RIGIDFIX 213709Recalls & reported problems
Medos International Sàrl · RIGIDFIX Femoral Rod 9mm, 30-120mm (Plus or Minus 6 Percent) · Class 1
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10886705003037), received between Jun 13, 2018 and Dec 11, 2023 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility3
- Break2
- Device Markings/Labelling Problem2
- Mechanical Jam2
- Connection Problem1
- Device Difficult to Setup or Prepare1
- Separation Failure1
Patient problems
- No Consequences Or Impact To Patient9
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 11, 2023Device-Device Incompatibility; Device-Device Incompatibility
THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) REPAIR PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE RIGIDFIX FEMORAL ST GUIDE FRAME N/S 1PK WAS BENT/TORQUED AND THE DRILL SLEEVES AND TROCAR WERE BEING PREVENTED FROM CROSSING THE WINDOW OF THE RIGIDFIX FEMORAL ROD 9 MM 1PK. THE DISTAL SLEEVE EVENTUALLY WENT ACROSS. THE PHYSICIAN TRIED TO REPOSITION THE GUIDE FOR THE PROXIMAL PIN, AFTER REAPPROXIMATING THE GUIDE AT A DIFFERENT ANGLE, AND REDRILLING, THEY WERE ABLE TO GET THE PROXIMAL PIN THROUGH THE FRAME. THE SKIVE MARKS WERE VISIBLE IN THE SLEEVE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18301182 ↗MalfunctionDec 11, 2023Device-Device Incompatibility; Device-Device Incompatibility
THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) REPAIR PROCEDURE ON 11/28/2023, IT WAS OBSERVED THAT THE RIGIDFIX FEMORAL ST GUIDE FRAME N/S 1PK WAS BENT/TORQUED AND THE DRILL SLEEVES AND TROCAR WERE BEING PREVENTED FROM CROSSING THE WINDOW OF THE RIGIDFIX FEMORAL ROD 9 MM 1PK. THE DISTAL SLEEVE EVENTUALLY WENT ACROSS. THE PHYSICIAN TRIED TO REPOSITION THE GUIDE FOR THE PROXIMAL PIN, AFTER REAPPROXIMATING THE GUIDE AT A DIFFERENT ANGLE, AND REDRILLING, THEY WERE ABLE TO GET THE PROXIMAL PIN THROUGH THE FRAME. THE SKIVE MARKS WERE VISIBLE IN THE SLEEVE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18301042 ↗MalfunctionMay 18, 2020Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED BY AFFILIATE VIA EMAIL, THAT PRE-OPERATIVELY TO AN ACL PROCEDURE, POTENTIAL PROBLEM WAS IDENTIFIED BY HOSPITAL CSD AND AN ALTERNATIVE SET WAS USED. DAMAGED INSTRUMENTS ARE: RIGIDFIX FEMROD 9MM *EA (EXPANDS AND CANNOT FIT INTO GUIDE FRAME), RIGIDFIX FEM B/T/B STEPTROC *EA (THEIR ONE IS BARBED AT THE TIP), RIGIDFIX FEM GDEFR B/T/B *EA (SCREW HAS BEEN CROSS THREADED AND WILL NOT GO INTO GUIDE). DEVICES ARE AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10067456 ↗MalfunctionSep 23, 2019Device Difficult to Setup or Prepare; Connection Problem
IT WAS REPORTED BY THE SALES REP VIA PHONE THAT PRIOR TO USE ON THE PATIENT DURING AN ACL PROCEDURE, THE SALES REP'S RIGIDFIX FEMORAL ROD 9MM WOULD NOT LATCH INTO PLACE WITH A MATING DEVICE. THE CASE WAS COMPLETED WITH A DIFFERENT SIZED LIKE-DEVICE WITH NO PATIENT HARM OR DELAYS TO THE CASE. THE SALES REP COULD NOT PROVIDE A LOT NUMBER FOR THIS DEVICE. THERE WAS NO DELAY IN THE SURGICAL PROCEDURE. THERE WAS A SPARE DEVICE AVAILABLE FOR USE TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9103959 ↗MalfunctionJan 14, 2019Mechanical Jam; Mechanical Jam
THIS IS REPORT 2 OF 2 FOR THE SAME EVENT. IT WAS REPORTED BY THE SALES REP THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE THUMB SCREW ON THEIR RIGIDFIX FEMORAL BTB GUIDE FRAME BROKE AND THEIR RIGIDFIX FEMORAL ROD 9MM WAS STUCK IN THE GUIDE FRAME. THE SALES REP STATED THAT WHERE THE DEVICE BROKE WAS OUTSIDE THE PATIENT. THE SALES REP REPORTED THAT THE SURGEON HAD COMPLETED THE PROCEDURE WITH THE DEVICE BEFORE THE DEVICE FAILURE OCCURRED. THE SALES REP REPORTED THAT THERE WERE NO PATIENT CONSEQUENCES OR DELAYS IN THE CASE. THE SALES REP COULD NOT PROVIDE LOT NUMBERS FOR THE DEVICES BUT STATED THAT THE GUIDE FRAME WAS LESS THAN A YEAR OLD AND THE ROD WAS APPROXIMATELY FOUR YEARS OLD, BOTH DEVICES HAVE SEEN HEAVY USE IN THE FIELD. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8246131 ↗MalfunctionNov 9, 2018Break; Break
