLEASE SEE ALSO MEDWATCH REPORT 1221934-2018-51369.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7693882 ↗
Other Guide, surgical, instrument devices
- NA 03.037.017Synthes GmbH · Guide, surgical, instrument346 reports
- EVOS 71174934Smith & Nephew, Inc. · Guide, surgical, instrument149 reports
- LCP 323.042Synthes GmbH · Guide, surgical, instrument129 reports
- DJO SURGICAL 342-10-705Encore Medical, L.P. · Guide, Surgical, Instrument1 recall51 reports
- DJO SURGICAL 341-10-707Encore Medical, L.P. · Guide, Surgical, Instrument81 reports
- NA 03.010.404Synthes GmbH · Guide, surgical, instrument73 reports
- DJO SURGICAL 341-14-708Encore Medical, L.P. · Guide, Surgical, Instrument1 recall22 reports
- DJO SURGICAL 341-10-711Encore Medical, L.P. · Guide, Surgical, Instrument1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page