THE SALES REPRESENTATIVE REPORTED THAT DURING AN ACL REPAIR THAT THE THUMBSCREW BROKE OFF HIS RIGIDFIX ST U GUIDE WHEN THE SURGEON WAS MALLETING THE DEVICE IN. THE SURGEON COMPLETED THE PROCEDURE WITH THE DEVICE WITH NO PATIENT CONSEQUENCES OR DELAY. THE 9MM ROD WAS NOT ABLE TO BE REMOVED FROM THE GUIDEFRAME AND WILL BE RETURNED WITH THE COMPLAINT DEVICE. THE SALES REPRESENTATIVE COULD NOT PROVIDE A LOT NUMBER FOR THE DEVICES BUT REPORTED THAT THEY WERE OVER THREE (3) YEARS OLD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8054873 ↗MalfunctionSep 17, 2018Separation Failure; Separation Failure
THE SALES REP REPORTED THE THUMBSCREW ON THE CUSTOMER'S RIGIDFIX BTB GUIDE BROKE OFF WHILE THE SURGEON WAS TIGHTENING THE THUMBSCREW TO SECURE THE FEMORAL ROD IN THE GUIDE DURING AN ACL PROCEDURE. THE SURGEON STILL USED THE GUIDE TO COMPLETE THE PROCEDURE WITH NO PATIENT CONSEQUENCES OR DELAYS. THE THREADS OF THE THUMBSCREW ARE STUCK IN THE GUIDE ALSO MAKING THE ROD STUCK IN THE GUIDE. THE ENTIRE GUIDE WITH THE ROD STUCK IN THE DEVICE ARE COMING BACK FOR EVALUATION. DUE TO THIS, THE ENTIRE GUIDE AND THE ROD WERE REPLACED FOR THE CUSTOMER. THE SALES REP DID NOT HAVE LOT NUMBERS BUT VERIFIED THE DEVICES WERE OVER 4 YEARS OLD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7879942 ↗MalfunctionAug 6, 2018Break; Break
IT WAS REPORTED BY THE HOSPITAL THAT DURING ANTERIOR CRUCIATE LIGAMENT (ACL) RECONSTRUCTION PROCEDURE; AFTER REAMING INTO THE FEMUR AND TIBIA, A 9MM AIMING GUIDE WAS SCREWED INTO RIGIDFIX U-SHAPED GUIDE. WHILE TIGHTENING THE BOLT ON THE U-SHAPED GUIDE TO THE 9MM AIMING GUIDE, THE BOLT BROKE OFF LEAVING A PIECE OF THE BOLT INSIDE THE U-SHAPED GUIDE CAUSING THE 9MM AIMING GUIDE TO BE STUCK INSIDE THE U-SHAPED GUIDE.-NO ADVERSE EFFECTS OCCURRED NOR ANY EXTENDED PROCEDURE TIME. SECURED THE AIMING GUIDE TO THE OPPOSITE SIDE OF THE U-SHAPED GUIDE WITH ANOTHER BOLT. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7751836 ↗MalfunctionJul 11, 2018Mechanical Jam; Mechanical Jam
THE SALES REPRESENTATIVE REPORTED THAT THE CUSTOMER'S SOFT TISSUE RIGIDFIX GUIDE AND 9' RIGIDFIX ROD BECAME STUCK TOGETHER WHEN THE SURGEON WAS PERFORMING AN ACL REPAIR. THE SALES REPRESENTATIVE REPORTED THAT WHEN THE SURGEON WAS UNSCREWING THE THUMBSCREW AT THE END OF THE PROCEDURE THE SCREW BROKE IN HIS HAND. THE SURGEON COMPLETED THE PROCEDURE WITH NO PATIENT CONSEQUENCES OR DELAY. THE SALES REPRESENTATIVE COULD NOT PROVIDE THE LOT NUMBER OR AGE OF THE DEVICES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7677775 ↗MalfunctionJun 13, 2018Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED BY THE SALES REP THAT "SOFT TISSUE ONLY" WAS ETCHED INTO TWO OF HIS RIGIDFIX FEMORAL ROD 10MM AND TWO OF HIS RIGIDFIX FEMORAL ROD 9MM WHEN THEY ARE IN FACT BTB DEVICES. THERE WAS NO CASE OR PATIENT INVOLVEMENT. THE DEVICES ARE BEING RETURNED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7596058 ↗MalfunctionJun 13, 2018Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED BY THE SALES REP THAT "SOFT TISSUE ONLY" WAS ETCHED INTO TWO OF HIS RIGIDFIX FEMORAL ROD 10MM AND TWO OF HIS RIGIDFIX FEMORAL ROD 9MM WHEN THEY ARE IN FACT BTB DEVICES. THERE WAS NO CASE OR PATIENT INVOLVEMENT. THE DEVICES ARE BEING RETURNED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7596051 ↗
Other Guide, surgical, instrument devices
- NA 03.037.017Synthes GmbH · Guide, surgical, instrument346 reports
- EVOS 71174934Smith & Nephew, Inc. · Guide, surgical, instrument149 reports
- LCP 323.042Synthes GmbH · Guide, surgical, instrument129 reports
- DJO SURGICAL 342-10-705Encore Medical, L.P. · Guide, Surgical, Instrument1 recall51 reports
- DJO SURGICAL 341-10-707Encore Medical, L.P. · Guide, Surgical, Instrument81 reports
- NA 03.010.404Synthes GmbH · Guide, surgical, instrument73 reports
- DJO SURGICAL 341-14-708Encore Medical, L.P. · Guide, Surgical, Instrument1 recall22 reports
- DJO SURGICAL 341-10-711Encore Medical, L.P. · Guide, Surgical, Instrument1 recall16 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.